The effect of injectable biocompatible elastomer (PDMS) on the strength of the proximal fixation of endovascular aneurysm repair grafts: An in vitro study

被引:10
作者
Bosman, Willem-Maarten P. F. [1 ]
van der Steenhoven, Tim J. [1 ,2 ]
Suarez, Daniel R. [3 ,4 ]
Valstar, Edward R. [3 ,4 ]
de Vries, Alexander C.
Brom, Hans L. F.
Jacobs, Michael J. [5 ]
Hamming, Jaap F. [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Surg, NL-2300 RC Leiden, Netherlands
[2] St Elizabeth Hosp, Dept Surg, Bochum, Germany
[3] Leiden Univ, Med Ctr, Dept Orthopaed, Biomech & Imaging Grp, NL-2300 RC Leiden, Netherlands
[4] Delft Univ Technol, Dept Precis & Microsyst Engn, Fac Mech Maritime & Mat Engn, NL-2600 AA Delft, Netherlands
[5] Maastricht Univ Med Ctr, Dept Surg, Maastricht, Netherlands
关键词
ABDOMINAL AORTIC-ANEURYSM; STENT-GRAFT; EUROSTAR EXPERIENCE; RISK-FACTORS; MIGRATION; RUPTURE; COMPLICATIONS; AUGMENTATION; SUITABILITY; DIAMETER;
D O I
10.1016/j.jvs.2010.01.026
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: One of the major concerns in the long-term success of endovascular aneurysm repair (EVAR) is stent graft migration, which can cause type I endoleak and even aneurysm rupture. Fixation depends on the mechanical forces between the graft and both the aortic neck and the blood flow. Therefore, there are anatomical restrictions for EVAR, such as short and angulated necks. To improve the fixation of EVAR grafts, elastomer (PDMS) can be injected in the aneurysm sac. The support given by the elastomer might prevent dislocation and migration of the graft. The aim of this study was to measure the influence of an injectable biocompatible elastomer on the fixation strength of different EVAR grafts in an in vitro model. Methods: The proximal part of three different stent grafts was inserted in a bovine artery with an attached latex aneurysm. The graft was connected to a tensile testing machine, applying force to the proximal fixation, while the artery with the aneurysm was fixated to the setup. The force to obtain graft dislodgement (DF) from the aorta was recorded in Newtons (N). Three different proximal seal lengths (5, 10, and 15 mm) were evaluated. The experiments were repeated after the space between the graft and the latex aneurysm was filled with the elastomer. Independent sample nests were used for the comparison between the DF before and after elastomer treatment for each seal length. Results: The mean DF (mean +/- SD) of all grafts without elastomer sac filling for a proximal seal length of 5, 10, and 15 mm were respectively, 4.4 +/- 3.1 N, 12.2 +/- 10.6 N, and 15.1 +/- 6.9 N. After elastomer sac filling, the dislodgement forces increased significantly (P < .001) to 20.9 +/- 3.8 N, 31.8 +/- 9.8 N, and 36.0 +/- 14.1 N, respectively. Conclusions: The present study shows that aneurysm sac filling may have a role as an adjuvant procedure to the present EVAR technique. The strength of the proximal fixation of three different stent grafts increases significantly in this in vitro setting. Further in vivo research must be done to see if this could facilitate the treatment of aneurysms with short infrarenal necks. (J Vase Surg 2010;52:152-8.)
引用
收藏
页码:152 / 158
页数:7
相关论文
共 43 条
  • [1] Anatomical risk factors for proximal perigraft endoleak and graft migration following endovascular repair of abdominal aortic aneurysms
    Albertini, JN
    Kalliafas, S
    Travis, S
    Yusuf, SW
    Macierewicz, JA
    Whitaker, SC
    Elmarasy, NM
    Hopkinson, BR
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2000, 19 (03) : 308 - 312
  • [2] Rupture of an abdominal aortic aneurysm after endovascular graft placement and aneurysm size reduction
    Alimi, YS
    Chakfe, N
    Rivoal, E
    Slimane, KK
    Valerio, N
    Riepe, G
    Kretz, JG
    Juhan, C
    [J]. JOURNAL OF VASCULAR SURGERY, 1998, 28 (01) : 178 - 183
  • [3] IN-VITRO EVALUATION OF ENDOVASCULAR STENTS TO ASSESS SUITABILITY FOR ENDOVASCULAR GRAFT FIXATION
    ANDREWS, SM
    ANSON, AW
    GREENHALGH, RM
    NOTT, DM
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1995, 9 (04) : 403 - 407
  • [4] ARKLES B, 1983, CHEMTECH, V13, P542
  • [5] Armon MP, 1997, BRIT J SURG, V84, P178, DOI 10.1046/j.1365-2168.1997.02596.x
  • [6] In Vitro Assessment of EmboGel and UltraGel Radiopaque Hydrogels for the Endovascular Treatment of Aneurysms
    Barnett, Brad P.
    Hughes, Andrew H.
    Lin, Sopo
    Arepally, Aravind
    Gailloud, Philippe H.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 20 (04) : 507 - 512
  • [7] The Proximal Fixation Strength of Modern EVAR Grafts in a Short Aneurysm Neck. An In Vitro Study
    Bosman, W. M. P. F.
    v. d. Steenhoven, T. J.
    Suarez, D. R.
    Hinnen, J. W.
    Valstar, E. R.
    Hamming, J. F.
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2010, 39 (02) : 187 - 192
  • [8] Aortic Customize: A new alternative endovascular approach to aortic aneurysm repair using injectable biocompatible elastomer. An in vitro study
    Bosman, Willem-Maarten P. F.
    van der Steenhoven, Tim J.
    Hinnen, Jan-Willem
    Kaptein, Bart L.
    de Vries, Alexander C.
    Brom, Hans L. F.
    Jacobs, Michael J.
    Hamming, Jaap F.
    [J]. JOURNAL OF VASCULAR SURGERY, 2010, 51 (05) : 1230 - 1237
  • [9] Outcomes of endovascular abdominal aortic aneurysm repair in patients with hostile neck anatomy
    Choke, Edward
    Munneke, Graham
    Morgan, Robert
    Belli, Anna-Maria
    Loftus, Ian
    McFarland, Robert
    Loosemore, Thomas
    Thompson, Matthew M.
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (06) : 975 - 980
  • [10] Endovascular abdominal aortic aneurysm repair and renal complications: A comparison between suprarenal and infrarenal fixation of stent grafts
    Cotroneo A.R.
    Iezzi R.
    Giancristofaro D.
    Santoro M.
    Pierro A.
    Spigonardo F.
    Storto M.L.
    [J]. La radiologia medica, 2007, 112 (2) : 252 - 263