Assessing the Intraoperative Accuracy of Pedicle Screw Placement by Using a Bone-Mounted Miniature Robot System through Secondary Registration

被引:55
作者
Kuo, Keng-Liang [2 ,3 ]
Su, Yu-Feng [1 ,3 ]
Wu, Chieh-Hsin [2 ,3 ]
Tsai, Cheng-Yu [1 ,3 ]
Chang, Chih-Hui [1 ,3 ]
Lin, Chih-Lung [1 ,2 ,3 ,4 ]
Tsai, Tai-Hsin [2 ,3 ,4 ]
机构
[1] Kaohsiung Med Univ, Coll Med, Grad Inst Clin Med, Kaohsiung, Taiwan
[2] Kaohsiung Med Univ, Grad Inst Med, Coll Med, Kaohsiung, Taiwan
[3] Kaohsiung Med Univ Hosp, Dept Neurosurg, Kaohsiung, Taiwan
[4] Kaohsiung Med Univ, Dept Neurosurg, Kaohsiung, Taiwan
关键词
LUMBAR SPINE; FIXATION; GUIDANCE;
D O I
10.1371/journal.pone.0153235
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Introduction Pedicle screws are commonly employed to restore spinal stability and correct deformities. The Renaissance robotic system was developed to improve the accuracy of pedicle screw placement. Purpose In this study, we developed an intraoperative classification system for evaluating the accuracy of pedicle screw placements through secondary registration. Furthermore, we evaluated the benefits of using the Renaissance robotic system in pedicle screw placement and postoperative evaluations. Finally, we examined the factors affecting the accuracy of pedicle screw implantation. Results Through use of the Renaissance robotic system, the accuracy of Kirschner-wire (K-wire) placements deviating <3 mm from the planned trajectory was determined to be 98.74%. According to our classification system, the robot-guided pedicle screw implantation attained an accuracy of 94.00% before repositioning and 98.74% after repositioning. However, the malposition rate before repositioning was 5.99%; among these placements, 4.73% were immediately repositioned using the robot system and 1.26% were manually repositioned after a failed robot repositioning attempt. Most K-wire entry points deviated caudally and laterally. Conclusion The Renaissance robotic system offers high accuracy in pedicle screw placement. Secondary registration improves the accuracy through increasing the precision of the positioning; moreover, intraoperative evaluation enables immediate repositioning. Furthermore, the K-wire tends to deviate caudally and laterally from the entry point because of skiving, which is characteristic of robot-assisted pedicle screw placement.
引用
收藏
页数:11
相关论文
共 19 条
[1]   Pedicle screw fixation in spinal disorders: A European view [J].
Boos N. ;
Webb J.K. .
European Spine Journal, 1997, 6 (1) :2-18
[2]   Clinical Acceptance and Accuracy Assessment of Spinal Implants Guided With SpineAssist Surgical Robot Retrospective Study [J].
Devito, Dennis P. ;
Kaplan, Leon ;
Dietl, Rupert ;
Pfeiffer, Michael ;
Horne, Dale ;
Silberstein, Boris ;
Hardenbrook, Mitchell ;
Kiriyanthan, George ;
Barzilay, Yair ;
Bruskin, Alexander ;
Sackerer, Dieter ;
Alexandrovsky, Vitali ;
Stueer, Carsten ;
Burger, Ralf ;
Maeurer, Johannes ;
Gordon, Donald G. ;
Schoenmayr, Robert ;
Friedlander, Alon ;
Knoller, Nachshon ;
Schmieder, Kirsten ;
Pechlivanis, Ioannis ;
Kim, In-Se ;
Meyer, Bernhard ;
Shoham, Moshe .
SPINE, 2010, 35 (24) :2109-2115
[3]  
DICK WT, 1988, CLIN ORTHOP RELAT R, P70
[4]   The use of pedicle screw fixation to improve correction in the lumbar spine of patients with idiopathic scoliosis - Is it warranted? [J].
Hamill, CL ;
Lenke, LG ;
Bridwell, KH ;
Chapman, MP ;
Blanke, K ;
Baldus, C .
SPINE, 1996, 21 (10) :1241-1249
[5]   Robotic-assisted pedicle screw placement: lessons learned from the first 102 patients [J].
Hu, Xiaobang ;
Ohnmeiss, Donna D. ;
Lieberman, Isador H. .
EUROPEAN SPINE JOURNAL, 2013, 22 (03) :661-666
[6]   Perioperative course and accuracy of screw positioning in conventional, open robotic-guided and percutaneous robotic-guided, pedicle screw placement [J].
Kantelhardt, Sven Rainer ;
Martinez, Ramon ;
Baerwinkel, Stefan ;
Burger, Ralf ;
Giese, Alf ;
Rohde, Veit .
EUROPEAN SPINE JOURNAL, 2011, 20 (06) :860-868
[7]   Free hand pedicle screw placement in the thoracic spine: Is it safe? [J].
Kim, YJ ;
Lenke, LG ;
Bridwell, KH ;
Cho, YSS ;
Riew, KD .
SPINE, 2004, 29 (03) :333-342
[8]   Pedicle screw placement accuracy - A meta-analysis [J].
Kosmopoulos, Victor ;
Schizas, Constantin .
SPINE, 2007, 32 (03) :E111-E120
[9]   The comparison of pedicle screw and cortical screw in posterior lumbar interbody fusion: a prospective randomized noninferiority trial [J].
Lee, Gun Woo ;
Son, Jung-Hwan ;
Ahn, Myun-Whan ;
Kim, Ho-Joong ;
Yeom, Jin S. .
SPINE JOURNAL, 2015, 15 (07) :1519-1526
[10]   Bone-mounted miniature robotic guidance for pedicle screw and translaminar facet screw placement: Part I - Technical development and a test case result [J].
Lieberman, Isador H. ;
Togawa, Daisuke ;
Kayanja, Mark M. ;
Reinhardt, Mary K. ;
Friedlander, Alon ;
Knoller, Nachshon ;
Benzel, Edward C. .
NEUROSURGERY, 2006, 59 (03) :641-650