Modifying a clinical linear accelerator for delivery of ultra-high dose rate irradiation

被引:150
作者
Lempart, Michael [1 ]
Blad, Borje [1 ]
Adrian, Gabriel [2 ]
Back, Sven [1 ,3 ]
Knoos, Tommy [1 ,3 ]
Ceberg, Crister [3 ]
Petersson, Kristoffer [1 ]
机构
[1] Skane Univ Hosp, Dept Hematol Oncol & Radiat Phys, Radiat Phys, Lund, Sweden
[2] Skane Univ Hosp, Div Oncol & Pathol, Clin Sci, Lund, Sweden
[3] Lund Univ, Clin Sci, Dept Med Radiat Phys, Lund, Sweden
关键词
FLASH; Ultra-high dose rate; Linac; Irradiation; ELECTRON-BEAM DOSIMETRY; FLASH IRRADIATION; RECOMBINATION;
D O I
10.1016/j.radonc.2019.01.031
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Objectives: The purpose of this study was to modify a clinical linear accelerator, making it capable of electron beam ultra-high dose rate (FLASH) irradiation. Modifications had to be quick, reversible, and without interfering with clinical treatments. Methods: Performed modifications: (1) reduced distance with three setup positions, (2) adjusted/optimized gun current, modulator charge rate and beam steering values for a high dose rate, (3) delivery was controlled with a microcontroller on an electron pulse level, and (4) moving the primary and/or secondary scattering foils from the beam path. Results: The variation in dose for a five-pulse delivery was measured to be 1% (using a diode, 4% using film) during 10 minutes after a warm-up procedure, later increasing to 7% (11% using film). A FLASH irradiation dose rate was reached at the cross-hair foil, MLC, and wedge position, with >= 30, >= 80, and >= 300 Gy/s, respectively. Moving the scattering foils resulted in an increased output of >= 120, >= 250, and >= 1000 Gy/s, at the three positions. The beam flatness was 5% at the cross-hair position for a 20 x 20 and a 10 x 10 cm(2) area, with and without both scattering foils in the beam. The beam flatness was 10% at the wedge position for a 6 and 2.5 cm diametric area, with and without the scattering foils in the beam path. Conclusions: A clinical accelerator was modified to produce ultra-high dose rates, high enough for FLASH irradiation. Future work aims to fine-tune the dose delivery, using the on-board transmission chamber signal and adjusting the dose-per-pulse. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:40 / 45
页数:6
相关论文
共 14 条
[1]  
[Anonymous], 2018, VAR MED SYST
[2]   The effect of free-electron collection on the recombination correction to ionization measurements of pulsed radiation [J].
Boag, JW ;
Hochhauser, E ;
Balk, OA .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (05) :885-897
[3]   Ultrahigh dose-rate, "flash" irradiation minimizes the side-effects of radiotherapy [J].
Favaudon, V. ;
Fouillade, C. ;
Vozenin, M. -C. .
CANCER RADIOTHERAPIE, 2015, 19 (6-7) :526-531
[4]   Ultrahigh dose-rate FLASH irradiation increases the differential response between normal and tumor tissue in mice [J].
Favaudon, Vincent ;
Caplier, Laura ;
Monceau, Virginie ;
Pouzoulet, Frederic ;
Sayarath, Mano ;
Fouillade, Charles ;
Poupon, Marie-France ;
Brito, Isabel ;
Hupe, Philippe ;
Bourhis, Jean ;
Hall, Janet ;
Fontaine, Jean-Jacques ;
Vozenin, Marie-Catherine .
SCIENCE TRANSLATIONAL MEDICINE, 2014, 6 (245)
[5]   High dose-per-pulse electron beam dosimetry: Commissioning of the Oriatron eRT6 prototype linear accelerator for preclinical use [J].
Jaccard, Maud ;
Duran, Maria Teresa ;
Petersson, Kristoffer ;
Germond, Jean-Francois ;
Liger, Philippe ;
Vozenin, Marie-Catherine ;
Bourhis, Jean ;
Bochud, Francois ;
Bailat, Claude .
MEDICAL PHYSICS, 2018, 45 (02) :863-874
[6]   High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films [J].
Jaccard, Maud ;
Petersson, Kristoffer ;
Buchillier, Thierry ;
Germond, Jean-Francois ;
Duran, Maria Teresa ;
Vozenin, Marie-Catherine ;
Bourhis, Jean ;
Bochud, Francois O. ;
Bailat, Claude .
MEDICAL PHYSICS, 2017, 44 (02) :725-735
[7]   Optimization of an on-board imaging system for extremely rapid radiation therapy [J].
Kemmerling, Erica M. Cherry ;
Wu, Meng ;
Yang, He ;
Maxim, Peter G. ;
Loo, Billy W., Jr. ;
Fahrig, Rebecca .
MEDICAL PHYSICS, 2015, 42 (11) :6757-6767
[8]   X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in mice [J].
Montay-Gruel, Pierre ;
Bouchet, Audrey ;
Jaccard, Maud ;
Patin, David ;
Serduc, Raphael ;
Aim, Warren ;
Petersson, Kristoffer ;
Petit, Benoit ;
Bailat, Claude ;
Bourhis, Jean ;
Brauer-Krisch, Elke ;
Vozenin, Marie-Catherine .
RADIOTHERAPY AND ONCOLOGY, 2018, 129 (03) :582-588
[9]   Irradiation in a flash: Unique sparing of memory in mice after whole brain irradiation with dose rates above 100 Gy/s [J].
Montay-Gruel, Pierre ;
Petersson, Kristoffer ;
Jaccard, Maud ;
Boivin, Gael ;
Germond, Jean-Francois ;
Petit, Benoit ;
Doenlen, Raphael ;
Favaudon, Vincent ;
Bochud, Francois ;
Bailat, Claude ;
Bourhis, Jean ;
Vozenin, Marie-Catherine .
RADIOTHERAPY AND ONCOLOGY, 2017, 124 (03) :365-369
[10]   Experimental Set-up for FLASH Proton Irradiation of Small Animals Using a Clinical System [J].
Patriarca, Annalisa ;
Fouillade, Charles ;
Auger, Michel ;
Martin, Frederic ;
Pouzoulet, Frederic ;
Nauraye, Catherine ;
Heinrich, Sophie ;
Favaudon, Vincent ;
Meyroneinc, Samuel ;
Dendale, Remi ;
Mazal, Alejandro ;
Poortmans, Philip ;
Verrelle, Pierre ;
De Marzi, Ludovic .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03) :619-626