64-section multidetector CT angiography for evaluation of intracranial aneurysms: comparison with 3D rotational angiography

被引:18
作者
Li, Qi [1 ,2 ]
Lv, Fajin
Yao, Guoen [3 ]
Li, Yongmei [2 ]
Xie, Peng [1 ]
机构
[1] Chongqing Med Univ, Affiliated Hosp 1, Dept Neurol, Chongqing 400016, Peoples R China
[2] Chongqing Med Univ, Affiliated Hosp 1, Dept Radiol, Chongqing 400016, Peoples R China
[3] Third Mil Med Univ, Daping Hosp, Dept Neurol, Chongqing, Peoples R China
基金
中国国家自然科学基金;
关键词
CT angiography; intracranial aneurysm; rotational angiography; DIGITAL-SUBTRACTION-ANGIOGRAPHY; COMPUTERIZED-TOMOGRAPHY ANGIOGRAPHY; SUBARACHNOID HEMORRHAGE; CEREBRAL ANEURYSMS; MICROANEURYSMS; SURGERY; 2D;
D O I
10.1177/0284185113506138
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To assess the diagnostic performance of 64-section computed tomography angiography (CTA) for detection of cerebral aneurysms by comparison with the new gold standard three-dimensional (3D) digital subtraction angiography (DSA). Material and Methods: A total of 128 patients who underwent both 64-section CTA and 3D DSA for suspected intracranial aneurysms were included. The location, size, and shape of the aneurysm were assessed and compared with 3D DSA results. The sensitivity, specificity, positive and negative predictive values of 64-section CTA were calculated by using 3D DSA as reference standard. Results: According to the reference standard, a total of 145 aneurysms were detected at 3D DSA in 118 patients. Of these 145 aneurysms, 64-section CTA detected 139 aneurysms in 118 patients. The sensitivity of CTA for detection of aneurysm was 95.8% on a per-aneurysm basis. All missed aneurysms were non-causative very small aneurysms in patients harboring multiple aneurysms. On a per-patient basis, the sensitivity, specificity, positive and negative predictive values were 100%. Conclusion: Compared with the new gold standard 3D DSA, 64-section CTA offers high sensitivity and specificity for detection of intracranial aneurysms. It could be readily used as a screening imaging method for detection of intracranial aneurysms.
引用
收藏
页码:840 / 846
页数:7
相关论文
共 50 条
  • [31] 64-Slice CT Angiography in the Detection of Intracranial Aneurysms: Comparison with DSA and Surgical Findings
    Medenica, S. Milosevic
    Vuckovic, V.
    Prstojevic, B.
    NEURORADIOLOGY JOURNAL, 2010, 23 (01) : 55 - 61
  • [32] CT angiography, MR angiography and rotational digital subtraction angiography for volumetric assessment of intracranial aneurysms.: An experimental study
    Piotin, M
    Gailloud, P
    Bidaut, L
    Mandai, S
    Muster, M
    Moret, J
    Rüfenacht, DA
    NEURORADIOLOGY, 2003, 45 (06) : 404 - 409
  • [33] CT angiography, MR angiography and rotational digital subtraction angiography for volumetric assessment of intracranial aneurysms. An experimental study
    M. Piotin
    P. Gailloud
    L. Bidaut
    S. Mandai
    M. Muster
    J. Moret
    D. A. Rüfenacht
    Neuroradiology, 2003, 45 : 404 - 409
  • [34] Multislice computerized tomography angiography in the evaluation of intracranial aneurysms: a comparison with intraarterial digital subtraction angiography
    Wintermark, M
    Uske, A
    Chalaron, M
    Regli, L
    Maeder, P
    Meuli, R
    Schnyder, P
    Binaghi, S
    JOURNAL OF NEUROSURGERY, 2003, 98 (04) : 828 - 836
  • [35] Intracranial aneurysms: evaluation in 200 patients with spiral CT angiography
    Young, N
    Dorsch, NWC
    Kingston, RJ
    Markson, G
    McMahon, J
    EUROPEAN RADIOLOGY, 2001, 11 (01) : 123 - 130
  • [36] Intracranial aneurysms: evaluation in 200 patients with spiral CT angiography
    N. Young
    N. W. C. Dorsch
    R. J. Kingston
    G. Markson
    J. McMahon
    European Radiology, 2001, 11 : 123 - 130
  • [37] Optimization of radiation settings for angiography using 3D fluoroscopy for imaging of intracranial aneurysms
    Linsenmann, Thomas
    Marz, Alexander
    Duffner, Vera
    Stetter, Christian
    Weiland, Judith
    Westermaier, Thomas
    COMPUTER ASSISTED SURGERY, 2021, 26 (01) : 22 - 30
  • [38] 320-Detector row CT angiography for detection and evaluation of intracranial aneurysms: Comparison with conventional digital subtraction angiography
    Wang, H.
    Li, W.
    He, H.
    Luo, L.
    Chen, C.
    Guo, Y.
    CLINICAL RADIOLOGY, 2013, 68 (01) : E15 - E20
  • [39] Subtracted 3D CT angiography for the evaluation of intracranial aneurysms in 256-slice multidetector CT: usefulness of the 80-kVp plus compact contrast medium bolus protocol
    Kidoh, Masafumi
    Nakaura, Takeshi
    Ogata, Takaaki
    Takashima, Hiroki
    Yoshikawa, Makoto
    Uemura, Shouzaburou
    Harada, Kazunori
    Yamashita, Yasuyuki
    EUROPEAN RADIOLOGY, 2013, 23 (11) : 3012 - 3019
  • [40] Comparison of 2D Digital Subtraction Angiography and 3D Rotational Angiography in the Evaluation of Dome-to-Neck Ratio