Optimization of liquid jet system for laser-induced breakdown spectroscopy analysis

被引:31
作者
Skocovska, Katarina [1 ]
Novotny, Jan [2 ]
Prochazka, David [2 ]
Porizka, Pavel [2 ]
Novotny, Karel [2 ]
Kaiser, Jozef [2 ]
机构
[1] Brno Univ Technol, Fac Mech Engn, Inst Phys Engn, CS-61090 Brno, Czech Republic
[2] Brno Univ Technol, CEITEC Cent European Inst Technol, CS-61090 Brno, Czech Republic
关键词
TRACE-ELEMENT; WATER; METALS; LIBS;
D O I
10.1063/1.4947233
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A complex optimization of geometrical and temporal parameters of a jet system (developed in Laser-induced breakdown spectroscopy (LIBS) laboratory of Brno University of Technology) for direct elemental analysis of samples in a liquid state of matter using LIBS was carried out. First, the peristaltic pump was synchronized with the flashlamp of the ablation laser, which reduced variation of the ablated sample amount. Also, the fluctuation of the laser ray angle incident on the jet surface was diminished. Such synchronization reduced signal standard deviations and thus increased repeatability of the measurements. Then, laser energy and distance of the focusing lens from the sample were optimized. The gate delay time and the gate width were optimized for single pulse (SP) experiments; the gate delay time and the inter-pulse delay were optimized for the use of double pulse (DP) variant. Results were assessed according to the highest signal to noise ratios and the lowest relative standard deviations of the signal. The sensitivity of the single pulse and the double pulse LIBS for the detection of heavy metals traces, copper (Cu. at 324.754 nm) and lead (Pb. at 405.781 nm), in aqueous solution of copper (Pi) sulfate and lead (Pi) acetate, was estimated in terms of limits of detection (LODs). As a result, sensitivity improvement of DP LIBS system was observed, the LOD of Cu obtained with DP was calculated 40% lower than LOD gained from SP technique. Published by AIP Publishing.
引用
收藏
页数:7
相关论文
共 26 条
[1]   Trace element analysis in water by the laser-induced breakdown spectroscopy technique [J].
Arca, G ;
Ciucci, A ;
Palleschi, V ;
Rastelli, S ;
Tognoni, E .
APPLIED SPECTROSCOPY, 1997, 51 (08) :1102-1105
[2]   EVALUATION OF AN ISOLATED DROPLET SAMPLE INTRODUCTION SYSTEM FOR LASER-INDUCED BREAKDOWN SPECTROSCOPY [J].
ARCHONTAKI, HA ;
CROUCH, SR .
APPLIED SPECTROSCOPY, 1988, 42 (05) :741-746
[3]   Laser-induced breakdown spectroscopy of liquids: Aqueous solutions of nickel and chlorinated hydrocarbons [J].
Berman, LM ;
Wolf, PJ .
APPLIED SPECTROSCOPY, 1998, 52 (03) :438-443
[4]   Sampling statistics and considerations for single-shot analysis using laser-induced breakdown spectroscopy [J].
Carranza, JE ;
Hahn, DW .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (04) :779-790
[5]   Panoramic laser-induced breakdown spectrometry of water [J].
Charfi, B ;
Harith, MA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (07) :1141-1153
[6]   SPECTROCHEMICAL ANALYSIS OF LIQUIDS USING THE LASER SPARK [J].
CREMERS, DA ;
RADZIEMSKI, LJ ;
LOREE, TR .
APPLIED SPECTROSCOPY, 1984, 38 (05) :721-729
[7]   Quantitative elemental determination in water and oil by laser induced breakdown spectroscopy [J].
Fichet, P ;
Mauchien, P ;
Wagner, JF ;
Moulin, C .
ANALYTICA CHIMICA ACTA, 2001, 429 (02) :269-278
[8]   Detection of heavy metals in Arabian crude oil residue using laser induced breakdown spectroscopy [J].
Gondal, M. A. ;
Hussain, T. ;
Yamani, Z. H. ;
Baig, M. A. .
TALANTA, 2006, 69 (05) :1072-1078
[9]   Laser-Induced Breakdown Spectroscopy (LIBS), Part II: Review of Instrumental and Methodological Approaches to Material Analysis and Applications to Different Fields [J].
Hahn, David W. ;
Omenetto, Nicolo .
APPLIED SPECTROSCOPY, 2012, 66 (04) :347-419
[10]   Detection of toxic metals in waste water from dairy products plant using Laser Induced Breakdown Spectroscopy [J].
Hussain, T. ;
Gondal, M. A. .
BULLETIN OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY, 2008, 80 (06) :561-565