MR-guided transcranial brain HIFU in small animal models

被引:80
作者
Larrat, B. [1 ]
Pernot, M. [1 ]
Aubry, J-F [1 ]
Dervishi, E. [2 ]
Sinkus, R. [1 ]
Seilhean, D. [3 ]
Marie, Y. [4 ]
Boch, A-L [2 ]
Fink, M. [1 ]
Tanter, M. [1 ]
机构
[1] Univ Paris 07, Inst Langevin, ESPCI ParisTech, Lab Ondes & Acoust,CNRS,INSERM,UMR 7587,U979, F-75231 Paris, France
[2] Hop La Pitie Salpetriere, INSERM, U495, F-75651 Paris 13, France
[3] Hop La Pitie Salpetriere, Dept Neuropathol, F-75651 Paris 13, France
[4] Hop La Pitie Salpetriere, Dept Neurosurg, F-75651 Paris 13, France
关键词
INTENSITY FOCUSED ULTRASOUND; MAGNETIC-RESONANCE-ELASTOGRAPHY; ACOUSTIC RADIATION FORCE; TIME-REVERSAL MIRROR; IN-VIVO; DIAGNOSTIC ULTRASOUND; MAMMALIAN-TISSUES; THERMAL ABLATION; TEMPERATURE; THERAPY;
D O I
10.1088/0031-9155/55/2/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recent studies have demonstrated the feasibility of transcranial high-intensity focused ultrasound (HIFU) therapy in the brain using adaptive focusing techniques. However, the complexity of the procedures imposes provision of accurate targeting, monitoring and control of this emerging therapeutic modality in order to ensure the safety of the treatment and avoid potential damaging effects of ultrasound on healthy tissues. For these purposes, a complete workflow and setup for HIFU treatment under magnetic resonance (MR) guidance is proposed and implemented in rats. For the first time, tissue displacements induced by the acoustic radiation force are detected in vivo in brain tissues and measured quantitatively using motion-sensitive MR sequences. Such a valuable target control prior to treatment assesses the quality of the focusing pattern in situ and enables us to estimate the acoustic intensity at focus. This MR-acoustic radiation force imaging is then correlated with conventional MR-thermometry sequences which are used to follow the temperature changes during the HIFU therapeutic session. Last, pre- and post-treatment magnetic resonance elastography (MRE) datasets are acquired and evaluated as a new potential way to non-invasively control the stiffness changes due to the presence of thermal necrosis. As a proof of concept, MR-guided HIFU is performed in vitro in turkey breast samples and in vivo in transcranial rat brain experiments. The experiments are conducted using a dedicated MR-compatible HIFU setup in a high-field MRI scanner (7 T). Results obtained on rats confirmed that both the MR localization of the US focal point and the pre- and post-HIFU measurement of the tissue stiffness, together with temperature control during HIFU are feasible and valuable techniques for efficient monitoring of HIFU in the brain. Brain elasticity appears to be more sensitive to the presence of oedema than to tissue necrosis.
引用
收藏
页码:365 / 388
页数:24
相关论文
共 80 条
  • [71] Compensating for bone interfaces and respiratory motion in high-intensity focused ultrasound
    Tanter, M.
    Pernot, M.
    Aubry, J.-F.
    Montaldo, G.
    Marquet, F.
    Fink, M.
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2007, 23 (02) : 141 - 151
  • [72] Focusing and steering through absorbing and aberrating layers: Application to ultrasonic propagation through the skull
    Tanter, M
    Thomas, JL
    Fink, M
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 103 (05) : 2403 - 2410
  • [73] Quantitative assessment of breast lesion viscoelasticity: Initial clinical results using supersonic shear imaging
    Tanter, Mickael
    Bercoff, Jeremy
    Athanasiou, Alexandra
    Deffieux, Thomas
    Gennisson, Jean-Luc
    Montaldo, Gabriel
    Muller, Marie
    Tardivon, Anne
    Fink, Mathias
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2008, 34 (09) : 1373 - 1386
  • [74] High intensity focused ultrasound: Physical principles and devices
    Ter Haar, Gail
    Coussios, Constantin
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2007, 23 (02) : 89 - 104
  • [75] Therapeutic applications of ultrasound
    ter Haar, Gail
    [J]. PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 2007, 93 (1-3) : 111 - 129
  • [76] Ultrasonic beam focusing through tissue inhomogeneities with a time reversal mirror: Application to transskull therapy
    Thomas, JL
    Fink, MA
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (06) : 1122 - 1129
  • [77] Image-guided acoustic therapy
    Vaezy, S
    Andrew, M
    Kaczkowski, P
    Crum, L
    [J]. ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2001, 3 : 375 - 390
  • [78] MRI detection of the thermal effects of focused ultrasound on the brain
    Vykhodtseva, N
    Sorrento, V
    Jolesz, FA
    Bronson, RT
    Hynynen, K
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (05) : 871 - 880
  • [79] Wu T, 2001, MAGN RESON MED, V45, P80, DOI 10.1002/1522-2594(200101)45:1<80::AID-MRM1012>3.0.CO
  • [80] 2-Y