Validation of a short-scan-time imaging protocol for thallium-201 myocardial SPECT with a multifocal collimator

被引:16
作者
Horiguchi, Yoriko [1 ]
Ueda, Tomohiro [2 ]
Shiomori, Tomofumi [2 ]
Kanna, Masahiko [1 ]
Matsushita, Hirooki [1 ]
Kawaminami, Tomoko [1 ]
Sudo, Yuta [1 ]
Kikuchi, Shinnosuke [1 ]
Sasaki, Ryo [2 ]
Hoshimiya, Jun [2 ]
Morita, Yukiko [1 ]
机构
[1] Sagamihara Natl Hosp, Natl Hosp Org, Dept Cardiol, Minami Ku, Sagamihara, Kanagawa 2520392, Japan
[2] Sagamihara Natl Hosp, Natl Hosp Org, Dept Radiol, Sagamihara, Kanagawa 2520392, Japan
关键词
CT-derived attenuation correction; IQ-SPECT; Multifocal collimator; Myocardial perfusion imaging; Short-time scan; AGREEMENT; CAMERAS;
D O I
10.1007/s12149-014-0880-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective IQ-SPECT (Siemens AG, Munich, Germany) is a highly sensitive single-photon-emission computed tomography (SPECT) myocardial perfusion imaging (MPI) system that uses a multifocal collimator. We searched for a suitable protocol for short-time imaging by IQ-SPECT in thallium-201 (Tl-201) MPI by evaluating phantom images and also by comparing human IQ-SPECT images with conventional SPECT images as reference standards. Methods We assessed the image quality using the normalized mean square error (NMSE) and drew up count profiles in Tl-201 SPECT images acquired with IQ-SPECT in a phantom study. We also performed Tl-201 stress myocardial SPECT/CT in 21 patients and compared delayed images acquired by using IQ-SPECT with 36 or 17 views per head with images obtained by using conventional SPECT. Results The NMSE of SPECT images from IQ-SPECT with 36 views was approximately one-fifth of that with 17 views. The myocardial count profile of images with 17 views was lower than those of images with 36 or 104 views in some regions. Defect scores were significantly lower, and image quality scores higher, in images from conventional SPECT than in those from IQ-SPECT with 17 views. Defect scores and image quality scores were equivalent in images from conventional SPECT and those from IQ-SPECT with 36 views. Agreement with the results of conventional SPECT in terms of coronary artery territory-based defect judgment was the best in IQ-SPECT with 36 views with computed tomography-derived attenuation correction (CTAC): the kappa values for IQ-SPECT with 36 views were 0.76 (without CTAC) and 0.83 (with CTAC), and those for IQ-SPECT with 17 views were 0.62 (without CTAC) and 0.59 (with CTAC). The difference in quantitative tracer uptake between conventional SPECT images and IQ-SPECT images was significantly greater for IQ-SPECT images with 17 views than for those with 36 views. Conclusions Scanning with 36 views per head with CTAC may be appropriate for Tl-201 MPI using IQSPECT, because it provides images equivalent to those using conventional SPECT.
引用
收藏
页码:707 / 715
页数:9
相关论文
共 16 条
[1]   TIME COURSE OF TL-201 REDISTRIBUTION AFTER TRANSIENT MYOCARDIAL ISCHEMIA [J].
BELLER, GA ;
WATSON, DD ;
ACKELL, P ;
POHOST, GM .
CIRCULATION, 1980, 61 (04) :791-797
[2]   SPECT/CT [J].
Buck, Andreas K. ;
Nekolla, Stephan ;
Ziegler, Sibylle ;
Beer, Ambros ;
Krause, Bernd J. ;
Herrmann, Ken ;
Scheidhauer, Klemens ;
Wester, Hans-Juergen ;
Rummeny, Ernst J. ;
Schwaiger, Markus ;
Drzezga, Alexander .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (08) :1305-1319
[3]   Evaluation of patients with coronary artery disease IQ-SPECT protocol in myocardial perfusion imaging: Preliminary results [J].
Caobelli, F. ;
Pizzocaro, C. ;
Paghera, B. ;
Guerra, U. P. .
NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2013, 52 (05) :178-185
[4]   Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart - A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association [J].
Cerqueira, MD ;
Weissman, NJ ;
Dilsizian, V ;
Jacobs, AK ;
Kaul, S ;
Laskey, WK ;
Pennell, DJ ;
Rumberger, JA ;
Ryan, T ;
Verani, MS .
CIRCULATION, 2002, 105 (04) :539-542
[6]   Cardiac Dedicated Ultrafast SPECT Cameras: New Designs and Clinical Implications [J].
Garcia, Ernest V. ;
Faber, Tracy L. ;
Esteves, Fabio P. .
JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (02) :210-217
[7]   Compared Performance of High-Sensitivity Cameras Dedicated to Myocardial Perfusion SPECT: A Comprehensive Analysis of Phantom and Human Images [J].
Imbert, Laetitia ;
Poussier, Sylvain ;
Franken, Philippe R. ;
Songy, Bernard ;
Verger, Antoine ;
Morel, Olivier ;
Wolf, Didier ;
Noel, Alain ;
Karcher, Gilles ;
Marie, Pierre-Yves .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (12) :1897-1903
[8]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174
[9]  
Nakata T, 2008, SHINZOUKAKUIGAKU, V9, P6
[10]   CONSTRAINED LEAST-SQUARES RESTORATION OF NUCLEAR-MEDICINE IMAGES - SELECTING THE COARSENESS FUNCTION [J].
PENNEY, BC ;
KING, MA ;
SCHWINGER, RB ;
BAKER, SP ;
STRITZKE, P ;
DOHERTY, PW .
MEDICAL PHYSICS, 1987, 14 (05) :849-858