Mass changes and dissolution of platinum during electrical stimulation in artificial perilymph solution

被引:31
作者
Hibbert, DB [1 ]
Weitzner, K
Tabor, B
Carter, P
机构
[1] Univ New S Wales, Sch Chem, Sydney, NSW 2052, Australia
[2] Cochlear Ltd, Lane Cove, NSW 2066, Australia
基金
澳大利亚研究理事会;
关键词
auditory prosthesis; platinum electrochemistry; corrosion; electrochemical quartz crystal microbalance;
D O I
10.1016/S0142-9612(00)00146-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The electrochemistry of platinum electrodes in artificial perilymph solution subjected to cyclic and steady-state potentials was studied by a quartz crystal electrochemical microbalance and by analysis of electrolyte for dissolved platinum. The effect of amino acid adsorption appears to be one of blocking sites for platinum oxidation and oxide reduction, a process in which the amino acid competes with chloride from phosphate-buffered saline. For amino acids such as cysteine, which are known to interact strongly with platinum, the voltammograms became nearly featureless and only a small change in mass was observed during cycling of the potential. There were no mass changes of an electrode in solution containing human serum albumen, but its presence did inhibit dissolution of platinum. The overall magnitudes of dissolved platinum found in the buffered solutions were low, remaining below 5 ppb in most cases. Dissolution was greatest in solutions containing high concentrations of cysteine. Extrapolation of the results to implanted auditory prosthesis electrodes indicated that platinum dissolution would not lead to toxic concentrations in the body. (C) 2000 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:2177 / 2182
页数:6
相关论文
共 15 条
[1]   ADSORPTION OF ANIONS ON SMOOTH PLATINUM ELECTRODES [J].
BAGOTZKY, VS ;
VASSIYEV, YB ;
WEBER, J ;
PIRTSKHA.JN .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1970, 27 (01) :31-&
[2]   PLATINUM OXIDE FILM FORMATION REDUCTION - AN IN-SITU MASS MEASUREMENT STUDY [J].
BIRSS, VI ;
CHANG, M ;
SEGAL, J .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1993, 355 (1-2) :181-191
[3]  
BLACK RC, 1979, APPL NEUROPHYSIOL, V42, P366
[4]   EXPERIMENTAL ASPECTS OF USE OF THE QUARTZ CRYSTAL MICROBALANCE IN SOLUTION [J].
BRUCKENSTEIN, S ;
SHAY, M .
ELECTROCHIMICA ACTA, 1985, 30 (10) :1295-1300
[5]   ELECTRICAL-STIMULATION WITH PT ELECTRODES - TRACE ANALYSIS FOR DISSOLVED PLATINUM AND OTHER DISSOLVED ELECTROCHEMICAL PRODUCTS [J].
BRUMMER, SB ;
MCHARDY, J ;
TURNER, MJ .
BRAIN BEHAVIOR AND EVOLUTION, 1977, 14 (1-2) :10-22
[6]  
BRUMMER SB, 1979, N01NS32313, P19
[7]   ELECTROCHEMICAL APPLICATIONS OF THE QUARTZ CRYSTAL MICROBALANCE [J].
DEAKIN, MR ;
BUTTRY, DA .
ANALYTICAL CHEMISTRY, 1989, 61 (20) :A1147-+
[8]   EFFECTS OF EXPERIMENTAL ROUND WINDOW MEMBRANE LACERATION ON THE FREE AMINO-ACID PROFILE OF PERILYMPH [J].
HARUTA, A ;
TONO, T ;
MORIMITSU, T .
ACTA OTO-LARYNGOLOGICA, 1995, 115 (04) :504-508
[9]  
HIBBERT DB, 2000, UNPUB J ELECTROCHEM
[10]   INFLUENCE OF PULSED CURRENT ON PLATINIZED TITANIUM AND TANTALUM ANODE DURABILITY [J].
JUCHNIEWICZ, R ;
WALASZKOWSKI, J ;
BOHDANOWICZ, W ;
SOKOLSKI, W ;
WIDUCHOWSKI, A .
CORROSION SCIENCE, 1986, 26 (01) :55-61