OPTIMIZED HYBRID MEGAVOLTAGE-KILOVOLTAGE IMAGING PROTOCOL FOR VOLUMETRIC PROSTATE ARC THERAPY

被引:16
作者
Liu, Wu [1 ]
Wiersma, Rodney D. [1 ]
Xing, Lei [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Sch Med, Stanford, CA 94305 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2010年 / 78卷 / 02期
关键词
Intrafraction prostate motion; onboard imaging; volumetric arc therapy; IMPLANTED FIDUCIAL MARKERS; INTRAFRACTION MOTION; GUIDED RADIOTHERAPY; TRACKING; KV; LOCALIZATION; IMRT; MV; SYSTEM; GLAND;
D O I
10.1016/j.ijrobp.2009.11.056
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To develop a real-time prostate position monitoring technique for modern arc radiotherapy through novel use of cine-megavoltage (MV) imaging, together with as-needed kilovoltage (kV) imaging. Methods and Materials: We divided the task of monitoring the intrafraction prostate motion into two steps for rotational deliveries: to detect potential target motion beyond a predefined threshold using MV images from different viewing angles by taking advantage of gantry rotation during arc therapy and to verify the displacement and determine whether intervention is needed using fiducial/tumor position information acquired from combined MV-kV imaging (by turning on the kV imager). A Varian Trilogy linear accelerator with an onboard kV imager was used to examine selected typical trajectories using a four-dimensional motion phantom. The performance of the algorithm was evaluated using phantom measurements and computer simulation for 536 Calypso-measured tracks from 17 patients. Results: Fiducial displacement relative to the MV beam was limited to within a range of 3 mm 99.9% of the time with <1 mm accuracy. On average, only similar to 0.5 intervention per arc delivery was needed to achieve this level of accuracy. Compared with other fluoroscopy-based tracking techniques, kV use was significantly reduced to an average of <15 times per arc delivery. Conclusion: By focusing the attention on detecting predefined abnormal motion (i.e., "failure" detection) and using the inherent mechanism of gantry rotation during arc radiotherapy, the current approach provides high confidence regarding the prostate position in real time without the unwanted overhead of continuous or periodic kV imaging. (C) 2010 Elsevier Inc.
引用
收藏
页码:595 / 604
页数:10
相关论文
共 16 条
  • [1] Optimizing monoscopic kV fluoro acquisition for prostate intrafraction motion evaluation
    Adamson, Justus
    Wu, Qiuwen
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (01) : 117 - 133
  • [2] Accuracy of a wireless localization system for radiotherapy
    Balter, JM
    Wright, JN
    Newell, LJ
    Friemel, B
    Dimmer, S
    Cheng, Y
    Wong, J
    Vertatschitsch, E
    Mate, TP
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (03): : 933 - 937
  • [3] Single-Arc IMRT?
    Bortfeld, Thomas
    Webb, Steve
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (01) : N9 - N20
  • [4] Three-dimensional intrafractional movement of prostate measured during real-time tumor-tracking radiotherapy in supine and prone treatment positions
    Kitamura, K
    Shirato, H
    Seppenwoolde, Y
    Onimaru, R
    Oda, M
    Fujita, K
    Shimizu, S
    Shinohara, N
    Harabayashi, T
    Miyasaka, K
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1117 - 1123
  • [5] Intrafraction motion of the prostate during external-beam radiation therapy: Analysis of 427 patients with implanted fiducial markers
    Kotte, Alexis N. T. J.
    Hofman, Pieter
    Lagendijk, Jan J. W.
    van Vulpen, Marco
    van der Heide, Uulke A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (02): : 419 - 425
  • [6] Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy
    Kupelian, Patrick
    Willoughby, Twyla
    Mahadevan, Arul
    Djemil, Toufik
    Weinstein, Geoffrey
    Jani, Shirish
    Enke, Charles
    Solberg, Timothy
    Flores, Nicholas
    Liu, David
    Beyer, David
    Levine, Lisa
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (04): : 1088 - 1098
  • [7] Image-guided radiotherapy for localized prostate cancer: Treating a moving target
    Kupelian, Patrick A.
    Langen, Katja M.
    Willoughby, Twyla R.
    Zeidan, Omar A.
    Meeks, Sanford L.
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2008, 18 (01) : 58 - 66
  • [8] Observations on real-time prostate gland motion using electromagnetic tracking
    Langen, Katja M.
    Willoughby, Twyla R.
    Meeks, Sanford L.
    Santhanam, Anand
    Cunningham, Alexis
    Levine, Lisa
    Kupelian, Patrick A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (04): : 1084 - 1090
  • [9] Dosimetric consequences of intrafraction prostate motion
    Li, Haisen S.
    Chetty, Indrin J.
    Enke, Charles A.
    Foster, Ryan D.
    Willoughby, Twyla R.
    Kupellian, Patrick A.
    Solberg, Timothy D.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (03): : 801 - 812
  • [10] Real-time 3D internal marker tracking during arc radiotherapy by the use of combined MV-kV imaging
    Liu, W.
    Wiersma, R. D.
    Mao, W.
    Luxton, G.
    Xing, L.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (24) : 7197 - 7213