Evaluation of new hip prosthesis design with finite element analysis

被引:15
作者
Talip, Celik [1 ]
Kisioglu, Yasin [1 ]
机构
[1] Kocaeli Univ, Fac Technol, Dept Biomed Engn, TR-41380 Kocaeli, Turkey
关键词
Hip prosthesis; Finite element analysis; New hip prosthesis; Computed tomography; STEM; CEMENT; ARTHROPLASTY; REPLACEMENT; FIXATION; FORCES;
D O I
10.1007/s13246-019-00802-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Cemented and cementless hip prostheses are used in total hip replacement surgery. Short, medium and long term success rates of these hip prostheses are controversial in the literature. Traditional cemented and cementless hip prostheses have advantages and disadvantages affecting the success of the implantation process. In this study, a new design of hip prosthesis is presented considering the advantages and disadvantages of the prostheses. Femur and prostheses were modeled and combined with each other to perform the finite element analysis (FEA). The new design of prosthesis was compared to the conventional prosthesis in terms of mechanical aspects. The evaluation criteria are the maximum von Mises stress and micro-movement of the contact between femur and prosthesis. In conclusion, the new design of prosthesis was found to provide a sufficient amount of primary stability and decreased the risk of stress shielding.
引用
收藏
页码:1033 / 1038
页数:6
相关论文
共 22 条
  • [1] Patient-specific design process and evaluation of a hip prosthesis femoral stem
    Abdelaa, Osama
    Darwish, Saied
    El-Hofy, Hassan
    Saito, Yoshio
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (06) : 271 - 290
  • [2] Cemented versus uncemented fixation in total hip replacement: a systematic review and meta-analysis of randomized controlled trials
    Abdulkarim, Ali
    Ellanti, Prasad
    Motterlini, Nicola
    Fahey, Tom
    O'Byrne, John M.
    [J]. ORTHOPEDIC REVIEWS, 2013, 5 (01) : 34 - 44
  • [3] [Anonymous], 2007, Med. Sci, DOI [10.3923/jms.2007.460.467, DOI 10.3923/JMS.2007.460.467]
  • [4] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [5] The effect of cement on hip stem fixation: a biomechanical study
    Celik, Talip
    Mutlu, Ibrahim
    Ozkan, Arif
    Kisioglu, Yasin
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (02) : 349 - 357
  • [6] Internal forces and moments in the femur during walking
    Duda, GN
    Schneider, E
    Chao, EYS
    [J]. JOURNAL OF BIOMECHANICS, 1997, 30 (09) : 933 - 941
  • [7] Determination of local micromotion at the stem-neck taper junction of a bi-modular total hip prosthesis design
    Falkenberg, Adrian
    Drummen, Paul
    Morlock, Michael M.
    Huber, Gerd
    [J]. MEDICAL ENGINEERING & PHYSICS, 2019, 65 : 31 - 38
  • [8] The biomechanical effect of anteversion and modular neck offset on stress shielding for short-stem versus conventional long-stem hip implants
    Goshulak, Peter
    Samiezadeh, Saeid
    Aziz, Mina S. R.
    Bougherara, Habiba
    Zdero, Radovan
    Schemitsch, Emil H.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2016, 38 (03) : 232 - 240
  • [9] Uncemented and cemented primary total hip arthroplasty in the Swedish Hip Arthroplasty Register Evaluation of 170,413 operations
    Hailer, Nils P.
    Garellick, Goran
    Karrholm, Johan
    [J]. ACTA ORTHOPAEDICA, 2010, 81 (01) : 34 - 41
  • [10] Three-dimensional shape optimization of a cemented hip stem and experimental validations
    Higa, Masaru
    Tanino, Hiromasa
    Nishimura, Ikuya
    Mitamura, Yoshinori
    Matsuno, Takeo
    Ito, Hiroshi
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2015, 18 (01) : 79 - 85