An experimental phantom study of the effect of gadolinium-based MR contrast agents on PET attenuation coefficients and PET quantification in PET-MR imaging: application to cardiac studies

被引:2
作者
O'Doherty, Jim [1 ]
Schleyer, Paul [2 ]
机构
[1] Kings Coll London, PET Imaging Ctr, Div Imaging Sci & Biomed Engn, Kings Hlth Partners,St Thomas Hosp, 1st Floor,Lambeth Wing,Westminster Bridge Rd, London SE1 7EH, England
[2] Siemens Healthcare Ltd, Frimley, Camberley, England
来源
EJNMMI PHYSICS | 2017年 / 4卷
基金
英国惠康基金; 英国工程与自然科学研究理事会;
关键词
PET-MR; Attenuation correction; Cardiac PET; MYOCARDIAL-PERFUSION;
D O I
10.1186/s40658-017-0173-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Simultaneous cardiac perfusion studies are an increasing trend in PET-MR imaging. During dynamic PET imaging, the introduction of gadolinium-based MR contrast agents (GBCA) at high concentrations during a dual injection of GBCA and PET radiotracer may cause increased attenuation effects of the PET signal, and thus errors in quantification of PET images. We thus aimed to calculate the change in linear attenuation coefficient (LAC) of a mixture of PET radiotracer and increasing concentrations of GBCA in solution and furthermore, to investigate if this change in LAC produced a measurable effect on the image-based PET activity concentration when attenuation corrected by three different AC strategies. Findings: We performed simultaneous PET-MR imaging of a phantom in a static scenario using a fixed activity of 40 MBq [18 F]-NaF, water, and an increasing GBCA concentration from 0 to 66 mM (based on an assumed maximum possible concentration of GBCA in the left ventricle in a clinical study). This simulated a range of clinical concentrations of GBCA. We investigated two methods to calculate the LAC of the solution mixture at 511 keV: (1) a mathematical mixture rule and (2) CT imaging of each concentration step and subsequent conversion to LAC at 511 keV. This comparison showed that the ranges of LAC produced by both methods are equivalent with an increase in LAC of the mixed solution of approximately 2% over the range of 0-66 mM. We then employed three different attenuation correction methods to the PET data: (1) each PET scan at a specific millimolar concentration of GBCA corrected by its corresponding CT scan, (2) each PET scan corrected by a CT scan with no GBCA present (i.e., at 0 mM GBCA), and (3) a manually generated attenuation map, whereby all CT voxels in the phantom at 0 mM were replaced by LAC = 0.1 cm(-1). All attenuation correction methods (1-3) were accurate to the true measured activity concentration within 5%, and there were no trends in image-based activity concentrations upon increasing the GBCA concentration of the solution. (Continued on next page) Conclusion: The presence of high GBCA concentration (representing a worst-case scenario in dynamic cardiac studies) in solution with PET radiotracer produces a minimal effect on attenuation-corrected PET quantification.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Whole-body FDG PET-MR oncologic imaging: pitfalls in clinical interpretation related to inaccurate MR-based attenuation correction
    Ulrike Attenberger
    Ciprian Catana
    Hersh Chandarana
    Onofrio A Catalano
    Kent Friedman
    Stefan A Schonberg
    James Thrall
    Marco Salvatore
    Bruce R. Rosen
    Alexander R. Guimaraes
    Abdominal Imaging, 2015, 40 : 1374 - 1386
  • [42] Comparison of MR-based attenuation correction and CT-based attenuation correction of whole-body PET/MR imaging
    David Izquierdo-Garcia
    Stephen J. Sawiak
    Karin Knesaurek
    Jagat Narula
    Valentin Fuster
    Joseph Machac
    Zahi A. Fayad
    European Journal of Nuclear Medicine and Molecular Imaging, 2014, 41 : 1574 - 1584
  • [43] Comparison of MR-based attenuation correction and CT-based attenuation correction of whole-body PET/MR imaging
    Izquierdo-Garcia, David
    Sawiak, Stephen J.
    Knesaurek, Karin
    Narula, Jagat
    Fuster, Valentin
    Machac, Joseph
    Fayad, Zahi A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 (08) : 1574 - 1584
  • [44] PET/MR brain imaging: evaluation of clinical UTE-based attenuation correction
    Lars Birger Aasheim
    Anna Karlberg
    Pål Erik Goa
    Asta Håberg
    Sveinung Sørhaug
    Unn-Merete Fagerli
    Live Eikenes
    European Journal of Nuclear Medicine and Molecular Imaging, 2015, 42 : 1439 - 1446
  • [45] Impact of Time-of-Flight PET on Quantification Errors in MR Imaging-Based Attenuation Correction
    Mehranian, Abolfazl
    Zaidi, Habib
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (04) : 635 - 641
  • [46] MLAA Based RF Surface Coil Attenuation Estimation in Hybrid PET/MR Imaging
    Heusser, Thorsten
    Rank, Christopher M.
    Freitag, Martin T.
    Kachelriess, Marc
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [47] PET/MR brain imaging: evaluation of clinical UTE-based attenuation correction
    Aasheim, Lars Birger
    Karlberg, Anna
    Goa, Pal Erik
    Haberg, Asta
    Sorhaug, Sveinung
    Fagerli, Unn-Merete
    Eikenes, Live
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (09) : 1439 - 1446
  • [48] Hybrid PET- and MR-driven attenuation correction for enhanced 18F-NaF and 18F-FDG quantification in cardiovascular PET/MR imaging
    Karakatsanis, Nicolas A.
    Abgral, Ronan
    Trivieri, Maria Giovanna
    Dweck, Marc R.
    Robson, Philip M.
    Calcagno, Claudia
    Boeykens, Gilles
    Senders, Max L.
    Mulder, Willem J. M.
    Tsoumpas, Charalampos
    Fayad, Zahi A.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2020, 27 (04) : 1126 - 1141
  • [49] Hybrid PET- and MR-driven attenuation correction for enhanced 18F-NaF and 18F-FDG quantification in cardiovascular PET/MR imaging
    Nicolas A. Karakatsanis
    Ronan Abgral
    Maria Giovanna Trivieri
    Marc R. Dweck
    Philip M. Robson
    Claudia Calcagno
    Gilles Boeykens
    Max L. Senders
    Willem J. M. Mulder
    Charalampos Tsoumpas
    Zahi A. Fayad
    Journal of Nuclear Cardiology, 2020, 27 : 1126 - 1141
  • [50] Quality control for quantitative multicenter whole-body PET/MR studies: A NEMA image quality phantom study with three current PET/MR systems
    Boellaard, Ronald
    Rausch, Ivo
    Beyer, Thomas
    Delso, Gaspar
    Yaqub, Maqsood
    Quick, Harald H.
    Sattler, Bernhard
    MEDICAL PHYSICS, 2015, 42 (10) : 5961 - 5969