Reproducibility of Retinal Nerve Fiber Layer and Macular Thickness Measurement with the RTVue-100 Optical Coherence Tomograph

被引:154
作者
Garas, Anita [1 ]
Vargha, Peter [2 ]
Hollo, Gabor [1 ]
机构
[1] Semmelweis Univ, Dept Ophthalmol, Budapest, Hungary
[2] Semmelweis Univ, Ctr Cardiovasc, Budapest, Hungary
关键词
STRATUS OCT; TIME-DOMAIN; SPECTRAL-DOMAIN; FOURIER-DOMAIN; GLAUCOMA; VARIABILITY; IMPROVES;
D O I
10.1016/j.ophtha.2009.08.039
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the reproducibility of peripapillary retinal nerve fiber layer thickness (RNFLT) and ganglion cell complex (GCC) measurements made with the RTVue-100 Fourier-domain optical coherence tomography (OCT) device (Optovue, Inc., Fremont, CA) and to determine the influence of 4 factors: pupil dilation, subject age, experience in imaging examinations, and glaucoma severity. Design: Prospective study for evaluation of a diagnostic test. Participants: Thirty-seven hospital-based participants (14 normal and ocular hypertensive subjects, 11 patients with moderate and 12 with severe glaucoma), all experienced in imaging examinations, and 40 consecutive screening trial participants lacking such experience. Methods: One eye of all participants was imaged 5 times. For the hospital-based participants, on the same day the measurement series was repeated after pupil dilation. For determination of intersession reproducibility, measurements were performed again 3 months later. Main Outcome Measures: Intrasession and intersession coefficient of variation (CV), intraclass correlation coefficient (ICC), intratest variability, and test-retest variability. Results: For the average and quadrant RNFLT parameters and the GCC parameters, intrasession ICC varied between 93.9% and 99.0%, intrasession CV between 1.95% and 5.69%, and intratest variability varied between 3.11 and 9.13 mu m. Most thickness values, all intrasession CV and intratest variability values, and the signal strength index were not changed significantly after pupil dilation. Intrasession CV of the 16 peripapillary RNFLT sectors varied between 4.90% and 11.66%. Most intrasession CV values increased significantly with increased disease severity, but intratest variability for average RNFLT and the quadrant RNFLT parameters, which are used for statistical comparison with the normative database, were unaffected by the severity of glaucoma. Patient age and experience in imaging examinations had no influence on intrasession CV. Intratest variability and intrasession CV represented 79.1% to 98.6% and 77.1% to 95.0% of test-retest variability and intervisit CV, respectively, for the average and quadrant RNFLT and the GCC parameters. Conclusions: Intrasession and intersession reproducibility of RNFLT and GCC measurements with the RTVue-100 OCT are satisfactory for diagnostic purposes, both in normals and in patients with different glaucoma severity. Pupil dilation, age, and experience in imaging examinations did not influence reproducibility in a clinically significant manner. Financial Disclosure(s): Proprietary or commercial disclosure may be found after the references. Ophthalmology 2010;117:738-746 (C) 2010 by the American Academy of Ophthalmology.
引用
收藏
页码:738 / 746
页数:9
相关论文
共 30 条
[1]   Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by optical coherence tomography [J].
Ajtony, Csilla ;
Balla, Zsolt ;
Somoskeoy, Szabolcs ;
Kovacs, Balint .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (01) :258-263
[2]   Thickness Profiles of Retinal Layers by Optical Coherence Tomography Image Segmentation [J].
Bagci, Ahmet Murat ;
Shahidi, Mahnaz ;
Ansari, Rashid ;
Blair, Michael ;
Blair, Norman Paul ;
Zelkha, Ruth .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 146 (05) :679-687
[3]   Macular symmetry testing for glaucoma detection [J].
Bagga, H ;
Greenfield, DS ;
Knighton, RW .
JOURNAL OF GLAUCOMA, 2005, 14 (05) :358-363
[4]   Inter-device Variability of the Stratus Optical Coherence Tomography [J].
Barkana, Yaniv ;
Burgansky-Eliash, Zvia ;
Gerber, Yariv ;
Melamed, Shlomo ;
Neudorfer, Meira ;
Avni, Isaac ;
Bartov, Elisha ;
Morad, Yair .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (02) :260-266
[5]   Reproducibility of retinal nerve fiber thickness measurements using the stratus OCT in normal and glaucomatous eyes [J].
Budenz, DL ;
Chang, RT ;
Huang, XR ;
Knighton, RW ;
Tielsch, JM .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (07) :2440-2443
[6]   Reproducibility of peripapillary retinal nerve fiber thickness measurements with stratus OCT in glaucomatous eyes [J].
Budenz, Donald L. ;
Fredette, Marie-Josee ;
Feuer, William J. ;
Anderson, Douglas R. .
OPHTHALMOLOGY, 2008, 115 (04) :661-666
[7]   Effects of scan circle displacement in optical coherence tomography retinal nerve fibre layer thickness measurement: a RNFL modelling study [J].
Cheung, C. Y. L. ;
Yiu, C. K. F. ;
Weinreb, R. N. ;
Lin, D. ;
Li, H. ;
Yung, A. Y. ;
Pang, C. P. ;
Lam, D. S. C. ;
Leung, C. K. S. .
EYE, 2009, 23 (06) :1436-1441
[8]   Relationship between retinal nerve fiber layer measurement and signal strength in optical coherence tomography [J].
Cheung, Carol Yim Lui ;
Leung, Christopher Kai Shun ;
Lin, Dusheung ;
Pang, Chi-Pui ;
Lam, Dennis Shun Chiu .
OPHTHALMOLOGY, 2008, 115 (08) :1347-1351
[9]   Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Detection of Band Atrophy of the Optic Nerve [J].
Costa-Cunha, Luciana V. F. ;
Cunha, Leonardo P. ;
Malta, Roberto F. S. ;
Monteiro, Mario L. R. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (01) :56-63
[10]   Reproducibility of RTVue Retinal Nerve Fiber Layer Thickness and Optic Disc Measurements and Agreement with Stratus Optical Coherence Tomography Measurements [J].
Gonzalez-Garcia, Alberto O. ;
Vizzeri, Gianmarco ;
Bowd, Christopher ;
Medeiros, Felipe A. ;
Zangwill, Linda M. ;
Weinreb, Robert N. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (06) :1067-1074