Anatomic tibial component design can increase tibial coverage and rotational alignment accuracy: a comparison of six contemporary designs

被引:66
作者
Dai, Yifei [1 ]
Scuderi, Giles R. [2 ]
Bischoff, Jeffrey E. [1 ]
Bertin, Kim
Tarabichi, Samih [3 ]
Rajgopal, Ashok [4 ]
机构
[1] Zimmer Inc, Warsaw, IN 46581 USA
[2] Insall Scott Kelly Inst, New York, NY 10065 USA
[3] Burjeel Hosp Adv Surg, Dubai, U Arab Emirates
[4] Medanta Bone & Joint Inst, Gurgaon 122002, Haryana, India
关键词
Total knee arthroplasty; Tibia; Fit; Morphology; Compromise; Rotational alignment; Coverage; Overhang; TOTAL KNEE ARTHROPLASTY; REPLACEMENT; PAIN; ANTHROPOMETRY; MORPHOLOGY; OVERHANG; SHAPE; SIZE;
D O I
10.1007/s00167-014-3282-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The aim of this study was to comprehensively evaluate contemporary tibial component designs against global tibial anatomy. We hypothesized that anatomically designed tibial components offer increased morphological fit to the resected proximal tibia with increased alignment accuracy compared to symmetric and asymmetric designs. Using a multi-ethnic bone dataset, six contemporary tibial component designs were investigated, including anatomic, asymmetric, and symmetric design types. Investigations included (1) measurement of component conformity to the resected tibia using a comprehensive set of size and shape metrics; (2) assessment of component coverage on the resected tibia while ensuring clinically acceptable levels of rotation and overhang; and (3) evaluation of the incidence and severity of component downsizing due to adherence to rotational alignment and overhang requirements, and the associated compromise in tibial coverage. Differences in coverage were statistically compared across designs and ethnicities, as well as between placements with or without enforcement of proper rotational alignment. Compared to non-anatomic designs investigated, the anatomic design exhibited better conformity to resected tibial morphology in size and shape, higher tibial coverage (92 % compared to 85-87 %), more cortical support (posteromedial region), lower incidence of downsizing (3 % compared to 39-60 %), and less compromise of tibial coverage (0.5 % compared to 4-6 %) when enforcing proper rotational alignment. The anatomic design demonstrated meaningful increase in tibial coverage with accurate rotational alignment compared to symmetric and asymmetric designs, suggesting its potential for less intra-operative compromises and improved performance. III.
引用
收藏
页码:2911 / 2923
页数:13
相关论文
共 37 条
  • [1] Barrack RL, 2001, CLIN ORTHOP RELAT R, P46
  • [2] Internal Rotation of the Tibial Component is Frequent in Stiff Total Knee Arthroplasty
    Bedard, Martin
    Vince, Kelly G.
    Redfern, John
    Collen, Stacy R.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (08) : 2346 - 2355
  • [3] Malrotation causing patellofemoral complications after total knee arthroplasty
    Berger, RA
    Crossett, LS
    Jacobs, JJ
    Rubash, HE
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (356) : 144 - 153
  • [4] Bindelglass D, 1991, TECH ORTHOP, V6, P80
  • [5] Mediolateral oversizing influences pain, function, and flexion after TKA
    Bonnin, Michel P.
    Schmidt, Axel
    Basiglini, Luca
    Bossard, Nadine
    Dantony, Emmanuelle
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) : 2314 - 2324
  • [6] PATIENT OUTCOMES FOLLOWING TRICOMPARTMENTAL TOTAL KNEE REPLACEMENT - A METAANALYSIS
    CALLAHAN, CM
    DRAKE, BG
    HECK, DA
    DITTUS, RS
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 271 (17): : 1349 - 1357
  • [7] Tibial component overhang following unicompartmental knee replacement-Does it matter?
    Chau, R.
    Gulati, A.
    Pandit, H.
    Beard, D. J.
    Price, A. J.
    Dodd, C. A. F.
    Gill, H. S.
    Murray, D. W.
    [J]. KNEE, 2009, 16 (05) : 310 - 313
  • [8] Three dimensional morphometry of the knee to design the total knee arthroplasty for Chinese population
    Cheng, Fu Bo
    Ji, Xiao Feng
    Lai, Ying
    Feng, Jia Chun
    Zheng, Wen Xu
    Sun, Yue Fang
    Fu, Yao Wen
    Li, You Qiong
    [J]. KNEE, 2009, 16 (05) : 341 - 347
  • [9] Knee cartilage topography, thickness, and contact areas from MRI: in-vitro calibration and in-vivo measurements
    Cohen, ZA
    McCarthy, DM
    Kwak, SD
    Legrand, P
    Fogarasi, F
    Ciaccio, EJ
    Ateshian, GA
    [J]. OSTEOARTHRITIS AND CARTILAGE, 1999, 7 (01) : 95 - 109
  • [10] Influence of landmark and surgical variability on virtual assessment of total knee arthroplasty
    Dai, Y.
    Seebeck, J.
    Henderson, A. D.
    Bischoff, J. E.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2014, 17 (10) : 1157 - 1164