The influence of contact ratio and its location on the primary stability of cementless total hip arthroplasty: A finite element analysis

被引:18
作者
Reimeringer, M. [1 ]
Nuno, N. [1 ]
机构
[1] Ecole Technol Super, Dept Genie Prod Automatisee, Lab Rech Imagerie & Orthopedie, 1100 Rue Notre Dame Ouest, Montreal, PQ H3C 1K3, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Total hip arthroplasty; Finite element analysis; Contact; Micromotion; Cementless stem; STEM; BONE; MODELS; FIXATION;
D O I
10.1016/j.jbiomech.2016.02.031
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Cementless hip stems are fixed to the surrounding bone by means of press-fit. To ensure a good press-fit, current surgical technique specifies an under-reaming of the bone cavity using successively larger broaches. Nevertheless, this surgical technique is inaccurate. Several studies show that the contact ratio (percentage of stem interface in contact with bone) achieved after surgery can vary between 20% and 95%. Therefore, this study aimed to investigate the influence of the contact ratio and its location on the primary stability of a cementless total hip arthroplasty using finite element analysis. A straight tapered femoral stem implanted in a composite bone was subjected to stair climbing. Micromotion of 7600 nodes at the stem-bone interface was computed for different configurations of contact ratios between 2% and 98%) along the hip stem. Considering the 15 configurations evaluated, the average micromotion ranges between 27 mu m and 54 m. The percentage of the porous interface of the stem having micromotion below 40 mu m that allows bone ingrowth range between 25-57%. The present numerical study shows that full contact (100%) between stem and bone is not necessary to obtain a good primary stability. The stem primary stability is influenced by both the contact ratio and its location. Several configurations with contact ratio lower than 100% and involving either the proximal or the cortical contact provide better primary stability than the full contact configuration. However, with contact ratio lower than 40%, the stem should be in contact with cortical bone to ensure a good primary stability. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1064 / 1070
页数:7
相关论文
共 43 条
  • [1] Abdul Kadir MR, 2007, J MEKANICAL, V23, P50
  • [2] Finite element modelling of primary hip stem stability: The effect of interference fit
    Abdul-Kadir, Mohammed Rafiq
    Hansen, Ulrich
    Klabunde, Ralf
    Lucas, Duncan
    Amis, Andrew
    [J]. JOURNAL OF BIOMECHANICS, 2008, 41 (03) : 587 - 594
  • [3] Cemented versus uncemented fixation in total hip replacement: a systematic review and meta-analysis of randomized controlled trials
    Abdulkarim, Ali
    Ellanti, Prasad
    Motterlini, Nicola
    Fahey, Tom
    O'Byrne, John M.
    [J]. ORTHOPEDIC REVIEWS, 2013, 5 (01) : 34 - 44
  • [4] Nonlinear three-dimensional finite element analysis of newly designed cementless total hip stems
    Ando, M
    Imura, S
    Omori, H
    Okumura, Y
    Bo, A
    Baba, H
    [J]. ARTIFICIAL ORGANS, 1999, 23 (04) : 339 - 346
  • [5] Comparison of an inhomogeneous orthotropic and isotropic material models used for FE analyses
    Baca, Vaclav
    Horak, Zdenek
    Mikulenka, Petr
    Dzupa, Valer
    [J]. MEDICAL ENGINEERING & PHYSICS, 2008, 30 (07) : 924 - 930
  • [6] Inter-subject variability effects on the primary stability of a short cementless femoral stem
    Bah, Mamadou T.
    Shi, Junfen
    Heller, Markus O.
    Suchier, Yanneck
    Lefebvre, Fabien
    Young, Philippe
    King, Leonard
    Dunlop, Doug G.
    Boettcher, Mick
    Draper, Edward
    Browne, Martin
    [J]. JOURNAL OF BIOMECHANICS, 2015, 48 (06) : 1032 - 1042
  • [7] Efficient computational method for assessing the effects of implant positioning in cementless total hip replacements
    Bah, Mamadou T.
    Nair, Prasanth B.
    Taylor, Mark
    Browne, Martin
    [J]. JOURNAL OF BIOMECHANICS, 2011, 44 (07) : 1417 - 1422
  • [8] Design process of cementless femoral stem using a nonlinear three dimensional finite element analysis
    Baharuddin, Mohd Yusof
    Salleh, Sh-Hussain
    Zulkifly, Ahmad Hafiz
    Lee, Muhammad Hisyam
    Noor, Alias Mohd
    Harris, Arief Ruhullah A.
    Majid, Norazman Abdul
    Abd Kader, Ab Saman
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2014, 15
  • [9] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [10] THE EFFECT OF FEMORAL STEM GEOMETRY ON INTERFACE MOTION IN UNCEMENTED POROUS-COATED TOTAL HIP PROSTHESES - COMPARISON OF STRAIGHT-STEM AND CURVED-STEM DESIGNS
    CALLAGHAN, JJ
    FULGHUM, CS
    GLISSON, RR
    STRANNE, SK
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (06) : 839 - 848