Characterization of a fiber-optic-coupled radioluminescent detector for application in the mammography energy range

被引:36
作者
Benevides, Luis A.
Huston, Alan L.
Justus, Brian L.
Falkenstein, Paul
Brateman, Libby F.
Hintenlang, David E.
机构
[1] Naval Dosimetry Ctr, Bethesda, MD 20889 USA
[2] USN, Res Lab, Washington, DC 20375 USA
[3] Univ Florida, Nucl Sci Ctr 202, Gainesville, FL 32611 USA
关键词
D O I
10.1118/1.2736788
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Fiber-optic-coupled radioluminescent (FOC) dosimeters are members of a new family of dosimeters that are finding increased clinical applications. This study provides the first characterization of a Cu doped quartz FOC dosimeter at diagnostic energies, specifically across the range of x-ray energies and intensities used in mammographies. We characterize the calibration factors, linearity, angular dependence, and reproducibility of the FOC dosimeters. The sensitive element of each dosimeter was coupled to a photon counting photomultiplier module via I m long optical fibers. A computer controlled interface permitted real-time monitoring of the dosimeter output and rapid data acquisition. The axial-angular responses for all dosimeter models show nearly uniform response without any marked decrease in sensitivity. However, the normal-to-axial angular response showed a marked decrease in sensitivity of about 0 degrees C and 180 degrees C. In most clinical applications, appropriate dosimeter positioning can minimize the contributions of the varying normal-to-axial response. The FOC dosimeters having the greatest sensitive length provided the greatest sensitivity, with greatest to lowest sensitivity observed for 4.0, 1.9, 1.6, and 1.1 mm length sensitive elements. The average sensitivity of the dosimeters varies linearly with sensitive volume (R-2 = 95%) and as a function of tube potential and target/filter combinations, generally exhibiting an increased sensitivity for higher energies. The dosimeter sensitivity as a function of tube potential had an average increase of 4.72 +/- 2.04% for dosimeter models and three target-filter combinations tested (Mo/Mo, Mo/Rh, and Rh/Rh) over a range of 25-31 kVp. All dosimeter models exhibited a linear response (R-2 >= 0.997) to exposure for all target-filter combinations, tube potentials, and tube current-time product stations evaluated and demonstrated reproducibility within 2%. All of the dosimeters examined in this study provided a response adequate for the accurate measurement of doses in clinical mammography applications. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:2220 / 2227
页数:8
相关论文
共 14 条
  • [1] Argo William P, 2004, J Appl Clin Med Phys, V5, P112, DOI 10.1120/jacmp.2022.25314
  • [2] Real-time optical-fibre luminescence dosimetry for radiotherapy:: physical characteristics and applications in photon beams
    Aznar, MC
    Andersen, CE
    Botter-Jensen, L
    Bäck, SÅJ
    Mattsson, S
    Kjær-Kristoffersen, F
    Medin, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09) : 1655 - 1669
  • [3] In vivo absorbed dose measurements in mammography using a new real-time luminescence technique
    Aznar, MC
    Hemdal, B
    Medin, J
    Marckmann, CJ
    Andersen, CE
    Botter-Jensen, L
    Andersson, I
    Mattsson, S
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (928) : 328 - 334
  • [4] Characterization of metal oxide semiconductor field effect transistor dosimeters for application in clinical mammography
    Benevides, LA
    Hintenlang, DE
    [J]. MEDICAL PHYSICS, 2006, 33 (02) : 514 - 520
  • [5] BENEVIDES LA, 2005, THESIS U FLORIDA GAI
  • [6] The characterization of a commercial MOSFET dosimeter system for use in diagnostic x ray
    Bower, MW
    Hintenlang, DE
    [J]. HEALTH PHYSICS, 1998, 75 (02): : 197 - 204
  • [7] Estimation of mean-glandular dose from monitoring breast entrance skin air kerma using a high sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeter system in mammography
    Dong, SL
    Chu, TC
    Lee, JS
    Lan, GY
    Wu, TH
    Yeh, YH
    Hwang, JJ
    [J]. APPLIED RADIATION AND ISOTOPES, 2002, 57 (06) : 791 - 799
  • [8] Huston AL, 2002, RADIAT PROT DOSIM, V101, P23, DOI 10.1093/oxfordjournals.rpd.a005974
  • [9] Remote optical fiber dosimetry
    Huston, AL
    Justus, BL
    Falkenstein, PL
    Miller, RW
    Ning, H
    Altemus, R
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2001, 184 (1-2) : 55 - 67
  • [10] Gated fiber-optic-coupled detector for in vivo real-time radiation dosimetry
    Justus, BL
    Falkenstein, P
    Huston, AL
    Plazas, MC
    Ning, H
    Miller, RW
    [J]. APPLIED OPTICS, 2004, 43 (08) : 1663 - 1668