Reproducibility of Retinal Nerve Fiber Layer and Macular Thickness Measurement with the RTVue-100 Optical Coherence Tomograph

被引:155
|
作者
Garas, Anita [1 ]
Vargha, Peter [2 ]
Hollo, Gabor [1 ]
机构
[1] Semmelweis Univ, Dept Ophthalmol, Budapest, Hungary
[2] Semmelweis Univ, Ctr Cardiovasc, Budapest, Hungary
关键词
STRATUS OCT; TIME-DOMAIN; SPECTRAL-DOMAIN; FOURIER-DOMAIN; GLAUCOMA; VARIABILITY; IMPROVES;
D O I
10.1016/j.ophtha.2009.08.039
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the reproducibility of peripapillary retinal nerve fiber layer thickness (RNFLT) and ganglion cell complex (GCC) measurements made with the RTVue-100 Fourier-domain optical coherence tomography (OCT) device (Optovue, Inc., Fremont, CA) and to determine the influence of 4 factors: pupil dilation, subject age, experience in imaging examinations, and glaucoma severity. Design: Prospective study for evaluation of a diagnostic test. Participants: Thirty-seven hospital-based participants (14 normal and ocular hypertensive subjects, 11 patients with moderate and 12 with severe glaucoma), all experienced in imaging examinations, and 40 consecutive screening trial participants lacking such experience. Methods: One eye of all participants was imaged 5 times. For the hospital-based participants, on the same day the measurement series was repeated after pupil dilation. For determination of intersession reproducibility, measurements were performed again 3 months later. Main Outcome Measures: Intrasession and intersession coefficient of variation (CV), intraclass correlation coefficient (ICC), intratest variability, and test-retest variability. Results: For the average and quadrant RNFLT parameters and the GCC parameters, intrasession ICC varied between 93.9% and 99.0%, intrasession CV between 1.95% and 5.69%, and intratest variability varied between 3.11 and 9.13 mu m. Most thickness values, all intrasession CV and intratest variability values, and the signal strength index were not changed significantly after pupil dilation. Intrasession CV of the 16 peripapillary RNFLT sectors varied between 4.90% and 11.66%. Most intrasession CV values increased significantly with increased disease severity, but intratest variability for average RNFLT and the quadrant RNFLT parameters, which are used for statistical comparison with the normative database, were unaffected by the severity of glaucoma. Patient age and experience in imaging examinations had no influence on intrasession CV. Intratest variability and intrasession CV represented 79.1% to 98.6% and 77.1% to 95.0% of test-retest variability and intervisit CV, respectively, for the average and quadrant RNFLT and the GCC parameters. Conclusions: Intrasession and intersession reproducibility of RNFLT and GCC measurements with the RTVue-100 OCT are satisfactory for diagnostic purposes, both in normals and in patients with different glaucoma severity. Pupil dilation, age, and experience in imaging examinations did not influence reproducibility in a clinically significant manner. Financial Disclosure(s): Proprietary or commercial disclosure may be found after the references. Ophthalmology 2010;117:738-746 (C) 2010 by the American Academy of Ophthalmology.
引用
收藏
页码:738 / 746
页数:9
相关论文
共 50 条
  • [1] Diagnostic accuracy of nerve fibre layer, macular thickness and optic disc measurements made with the RTVue-100 optical coherence tomograph to detect glaucoma
    Garas, A.
    Vargha, P.
    Hollo, G.
    EYE, 2011, 25 (01) : 57 - 65
  • [2] Diagnostic accuracy of nerve fibre layer, macular thickness and optic disc measurements made with the RTVue-100 optical coherence tomograph to detect glaucoma
    A Garas
    P Vargha
    G Holló
    Eye, 2011, 25 : 57 - 65
  • [3] Automatic, Operator-adjusted, and Manual Disc-definition for Optic Nerve Head and Retinal Nerve Fiber Layer Measurements With the RTVue-100 Optical Coherence Tomograph
    Garas, Anita
    Vargha, Peter
    Hollo, Gabor
    JOURNAL OF GLAUCOMA, 2011, 20 (02) : 80 - 86
  • [4] OPTICAL COHERENCE TOMOGRAPHY IN RETINITIS PIGMENTOSA Reproducibility and Capacity to Detect Macular and Retinal Nerve Fiber Layer Thickness Alterations
    Garcia-Martin, Elena
    Pinilla, Isabel
    Sancho, Eva
    Almarcegui, Carmen
    Dolz, Isabel
    Rodriguez-Mena, Diego
    Fuertes, Isabel
    Cuenca, Nicolas
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2012, 32 (08): : 1581 - 1591
  • [5] Reproducibility of corneal, macular and retinal nerve fiber layer thickness measurements using the iVue-100 optical coherence tomography
    Mashige, Khathutshelo P.
    AFRICAN HEALTH SCIENCES, 2017, 17 (04) : 1222 - 1228
  • [6] Evaluation of the reproducibility of retinal nerve fiber layer thickness and optic nerve head parameters with optical coherence tomography
    Rau, Ricardo
    de Freitas Silva, Andre Luiz
    Alves Salame, Andre Luiz
    Soares Melo Junior, Luiz Alberto
    REVISTA BRASILEIRA DE OFTALMOLOGIA, 2012, 71 (05) : 280 - 284
  • [7] Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography
    Wu, Huijuan
    de Boer, Johannes F.
    Chen, Teresa C.
    JOURNAL OF GLAUCOMA, 2011, 20 (08) : 470 - 476
  • [8] Comparison of Repeatability of Retinal Nerve Fiber Layer Thickness Measurement Made Using the RTVue Fourier-domain Optical Coherence Tomograph and the GDx Scanning Laser Polarimeter With Variable or Enhanced Corneal Compensation
    Garas, Anita
    Toth, Marta
    Vargha, Peter
    Hollo, Gabor
    JOURNAL OF GLAUCOMA, 2010, 19 (06) : 412 - 417
  • [9] Differences of Intrasession Reproducibility of Circumpapillary Total Retinal Thickness and Circumpapillary Retinal Nerve Fiber Layer Thickness Measurements Made with the RS-3000 Optical Coherence Tomograph
    Kita, Yoshiyuki
    Hollo, Gabor
    Kita, Ritsuko
    Horie, Daisuke
    Inoue, Makoto
    Hirakata, Akito
    PLOS ONE, 2015, 10 (12):
  • [10] Comparison of two retinal nerve fibre layer thickness measurement patterns of RTvue optical coherence tomography
    Li, Shuning
    Wang, Xiaozhen
    Wu, Gewei
    Fu, Jing
    Li, Songfeng
    Wan, Xiuhua
    Wang, Ningli
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 39 (03): : 222 - 229