Implantable Cardioverter Defibrillator Outcomes in Pediatric and Congenital Heart Disease: Time to System Revision

被引:16
作者
Dechert, Brynn E. [1 ]
Bradley, David J. [1 ]
Serwer, Gerald A. [1 ]
Dick, Macdonald, II [1 ]
Lapage, Martin J. [1 ]
机构
[1] Univ Michigan, Div Pediat Cardiol, Dept Pediat & Communicable Dis, Ann Arbor, MI 48109 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2016年 / 39卷 / 07期
关键词
pediatric; congenital heart disease; implantable cardioverter defibrillator; complications; COMPLICATIONS; THERAPY;
D O I
10.1111/pace.12878
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundImplantable cardioverter defibrillators (ICDs) are intended to prevent sudden cardiac death yet also impose a risk of morbidity. This study describes the outcomes of ICDs in a pediatric and congenital heart disease (CHD) population from a single center. MethodsRetrospective cohort study of all patients with an ICD followed at the University of Michigan Congenital Heart Center from 2005-2013. The primary outcome was ICD system revision for any reason excluding routine generator change for battery depletion. ResultsThere were 191 ICD systems in 131 patients, including 57 with CHD, 24 with hypertrophic cardiomyopathy, and 45 with structurally normal hearts. Median age was 16 years at initial implant. Total follow-up was 850 patient-years; median 4.9 years/patient. There were 43 (33%) patients who required 60 ICD revisions; 70 revisions/1,000 patient-years of follow-up. Revisions included 25 lead extractions with replacement, 21 lead additions, five lead repositions, and four full system revisions. Kaplan-Meier (K-M) median time to appropriate shock was similar to the median time to system revision. K-M time to system revision was significantly affected by recalled lead performance. ConclusionsThe need for ICD system revision is high in this pediatric and CHD population and occurs at a rate similar to the rate of receiving appropriate therapy. These results highlight the need for judicious implant criteria and improved device longevity.
引用
收藏
页码:703 / 708
页数:6
相关论文
共 13 条
[1]   Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients [J].
Alter, P ;
Waldhans, S ;
Plachta, E ;
Moosdorf, R ;
Grimm, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (09) :926-932
[2]   Multi-Institutional Study of Implantable Defibrillator Lead Performance in Children and Young Adults: Results of the Pediatric Lead Extractability and Survival Evaluation (PLEASE) Study [J].
Atallah, Joseph ;
Erickson, Christopher C. ;
Cecchin, Frank ;
Dubin, Anne M. ;
Law, Ian H. ;
Cohen, Mitchell I. ;
LaPage, Martin J. ;
Cannon, Bryan C. ;
Chun, Terrence U. H. ;
Freedenberg, Vicki ;
Gierdalski, Marcin ;
Berul, Charles I. .
CIRCULATION, 2013, 127 (24) :2393-2402
[3]   Time Dependence of Risks and Benefits in Pediatric Primary Prevention Implantable Cardioverter-Defibrillator Therapy [J].
DeWitt, Elizabeth S. ;
Triedman, John K. ;
Cecchin, Frank ;
Mah, Doug Y. ;
Abrams, Dominic J. ;
Walsh, Edward P. ;
Gauvreau, Kimberlee ;
Alexander, Mark E. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (06) :1057-U137
[4]   Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure [J].
Hauser, Robert G. ;
Hayes, David L. .
HEART RHYTHM, 2009, 6 (05) :605-610
[5]   Implantable cardioverter defibrillator therapy for congenital long QT syndrome: A single-center experience [J].
Horner, Justin M. ;
Kinoshita, Masayoshi ;
Webster, Tracy L. ;
Haglund, Carla M. ;
Friedman, Paul A. ;
Ackerman, Michael J. .
HEART RHYTHM, 2010, 7 (11) :1616-1622
[6]   Implanted Defibrillators in Young Hypertrophic Cardiomyopathy Patients: A Multicenter Study [J].
Kamp, Anna N. ;
Von Bergen, Nicholas H. ;
Henrikson, Charles A. ;
Makhoul, Majd ;
Saarel, Elizabeth V. ;
LaPage, Martin J. ;
Russell, Mark W. ;
Strieper, Margaret ;
Yu, Sunkyung ;
Dick, Macdonald ;
Day, Sharlene M. ;
Bradley, David J. .
PEDIATRIC CARDIOLOGY, 2013, 34 (07) :1620-1627
[7]   The National ICD Registry Report: Version 2.1 including leads and pediatrics for years 2010 and 2011 [J].
Kremers, Mark S. ;
Hammill, Stephen C. ;
Berul, Charles I. ;
Koutras, Christina ;
Curtis, Jeptha S. ;
Wang, Yongfei ;
Beachy, Jim ;
Meisnere, Laura Blum ;
Conyers, Del M. ;
Reynolds, Matthew R. ;
Heidenreich, Paul A. ;
Al-Khatib, Sana M. ;
Pina, Ileana L. ;
Blake, Kathleen ;
Walsh, Mary Norine ;
Wilkoff, Bruce L. ;
Shalaby, Alaa ;
Masoudi, Frederick A. ;
Rumsfeld, John .
HEART RHYTHM, 2013, 10 (04) :E59-E65
[8]   Inappropriate ICD Discharges in Single-Chamber Versus Dual-Chamber Devices in the Pediatric and Young Adult Population [J].
Lawrence, David ;
Von Bergen, Nicholas ;
Law, Ian H. ;
Bradley, David J. ;
Dick, Macdonald, II ;
Frias, Patricio A. ;
Streiper, Margaret J. ;
Fischbach, Peter S. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (03) :287-290
[9]   Multicenter experience with extraction of the Riata/Riata ST ICD lead [J].
Maytin, Melanie ;
Wilkoff, Bruce L. ;
Brunner, Michael ;
Cronin, Edmond ;
Love, Charles J. ;
Bongiorni, Maria Grazia ;
Segreti, Luca ;
Carrillo, Roger G. ;
Garisto, Juan D. ;
Kutalek, Steven ;
Subzposh, Faiz ;
Fischer, Avi ;
Coffey, James O. ;
Gangireddy, Sandeep R. ;
Saba, Samir ;
Mittal, Suneet ;
Arshad, Aysha ;
O'Keefe, Ryan Michael ;
Henrikson, Charles A. ;
Belott, Peter ;
John, Roy M. ;
Epstein, Laurence M. .
HEART RHYTHM, 2014, 11 (09) :1613-1618
[10]   Efficacy of Implantable Cardioverter Defibrillators in Young Patients With Catecholaminergic Polymorphic Ventricular Tachycardia: Success Depends on Substrate [J].
Miyake, Christina Y. ;
Webster, Gregory ;
Czosek, Richard J. ;
Kantoch, Michal J. ;
Dubin, Anne M. ;
Avasarala, Kishor ;
Atallah, Joseph .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2013, 6 (03) :579-587