Automated Non-Contact Respiratory Rate Monitoring of Neonates Based on Synchronous Evaluation of a 3D Time-of-Flight Camera and a Microwave Interferometric Radar Sensor

被引:14
|
作者
Gleichauf, Johanna [1 ]
Herrmann, Sven [1 ]
Hennemann, Lukas [1 ]
Krauss, Hannes [1 ]
Nitschke, Janina [1 ]
Renner, Philipp [1 ]
Niebler, Christine [1 ]
Koelpin, Alexander [2 ]
机构
[1] Nuremberg Inst Technol, D-90489 Nurnberg, Germany
[2] Hamburg Univ Technol, Inst High Frequency Technol, D-21073 Hamburg, Germany
关键词
non-contact monitoring; neonates; synchronous evaluation; respiratory rate; RGB-D CAMERA; REAL-TIME; VITAL SIGNS; 6-PORT; SYSTEM; UWB;
D O I
10.3390/s21092959
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper introduces an automatic non-contact monitoring method based on the synchronous evaluation of a 3D time-of-flight (ToF) camera and a microwave interferometric radar sensor for measuring the respiratory rate of neonates. The current monitoring on the Neonatal Intensive Care Unit (NICU) has several issues which can cause pressure marks, skin irritations and eczema. To minimize these risks, a non-contact system made up of a 3D time-of-flight camera and a microwave interferometric radar sensor is presented. The 3D time-of-flight camera delivers 3D point clouds which can be used to calculate the change in distance of the moving chest and from it the respiratory rate. The disadvantage of the ToF camera is that the heartbeat cannot be determined. The microwave interferometric radar sensor determines the change in displacement caused by the respiration and is even capable of measuring the small superimposed movements due to the heartbeat. The radar sensor is very sensitive towards movement artifacts due to, e.g., the baby moving its arms. To allow a robust vital parameter detection the data of both sensors was evaluated synchronously. In this publication, we focus on the first step: determining the respiratory rate. After all processing steps, the respiratory rate determined by the radar sensor was compared to the value received from the 3D time-of-flight camera. The method was validated against our gold standard: a self-developed neonatal simulation system which can simulate different breathing patterns. In this paper, we show that we are the first to determine the respiratory rate by evaluating the data of an interferometric microwave radar sensor and a ToF camera synchronously. Our system delivers very precise breaths per minute (BPM) values within the norm range of 20-60 BPM with a maximum difference of 3 BPM (for the ToF camera itself at 30 BPM in normal mode). Especially in lower respiratory rate regions, i.e., 5 and 10 BPM, the synchronous evaluation is required to compensate the drawbacks of the ToF camera. In the norm range, the ToF camera performs slightly better than the radar sensor.
引用
收藏
页数:26
相关论文
共 31 条
  • [1] Automatic non-contact monitoring of the respiratory rate of neonates using a structured light camera
    Gleichauf, Johanna
    Niebler, Christine
    Koelpin, Alexander
    42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 4118 - 4121
  • [2] Non-contact Breathing Rate Detection Based on Time of Flight Sensor
    Yang, Chengxu
    Huang, Xinxin
    Zheng, Yu
    Xie, Yufei
    Duan, Xiaohui
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 7284 - 7287
  • [3] FPGA Based Time-of-Flight 3D Camera Characterization System
    Seiter, Johannes
    Hofbauer, Michael
    Davidovic, Milos
    Zimmermann, Horst
    PROCEEDINGS OF THE 2013 IEEE 16TH INTERNATIONAL SYMPOSIUM ON DESIGN AND DIAGNOSTICS OF ELECTRONIC CIRCUITS & SYSTEMS (DDECS), 2013, : 240 - 245
  • [4] Non-contact microwave sensor for FDM 3D printing quality control
    Slot, M.
    Samolej, K.
    Bartosik, M.
    Drabik, P.
    Zasada, I.
    VIRTUAL AND PHYSICAL PROTOTYPING, 2024, 19 (01)
  • [5] Automated Detection of Movements During Sleep Using a 3D Time-of-Flight Camera: Design and Experimental Evaluation
    Gall, Markus
    Garn, Heinrich
    Kohn, Bernhard
    Bajic, Katharina
    Coronel, Carmina
    Seidel, Stefan
    Mandl, Magdalena
    Kaniusas, Eugenijus
    IEEE ACCESS, 2020, 8 : 109144 - 109155
  • [6] MOEMS-based Time-of-Flight Camera for 3D Video Capturing
    You, Jang-Woo
    Park, Yong-Hwa
    Cho, Yong-Chul
    Park, Chang-Young
    Yoon, Heesun
    Lee, Sang-Hun
    Lee, Seung-Wan
    MOEMS AND MINIATURIZED SYSTEMS XII, 2013, 8616
  • [7] Time-of-Flight camera based 3D point cloud reconstruction of a car
    Hoegg, Thomas
    Lefloch, Damien
    Kolb, Andreas
    COMPUTERS IN INDUSTRY, 2013, 64 (09) : 1099 - 1114
  • [8] Non-contact monitoring of current distribution in a PCB using 3D magnetic sensor
    Tran, Do Phuong Uyen
    Nguyen, Tien Anh
    Mathias, Herve
    Martincic, Emile
    Lefebvre, Stephane
    Joubert, Pierre-Yves
    2022 SYMPOSIUM ON DESIGN, TEST, INTEGRATION AND PACKAGING OF MEMS/MOEMS, DTIP, 2022,
  • [9] An Autonomous Forklift with 3D Time-of-Flight Camera-Based Localization and Navigation
    Behrje, Ulrich
    Himstedt, Marian
    Maehle, Erik
    2018 15TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION, ROBOTICS AND VISION (ICARCV), 2018, : 1739 - 1746
  • [10] DETECTING RESPIRATORY EVENTS BY RESPIRATORY EFFORT DERIVED FROM 3D TIME-OF-FLIGHT CAMERA AND SPO2
    Coronel, Carmina
    Wiesmeyr, Christoph
    Garn, Heinrich
    Kohn, Bernhard
    Wimmer, Markus
    Mandl, Magdalena
    Glos, Martin
    Penzel, Thomas
    Kloesch, Gerhard
    Stefanic-Kejik, Andrijana
    Boeck, Marion
    Kaniusas, Eugenijus
    Seidel, Stefan
    SLEEP, 2019, 42