Mechanical properties of patellar tendon allografts subjected to chemical sterilization

被引:42
作者
Jones, David B.
Huddleston, Paul M.
Zobitz, Mark E.
Stuart, Michael J.
机构
[1] Mayo Clin, Dept Orthoped Surg, Mayo Med Sch, Coll Med, Rochester, MN 55905 USA
[2] Mayo Clin, Div Orthoped Res, Orthoped Biomech Lab, Rochester, MN 55905 USA
[3] Mayo Clin, Dept Orthoped Surg, Rochester, MN 55905 USA
关键词
allograft; anterior cruciate ligament; biomechanical; reconstruction; sterilization;
D O I
10.1016/j.arthro.2006.11.031
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: We tested the hypothesis that the preimplantation mechanical properties of BioCleanse-treated bone-patellar tendon-bone (BPTB) allografts are not significantly different from those of untreated specimens. Methods: For this controlled laboratory study, specimens were harvested as central third or hemi-BPTB units from both knees of 17 cadaveric tissue donors (I I men and 6 women) aged 19 to 88 years. Donor-matched specimens (20 per group) were randomly assigned to either BioCleanse-treated or untreated control groups. Specimens were subjected to 25 minutes of pretensioning at 89 N and then cyclically loaded under longitudinal tension between 50 N and 250 N for 1,000 cycles at 1 Hz, followed by ramp to failure at 50 mm/min. Results: No statistically significant difference was found between untreated and BioCleanse-treated specimens in stiffness (235.3 +/- 37.6 N/mm v 222.3 +/- 53.4 N/mm, P =.37), cyclic creep (0.38 +/- 0.42 mm v 0.40 +/- 0.26 rum, P = .81), maximum force (1,685.7 +/- 471.6 N v 1,807.0 +/- 657.8 N, P = .47), or ultimate stress (29.0 +/- 9.8 MPa v 29.0 +/- 12.8 MPa, P = .98). Conclusions: The preimplantation mechanical properties of BPTB allografts treated with BioCleanse are not significantly different from those of untreated controls. Clinical Relevance: This laboratory study compares the biomechanical properties of chemically treated allografts, which are currently being used in anterior cruciate ligament reconstruction, with those of nontreated fresh-frozen allografts.
引用
收藏
页码:400 / 404
页数:5
相关论文
共 33 条
[1]  
Archibald LK, 2002, JAMA-J AM MED ASSOC, V287, P1642
[2]   THE EFFECTS OF FREEZE-DRYING AND ETHYLENE-OXIDE STERILIZATION ON THE MECHANICAL-PROPERTIES OF HUMAN PATELLAR TENDON [J].
BECHTOLD, JE ;
EASTLUND, DT ;
BUTTS, MK ;
LAGERBORG, DF ;
KYLE, RF .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1994, 22 (04) :562-566
[3]   Anterior cruciate ligament reconstruction: Allograft versus autograft [J].
Chang, SKY ;
Egami, DK ;
Shaieb, MD ;
Kan, DM ;
Richardson, AB .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2003, 19 (05) :453-462
[4]   Cost comparison of anterior cruciate ligament reconstruction: Autograft versus allograft [J].
Cole, DW ;
Ginn, TA ;
Chen, GJ ;
Smith, BP ;
Curl, WW ;
Martin, DF ;
Poehling, GG .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2005, 21 (07) :786-790
[5]   The biomechanical effects of low-dose irradiation on bone-patellar tendon-bone allografts [J].
Curran, AR ;
Adams, DJ ;
Gill, JL ;
Steiner, ME ;
Scheller, AD .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2004, 32 (05) :1131-1135
[6]   EFFECTS OF GAMMA-IRRADIATION ON THE HUMAN-IMMUNODEFICIENCY-VIRUS - A STUDY IN FROZEN HUMAN BONE-PATELLAR LIGAMENT-BONE GRAFTS OBTAINED FROM INFECTED CADAVERA [J].
FIDELER, BM ;
VANGSNESS, CT ;
MOORE, T ;
LI, ZL ;
RASHEED, S .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (07) :1032-1035
[7]   GAMMA-IRRADIATION - EFFECTS ON BIOMECHANICAL PROPERTIES OF HUMAN BONE-PATELLAR TENDON-BONE ALLOGRAFTS [J].
FIDELER, BM ;
VANGSNESS, T ;
LU, B ;
ORLANDO, C ;
MOORE, T .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1995, 23 (05) :643-646
[8]   BIOMECHANICAL ANALYSIS OF PATELLAR TENDON ALLOGRAFTS AS A FUNCTION OF DONOR AGE [J].
FLAHIFF, CM ;
BROOKS, AT ;
HOLLIS, M ;
VANDERSCHILDEN, JL ;
NICHOLAS, RW .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1995, 23 (03) :354-358
[9]   EFFECTS OF GAMMA-IRRADIATION ON THE INITIAL MECHANICAL AND MATERIAL PROPERTIES OF GOAT BONE-PATELLAR TENDON-BONE ALLOGRAFTS [J].
GIBBONS, MJ ;
BUTLER, DL ;
GROOD, ES ;
BYLSKIAUSTROW, DI ;
LEVY, MS ;
NOYES, FR .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (02) :209-218
[10]  
Harner CD, 1996, CLIN ORTHOP RELAT R, P134