共 7 条
Development of a two-dimensional multi-grid-type microstrip gas chamber for spallation neutron source
被引:15
作者:
Takahashi, H
Siritiprussamee, P
Kai, M
Nakazawa, M
Ino, T
Furusaka, M
Kishimoto, S
Taniguchi, M
Ito, S
Kanazawa, M
机构:
[1] Univ Tokyo, Sch Engn, Bunkyo Ku, Tokyo 1138656, Japan
[2] High Energy Accelerator Res Org, Inst Mat Struct Sci, Tsukuba, Ibaraki 3050801, Japan
[3] Futaba Corp, Chosei, Chiba 2994393, Japan
来源:
关键词:
MSGC;
gas detector;
position-sensitive detector;
encoding method;
GLG method;
D O I:
10.1016/j.nima.2004.05.010
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
Several types of two-dimensional multi-grid-type micro strip gas chambers (M-MSGCs) are being developed for the new spallation neutron source in Japan. In this paper, we propose a new type of M-MSGC using multi-layer technology. In order to reduce the number of readout lines, we introduce a new encoding scheme, the Global-Local Grouping (GLG) method. In this approach, global and local signals are used to locate the coarse and fine position, respectively. We have designed, fabricated and tested the GLG M-MSGC plate with an active area of 10 x 10 cm(2). The results confirmed that the new type of M-MSGC can be operated properly. (C) 2004 Published by Elsevier B.V.
引用
收藏
页码:348 / 353
页数:6