Quantitative Testing of fMRI-Compatibility of an Electrically Active Mechatronic Device for Robot-Assisted Sensorimotor Protocols

被引:6
作者
Farrens, Andria J. [1 ]
Zonnino, Andrea [1 ]
Erwin, Andrew [2 ]
O'Malley, Marcia K. [2 ]
Johnson, Curtis L. [3 ]
Ress, David [4 ]
Sergi, Fabrizio [1 ]
机构
[1] Univ Delaware, Dept Biomed Engn, Human Robot Lab, Newark, DE 19713 USA
[2] Rice Univ, Dept Mech Engn, Mech & Hapt Interfaces Lab, Houston, TX 77251 USA
[3] Univ Delaware, Dept Biomed Engn, Newark, DE USA
[4] Baylor Coll Med, Dept Neurosci, Houston, TX 77030 USA
关键词
fMRI; MR-compatible mechatronics; RF interference; robotics; signal-to-noise ratio; sensorimotor learning; FUNCTIONAL CONNECTIVITY; MRI; BRAIN; MULTICENTER; MOVEMENTS; DYNAMICS; DESIGN; ERRORS;
D O I
10.1109/TBME.2017.2741346
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: To develop a quantitative set of methods for testing the functional magnetic resonance imaging (fMRI) compatibility of an electrically-active mechatronic device developed to support sensorimotor protocols during fMRI. Methods: The set of methods includes phantom and in vivo experiments to measure the effect of a progressively broader set of noise sources potentially introduced by the device. Phantom experiments measure the radio-frequency (RF) noise and temporal noise-to-signal ratio (tNSR) introduced by the device. The in vivo experiment assesses the effect of the device on measured brain activation for a human subject performing a representative sensorimotor task. The proposed protocol was validated via experiments using a 3T MRI scanner operated under nominal conditions and with the inclusion of an electrically-active mechatronic device - the MR-SoftWrist - as the equipment under test (EUT). Results: Quantitative analysis of RF noise data allows detection of active RF noise sources both in controlled RF noise conditions, and in conditions resembling improper filtering of the EUT's electrical signals. In conditions where no RF noise was detectable, the presence and operation of the EUT did not introduce any significant increase in tNSR. A quantitative analysis conducted on in vivo measurements of the number of active voxels in visual and motor areas further showed no significant difference between EUT and baseline conditions. Conclusion and significance: The proposed set of quantitative methods supports the development and troubleshooting of electrically-active mechatronic devices for use in sensorimotor protocols with fMRI, and may be used for future testing of such devices.
引用
收藏
页码:1595 / 1606
页数:12
相关论文
共 40 条
  • [31] The neural correlates of learned motor acuity
    Shmuelof, Lior
    Yang, Juemin
    Caffo, Brian
    Mazzoni, Pietro
    Krakauer, John W.
    [J]. JOURNAL OF NEUROPHYSIOLOGY, 2014, 112 (04) : 971 - 980
  • [32] Automated quality assurance routines for fMRI data applied to a multicenter study
    Stöcker, T
    Schneider, F
    Klein, M
    Habel, U
    Kellermann, T
    Zilles, K
    Shah, NJ
    [J]. HUMAN BRAIN MAPPING, 2005, 25 (02) : 237 - 246
  • [33] Sulzer J., 2013, IEEE INT C REHABILIT, P1
  • [34] Investigation and validation of intersite fMRI studies using the same Imaging hardware
    Sutton, Bradley P.
    Goh, Joshua
    Hebrank, Andrew
    Welsh, Robert C.
    Chee, Michael W. L.
    Park, Denise C.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (01) : 21 - 28
  • [35] Vanello N., 2015, IEEE-ASME T MECH, V13, P345
  • [36] Design and Evaluation of a Cable-Driven fMRI-Compatible Haptic Interface to Investigate Precision Grip Control
    Vigaru, Bogdan
    Sulzer, James
    Gassert, Roger
    [J]. IEEE TRANSACTIONS ON HAPTICS, 2016, 9 (01) : 20 - 32
  • [37] Weinreb Jeffrey, 2005, J Am Coll Radiol, V2, P494, DOI 10.1016/j.jacr.2004.11.004
  • [38] Simple measurement of scanner stability for functional NMR imaging of activation in the brain
    Weisskoff, RM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (04) : 643 - 645
  • [39] Yu N., 2010, INT J COMPUT ASSIST, V6, P473
  • [40] Yu N., 2010, INT J COMPUT ASSIST, V6, P447