Quantitative Testing of fMRI-Compatibility of an Electrically Active Mechatronic Device for Robot-Assisted Sensorimotor Protocols

被引:6
作者
Farrens, Andria J. [1 ]
Zonnino, Andrea [1 ]
Erwin, Andrew [2 ]
O'Malley, Marcia K. [2 ]
Johnson, Curtis L. [3 ]
Ress, David [4 ]
Sergi, Fabrizio [1 ]
机构
[1] Univ Delaware, Dept Biomed Engn, Human Robot Lab, Newark, DE 19713 USA
[2] Rice Univ, Dept Mech Engn, Mech & Hapt Interfaces Lab, Houston, TX 77251 USA
[3] Univ Delaware, Dept Biomed Engn, Newark, DE USA
[4] Baylor Coll Med, Dept Neurosci, Houston, TX 77030 USA
关键词
fMRI; MR-compatible mechatronics; RF interference; robotics; signal-to-noise ratio; sensorimotor learning; FUNCTIONAL CONNECTIVITY; MRI; BRAIN; MULTICENTER; MOVEMENTS; DYNAMICS; DESIGN; ERRORS;
D O I
10.1109/TBME.2017.2741346
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: To develop a quantitative set of methods for testing the functional magnetic resonance imaging (fMRI) compatibility of an electrically-active mechatronic device developed to support sensorimotor protocols during fMRI. Methods: The set of methods includes phantom and in vivo experiments to measure the effect of a progressively broader set of noise sources potentially introduced by the device. Phantom experiments measure the radio-frequency (RF) noise and temporal noise-to-signal ratio (tNSR) introduced by the device. The in vivo experiment assesses the effect of the device on measured brain activation for a human subject performing a representative sensorimotor task. The proposed protocol was validated via experiments using a 3T MRI scanner operated under nominal conditions and with the inclusion of an electrically-active mechatronic device - the MR-SoftWrist - as the equipment under test (EUT). Results: Quantitative analysis of RF noise data allows detection of active RF noise sources both in controlled RF noise conditions, and in conditions resembling improper filtering of the EUT's electrical signals. In conditions where no RF noise was detectable, the presence and operation of the EUT did not introduce any significant increase in tNSR. A quantitative analysis conducted on in vivo measurements of the number of active voxels in visual and motor areas further showed no significant difference between EUT and baseline conditions. Conclusion and significance: The proposed set of quantitative methods supports the development and troubleshooting of electrically-active mechatronic devices for use in sensorimotor protocols with fMRI, and may be used for future testing of such devices.
引用
收藏
页码:1595 / 1606
页数:12
相关论文
共 40 条
  • [1] Study design in MRI: Basic principles
    Amaro, E
    Barker, GJ
    [J]. BRAIN AND COGNITION, 2006, 60 (03) : 220 - 232
  • [2] FUNCTIONAL CONNECTIVITY IN THE MOTOR CORTEX OF RESTING HUMAN BRAIN USING ECHO-PLANAR MRI
    BISWAL, B
    YETKIN, FZ
    HAUGHTON, VM
    HYDE, JS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (04) : 537 - 541
  • [3] Bo Yang, 2011, IEEE International Conference on Robotics and Automation, P674
  • [4] The central nervous system stabilizes unstable dynamics by learning optimal impedance
    Burdet, E
    Osu, R
    Franklin, DW
    Milner, TE
    Kawato, M
    [J]. NATURE, 2001, 414 (6862) : 446 - 449
  • [5] Neural correlates of reach errors
    Diedrichsen, J
    Hashambhoy, Y
    Rane, T
    Shadmehr, R
    [J]. JOURNAL OF NEUROSCIENCE, 2005, 25 (43) : 9919 - 9931
  • [6] Kinesthetic Feedback During 2DOF Wrist Movements via a Novel MR-Compatible Robot
    Erwin, Andrew
    O'Malley, Marcia K.
    Ress, David
    Sergi, Fabrizio
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2017, 25 (09) : 1489 - 1499
  • [7] Erwin A, 2015, INT C REHAB ROBOT, P187, DOI 10.1109/ICORR.2015.7281197
  • [8] Fischer GS, 2008, LECT NOTES COMPUT SC, V5242, P509, DOI 10.1007/978-3-540-85990-1_61
  • [9] THE COORDINATION OF ARM MOVEMENTS - AN EXPERIMENTALLY CONFIRMED MATHEMATICAL-MODEL
    FLASH, T
    HOGAN, N
    [J]. JOURNAL OF NEUROSCIENCE, 1985, 5 (07) : 1688 - 1703
  • [10] Report on a multicenter fMRI quality assurance protocol
    Friedman, L
    Glover, GH
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 23 (06) : 827 - 839