Repeatability of Keratometry Measurements Obtained With Three Topographers in Keratoconic and Normal Corneas

被引:29
作者
Guilbert, Emmanuel [1 ,2 ]
Saad, Alain [1 ,2 ]
Elluard, Maud [1 ]
Grise-Dulac, Alice [1 ,2 ]
Rouger, Helene [1 ]
Gatinel, Damien [1 ,2 ]
机构
[1] Rothschild Fdn, Dept Cataract & Refract Surg, Paris, France
[2] Ctr Expertise & Res Opt Clinicians, Paris, France
关键词
COLLAGEN CROSS-LINKING; OPTICAL COHERENCE TOMOGRAPHY; FORME-FRUSTE KERATOCONUS; ORBSCAN-II; SYSTEM; ABERRATIONS; THICKNESS; READINGS; EYE;
D O I
10.3928/1081597X-20160113-01
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the repeatability of the corneal topography functions of Orbscan II (Bausch & Lomb, Rochester, NY), OPD-Scan III (Nidek, Gamagori, Japan), and iTrace (Tracey Technologies, Houston, TX) in keratoconic eyes and in a control group of normal patients. METHODS: In this prospective cohort study, patients were recruited between November 2011 and May 2012. Measurements were performed with a combined Placido-scanning slit system (Orbscan II) and two combined Placido-aberrometer systems (OPDScan III and iTrace). Repeatability limit and intraclass correlation coefficients (ICCs) of keratometric readings were calculated. RESULTS: Fifty-nine keratoconic eyes of 34 patients and 54 normal eyes of 27 patients were included. Three groups were evaluated: all stage I-IV keratoconic eyes (59 eyes), a subgroup consisting of stage I-II keratoconic eyes (41 eyes), and normal eyes (54 eyes). For almost all parameters studied, the repeatability limit was higher in the two groups of keratoconic eyes compared to normal eyes with all three topographers, indicating lower repeatability. For the maximum keratometry measurement, repeatability limit was 1.73, 1.49, and 1.41 diopters (D) in the stage I-IV keratoconic eyes group, 1.11, 1.02, and 0.98 D in the stage I-II keratoconic eyes group, and 0.61, 0.37, and 1.02 D in the normal eyes group with Orbscan II, OPD-Scan III, and iTrace, respectively. CONCLUSIONS: Topographies performed in keratoconic eyes are less repeatable than those performed in normal eyes. Threshold values of keratometric changes used to ascertain keratoconus progression should be carefully considered. Caution should be taken when interpreting the topographies of such patients. The higher variability should be taken into account before performing any treatment.
引用
收藏
页码:187 / +
页数:8
相关论文
共 34 条
  • [1] Keratoconus evaluation using the Orbscan Topography System
    Auffarth, GU
    Wang, L
    Völcker, HE
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (02) : 222 - 228
  • [2] Average 3-Dimensional Models for the Comparison of Orbscan II and Pentacam Pachymetry Maps in Normal Corneas
    Bourges, Jean-Louis
    Alfonsi, Nicolas
    Laliberte, Jean-Francois
    Chagnon, Miguel
    Renard, Gilles
    Legeais, Jean-Marc
    Brunette, Isabelle
    [J]. OPHTHALMOLOGY, 2009, 116 (11) : 2064 - 2071
  • [3] Contralateral Eye Study of Corneal Collagen Cross-linking With Riboflavin and UVA Irradiation in Patients With Keratoconus
    Coskunseven, Efekan
    Jankov, Mirko R., II
    Hafezi, Farhad
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (04) : 371 - 376
  • [4] Comparison and Repeatability of Keratometric and Corneal Power Measurements Obtained by Orbscan II, Pentacam, and Galilei Corneal Tomography Systems
    Crawford, Alexandra Z.
    Patel, Dipika V.
    McGhee, Charles N. J.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 156 (01) : 53 - 60
  • [5] Corneal thickness, curvature, and elevation readings in normal corneas: Combined Placido-Scheimpflug system versus combined Placido-scanning-slit system
    Guilbert, Emmanuel
    Saad, Alain
    Grise-Dulac, Alice
    Gatinel, Damien
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (07) : 1198 - 1206
  • [6] Repeatability and reproducibility of Galilei measurements in normal keratoconic and postrefractive corneas
    Guler, Emre
    Yagci, Ramazan
    Akyol, Mesut
    Arslanyilmaz, Zeynel
    Balci, Mehmet
    Hepsen, Ibrahim F.
    [J]. CONTACT LENS & ANTERIOR EYE, 2014, 37 (05) : 331 - 336
  • [7] Corneal and epithelial thickness in keratoconus: A comparison of ultrasonic pachymetry Orbscan II, and optical coherence tomography
    Haque, S
    Simpson, T
    Jones, L
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (05) : 486 - 493
  • [8] Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devices
    Hashemi, Hassan
    Yekta, Abbasali
    Khabazkhoob, Mehdi
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (05) : 1065 - 1072
  • [9] International Organization for Standardization, 1977, 3534 ISO
  • [10] Repeatability of ocular aberration measurements in patients with keratoconus
    Jinabhai, Amit
    Radhakrishnan, Hema
    O'Donnell, Clare
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2011, 31 (06) : 588 - 594