Investigation on Blind Tip Reconstruction Errors Caused by Sample Features

被引:10
作者
Wan, Jiahuan [1 ]
Xu, Linyan [1 ]
Wu, Sen [1 ]
Hu, Xiaodong [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
基金
中国国家自然科学基金;
关键词
blind tip reconstruction; 3D tip shape; AFM; tip characterizer; ATOMIC-FORCE MICROSCOPE; SCANNING PROBE MICROSCOPY; SURFACE RECONSTRUCTION; ELECTRON-MICROSCOPE; IMAGE SIMULATION; SHAPE; CALIBRATION; CONTACT;
D O I
10.3390/s141223159
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Precision measurements of a nanoscale sample surface using an atomic force microscope (AFM) require a precise quantitative knowledge of the 3D tip shape. Blind tip reconstruction (BTR), established by Villarrubia, gives an outer bound with larger errors if the tip characterizer is not appropriate. In order to explore the errors of BTR, a series of simulation experiments based on a conical model were carried out. The results show that, to reconstruct the tip precisely, the cone angle of the tip characterizer must be smaller than that of the tip. Furthermore, the errors decrease as a function of the tip cone angle and increase linearly with the sample radius of curvature, irrespective of the tip radius of curvature. In particular, for sharp (20 nm radius) and blunt (80 nm radius) tips, the radius of curvature of the tip characterizer must be smaller than 5 nm. Based on these simulation results, a local error model of BTR was established. The maximum deviation between the errors derived from the model and the simulated experiments is 1.22 nm. Compared with the lateral resolution used in the above simulated experiments (4 nm/pixel), it is valid to ignore the deviations and consider the local error model of BTR is indeed in quantitative agreement with the simulation results. Finally, two simulated ideal structures are proposed here, together with their corresponding real samples. The simulation results show they are suitable for BTR.
引用
收藏
页码:23159 / 23175
页数:17
相关论文
共 38 条
[1]   Investigation of blind tip estimation [J].
Bakucz, P. ;
Krueger-Sehm, R. ;
Koenders, L. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (07)
[2]  
Bukharaev A.A., 1998, SCANNING MICROSCOPY, V12, P225
[3]   Force measurements with the atomic force microscope: Technique, interpretation and applications [J].
Butt, HJ ;
Cappella, B ;
Kappl, M .
SURFACE SCIENCE REPORTS, 2005, 59 (1-6) :1-152
[4]   AFM nanoindentation: tip shape and tip radius of curvature effect on the hardness measurement [J].
Calabri, L. ;
Pugno, N. ;
Menozzi, C. ;
Valeri, S. .
JOURNAL OF PHYSICS-CONDENSED MATTER, 2008, 20 (47)
[5]   Micromechanical properties of elastic polymeric materials as probed by scanning force microscopy [J].
Chizhik, SA ;
Huang, Z ;
Gorbunov, VV ;
Myshkin, NK ;
Tsukruk, VV .
LANGMUIR, 1998, 14 (10) :2606-2609
[6]   Self-assembled polystyrene nanospheres for the evaluation of atomic force microscopy tip curvature radius [J].
Colombi, P. ;
Alessandri, I. ;
Bergese, P. ;
Federici, S. ;
Depero, L. E. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2009, 20 (08)
[7]   Examination of atomic (scanning) force microscopy probe tips with the transmission electron microscope [J].
DeRose, JA ;
Revel, JP .
MICROSCOPY AND MICROANALYSIS, 1997, 3 (03) :203-213
[8]   Experimental test of blind tip reconstruction for scanning probe microscopy [J].
Dongmo, LS ;
Villarrubia, JS ;
Jones, SN ;
Renegar, TB ;
Postek, M ;
Song, JF .
ULTRAMICROSCOPY, 2000, 85 (03) :141-153
[9]   Towards easy and reliable AFM tip shape determination using blind tip reconstruction [J].
Flater, Erin E. ;
Zacharakis-Jutz, George E. ;
Dumba, Braulio G. ;
White, Isaac A. ;
Clifford, Charles A. .
ULTRAMICROSCOPY, 2014, 146 :130-143
[10]   Implications of the idea of effective tip shape on nanoindentation unloading curves: AFM measurements and FE simulation [J].
Gadelrab, Karim R. ;
Chiesa, Matteo ;
Bonilla, Flavio A. .
JOURNAL OF MATERIALS RESEARCH, 2012, 27 (01) :126-131