An experimental phantom study of the effect of gadolinium-based MR contrast agents on PET attenuation coefficients and PET quantification in PET-MR imaging: application to cardiac studies

被引:2
作者
O'Doherty, Jim [1 ]
Schleyer, Paul [2 ]
机构
[1] Kings Coll London, PET Imaging Ctr, Div Imaging Sci & Biomed Engn, Kings Hlth Partners,St Thomas Hosp, 1st Floor,Lambeth Wing,Westminster Bridge Rd, London SE1 7EH, England
[2] Siemens Healthcare Ltd, Frimley, Camberley, England
来源
EJNMMI PHYSICS | 2017年 / 4卷
基金
英国惠康基金; 英国工程与自然科学研究理事会;
关键词
PET-MR; Attenuation correction; Cardiac PET; MYOCARDIAL-PERFUSION;
D O I
10.1186/s40658-017-0173-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Simultaneous cardiac perfusion studies are an increasing trend in PET-MR imaging. During dynamic PET imaging, the introduction of gadolinium-based MR contrast agents (GBCA) at high concentrations during a dual injection of GBCA and PET radiotracer may cause increased attenuation effects of the PET signal, and thus errors in quantification of PET images. We thus aimed to calculate the change in linear attenuation coefficient (LAC) of a mixture of PET radiotracer and increasing concentrations of GBCA in solution and furthermore, to investigate if this change in LAC produced a measurable effect on the image-based PET activity concentration when attenuation corrected by three different AC strategies. Findings: We performed simultaneous PET-MR imaging of a phantom in a static scenario using a fixed activity of 40 MBq [18 F]-NaF, water, and an increasing GBCA concentration from 0 to 66 mM (based on an assumed maximum possible concentration of GBCA in the left ventricle in a clinical study). This simulated a range of clinical concentrations of GBCA. We investigated two methods to calculate the LAC of the solution mixture at 511 keV: (1) a mathematical mixture rule and (2) CT imaging of each concentration step and subsequent conversion to LAC at 511 keV. This comparison showed that the ranges of LAC produced by both methods are equivalent with an increase in LAC of the mixed solution of approximately 2% over the range of 0-66 mM. We then employed three different attenuation correction methods to the PET data: (1) each PET scan at a specific millimolar concentration of GBCA corrected by its corresponding CT scan, (2) each PET scan corrected by a CT scan with no GBCA present (i.e., at 0 mM GBCA), and (3) a manually generated attenuation map, whereby all CT voxels in the phantom at 0 mM were replaced by LAC = 0.1 cm(-1). All attenuation correction methods (1-3) were accurate to the true measured activity concentration within 5%, and there were no trends in image-based activity concentrations upon increasing the GBCA concentration of the solution. (Continued on next page) Conclusion: The presence of high GBCA concentration (representing a worst-case scenario in dynamic cardiac studies) in solution with PET radiotracer produces a minimal effect on attenuation-corrected PET quantification.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] An experimental phantom study of the effect of gadolinium-based MR contrast agents on PET attenuation coefficients and PET quantification in PET-MR imaging: application to cardiac studies
    Jim O’ Doherty
    Paul Schleyer
    EJNMMI Physics, 4
  • [2] Gadolinium-based contrast agent attenuation does not impact PET quantification in simultaneous dynamic contrast enhanced breast PET/MR
    Allen, Timothy J.
    Henze Bancroft, Leah C.
    Kumar, Manoj
    Bradshaw, Tyler J.
    Strigel, Roberta M.
    McMillan, Alan B.
    Fowler, Amy M.
    MEDICAL PHYSICS, 2022, 49 (08) : 5206 - 5215
  • [3] PET-MR imaging: A great opportunity for enhanced quantification in PET and beyond
    Buvat, I.
    MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2015, 39 (03): : 291 - 292
  • [4] The effect of MR surface coils on PET quantification in whole-body PET/MR: Results from a pseudo-PET/MR phantom study
    Tellmann, L.
    Quick, H. H.
    Bockisch, A.
    Herzog, H.
    Beyer, T.
    MEDICAL PHYSICS, 2011, 38 (05) : 2795 - 2805
  • [5] Effect of MR contrast agents on quantitative accuracy of PET in combined whole-body PET/MR imaging
    Cristina Lois
    Ilja Bezrukov
    Holger Schmidt
    Nina Schwenzer
    Matthias K. Werner
    Jürgen Kupferschläger
    Thomas Beyer
    European Journal of Nuclear Medicine and Molecular Imaging, 2012, 39 : 1756 - 1766
  • [6] Effect of MR contrast agents on quantitative accuracy of PET in combined whole-body PET/MR imaging
    Lois, Cristina
    Bezrukov, Ilja
    Schmidt, Holger
    Schwenzer, Nina
    Werner, Matthias K.
    Kupferschlaeger, Juergen
    Beyer, Thomas
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 (11) : 1756 - 1766
  • [7] MR-based motion correction for PET imaging using wired active MR microcoils in simultaneous PET-MR: Phantom studya)
    Huang, Chuan
    Ackerman, Jerome L.
    Petibon, Yoann
    Brady, Thomas J.
    El Fakhri, Georges
    Ouyang, Jinsong
    MEDICAL PHYSICS, 2014, 41 (04)
  • [8] The effect of high count rates on cardiac perfusion quantification in a simultaneous PET-MR system using a cardiac perfusion phantom
    O'Doherty, Jim
    Chalampalakis, Zacharias
    Schleyer, Paul
    Nazir, Muhummad Sohaib
    Chiribiri, Amedeo
    Marsden, Paul K.
    EJNMMI PHYSICS, 2017, 4
  • [9] Comparison of Image Quality According to Phantom Fluid Material in MR-based Attenuation Correction PET Imaging in PET/MR
    Park, Chan Rok
    JOURNAL OF MAGNETICS, 2021, 26 (03) : 299 - 304
  • [10] The effect of high count rates on cardiac perfusion quantification in a simultaneous PET-MR system using a cardiac perfusion phantom
    Jim O’ Doherty
    Zacharias Chalampalakis
    Paul Schleyer
    Muhummad Sohaib Nazir
    Amedeo Chiribiri
    Paul K. Marsden
    EJNMMI Physics, 4