Comparison of the morphological and biomechanical characteristics of keratoconus, forme fruste keratoconus, and normal corneas

被引:20
|
作者
Guo, Li-Li [1 ]
Tian, Lei [1 ,2 ,3 ]
Cao, Kai [1 ]
Li, Yu-Xin [1 ]
Li, Na [1 ]
Yang, Wei-Qiang [1 ]
Jie, Ying [1 ]
机构
[1] Capital Med Univ, Beijing Ophthalmol & Visual Sci Key Lab, Beijing Tongren Hosp, Beijing Inst Ophthalmol,Beijing Tongren Eye Ctr, Beijing, Peoples R China
[2] Beihang Univ, Beijing Adv Innovat Ctr Big Data Based Precis Med, Beijing, Peoples R China
[3] Capital Med Univ, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
keratoconus; forme fruste keratoconus; corneal morphology; corneal biomechanics; Corvis ST;
D O I
10.1080/08820538.2021.1896752
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To explore the feasibility of corneal morphological and biomechanical parameters for keratoconus and forme fruste keratoconus diagnosis. Methods: This case-control study included a total of 517 eyes from 408 keratoconus patients (KC group), 83 eyes from 83 forme fruste keratoconus patients (FFKC group), and 158 eyes from 158 patients with normal corneas (NL group). All subjects underwent routine ophthalmologic examinations. Pentacam and Corneal Visualization Scheimpflug Technology (Corvis ST) were used to obtain corneal morphological and biomechanical parameters. Differences between groups were compared using receiver operating characteristic (ROC) curve analysis. Results: ROC analysis showed that all corneal morphological parameters and most biomechanical parameters distinguished KC from NL, with an area under the curve (AUC) greater than 0.80, of which Belin-Ambrosio enhanced ectasia total deviation index (BAD-D) and tomographic and biomechanical index (TBI) were most efficient. The AUC for distinguishing KC from NL of the BAD-D was 0.989 and the TBI was 0.993, which were not statistically significant (DeLong et al., P= .232). The BAD-D cut-off point of 1.595 provided 95.9% sensitivity for distinguishing KC from NL with 100% specificity. The TBI cut-off point of 0.515 provided 96.7% sensitivity for distinguishing KC from NL with 100% specificity. The ability of other parameters to distinguish KC from NL was lower than that of BAD and TBI. Except for central astigmatism from the anterior corneal surface (AstigF), the AUC that distinguished FFKC from NL was 0.862. The AstigF cut-off point of 4.65 provided 73.5% sensitivity for distinguishing FFKC from NL with 99.3% specificity. Other parameters distinguished FFKC from NL with low efficiency. Among them, the AUC for distinguishing FFKC from NL of the TBI was 0.722, whose cut-off point of 0.273 provided 55.4% sensitivity for distinguishing KC from NL with 79.7% specificity. Conclusion: BAD-D and TBI have the highest efficiency, sensitivity, and specificity for distinguishing KC from NL. Except for AstigF, other corneal morphological and biomechanical parameters have a relatively low ability to distinguish FFKC from NL.
引用
收藏
页码:671 / 678
页数:8
相关论文
共 50 条
  • [1] Comprehensive evaluation of corneas from normal, forme fruste keratoconus and clinical keratoconus patients using morphological and biomechanical properties
    Zhang, Hui
    Tian, Lei
    Guo, Lili
    Qin, Xiao
    Zhang, Di
    Li, Lin
    Jie, Ying
    Zhang, Haixia
    INTERNATIONAL OPHTHALMOLOGY, 2021, 41 (04) : 1247 - 1259
  • [2] Comprehensive evaluation of corneas from normal, forme fruste keratoconus and clinical keratoconus patients using morphological and biomechanical properties
    Hui Zhang
    Lei Tian
    Lili Guo
    Xiao Qin
    Di Zhang
    Lin Li
    Ying Jie
    Haixia Zhang
    International Ophthalmology, 2021, 41 : 1247 - 1259
  • [3] Biomechanical properties analysis of forme fruste keratoconus and subclinical keratoconus
    Yan Huo
    Xuan Chen
    Huazheng Cao
    Jing Li
    Jie Hou
    Yan Wang
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2023, 261 : 1311 - 1320
  • [4] Biomechanical properties analysis of forme fruste keratoconus and subclinical keratoconus
    Huo, Yan
    Chen, Xuan
    Cao, Huazheng
    Li, Jing
    Hou, Jie
    Wang, Yan
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2023, 261 (05) : 1311 - 1320
  • [5] Topographic and Tomographic Properties of Forme Fruste Keratoconus Corneas
    Saad, Alain
    Gatinel, Damien
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (11) : 5546 - 5555
  • [6] Comparisons of corneal biomechanical and tomographic parameters among thin normal cornea, forme fruste keratoconus, and mild keratoconus
    Lei Tian
    Di Zhang
    Lili Guo
    Xiao Qin
    Hui Zhang
    Haixia Zhang
    Ying Jie
    Lin Li
    Eye and Vision, 8
  • [7] SENSITIVITY AND SPECIFICITY OF NOVEL BIOMECHANICAL WAVEFORM PARAMETERS IN DETECTING FORME FRUSTE KERATOCONUS AND KERATOCONUS
    Viswanathan, Deepa
    Males, John
    Kumar, Nikhil
    Graham, Stuart
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 42 : 87 - 87
  • [8] Comparisons of corneal biomechanical and tomographic parameters among thin normal cornea, forme fruste keratoconus, and mild keratoconus
    Tian, Lei
    Zhang, Di
    Guo, Lili
    Qin, Xiao
    Zhang, Hui
    Zhang, Haixia
    Jie, Ying
    Li, Lin
    EYE AND VISION, 2021, 8 (01)
  • [9] A Systematic Review of Subclinical Keratoconus and Forme Fruste Keratoconus
    Henriquez, Maria A.
    Hadid, Marta
    Izquierdo, Luis, Jr.
    JOURNAL OF REFRACTIVE SURGERY, 2020, 36 (04) : 270 - +
  • [10] Corneal Biomechanical Properties in Normal, Forme Fruste Keratoconus, and Manifest Keratoconus After Statistical Correction for Potentially Confounding Factors
    Johnson, R. Duncan
    Nguyen, Myhanh T.
    Lee, Nancy
    Hamilton, D. Rex
    CORNEA, 2011, 30 (05) : 516 - 523