Modeling of a Piezoelectric MEMS Micropump Dedicated to Insulin Delivery and Experimental Validation Using Integrated Pressure Sensors: Application to Partial Occlusion Management

被引:12
作者
Fournier, S. [1 ]
Chappel, E. [1 ]
机构
[1] Debiotech SA, 28 Ave Sevelin, CH-1004 Lausanne, Switzerland
关键词
SIMULATION;
D O I
10.1155/2017/3719853
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A numerical model based on equivalent electrical networks has been built to simulate the dynamic behavior of a positive-displacement MEMS micropump dedicated to insulin delivery. This device comprises a reservoir in direct communication with the inlet check valve, a pumping membrane actuated by a piezo actuator, two integrated piezoresistive pressure sensors, an anti-free-flow check valve at the outlet, and finally a fluidic pathway up to the patient cannula. The pressure profiles delivered by the sensors are continuously analyzed during the therapy in order to detect failures like occlusion. The numerical modeling is a reliable way to better understand the behavior of the micropump in case of failure. The experimental pressure profiles measured during the actuation phase have been used to validate the numerical modeling. The effect of partial occlusion on the pressure profiles has been also simulated. Based on this analysis, a new management of partial occlusion for MEMS micropump is finally proposed.
引用
收藏
页数:7
相关论文
共 12 条
[1]   Modeling micropumps with electrical equivalent networks [J].
Bourouina, T ;
Grandchamp, JP .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1996, 6 (04) :398-404
[2]   Design and simulation of an electrostatic micropump for drug-delivery applications [J].
Bourouina, T ;
Bosseboeuf, A ;
Grandchamp, JP .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1997, 7 (03) :186-188
[3]   Toward understanding insulin fibrillation [J].
Brange, J ;
Andersen, L ;
Laursen, ED ;
Meyn, G ;
Rasmussen, E .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1997, 86 (05) :517-525
[4]  
Chappel E., 2014, P MICR BIOMEMS MED M
[5]   Simulation and optimization of a piezoelectric micropump for medical applications [J].
Cui, Qifeng ;
Liu, Chengliang ;
Zha, Xuan F. .
INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2008, 36 (5-6) :516-524
[6]   Insulin Micropump with Embedded Pressure Sensors for Failure Detection and Delivery of Accurate Monitoring [J].
Dumont-Fillon, Dimitry ;
Tahriou, Hassen ;
Conan, Christophe ;
Chappel, Eric .
MICROMACHINES, 2014, 5 (04) :1161-1172
[7]   Equivalent electrical network for performance characterization of piezoelectric peristaltic micropump [J].
Hsu, Y. C. ;
Le, N. B. .
MICROFLUIDICS AND NANOFLUIDICS, 2009, 7 (02) :237-248
[8]   Modeling and experimental validation of a piezoelectric micropump with novel no-moving-part valves [J].
Izzo, Ivano ;
Accoto, Dino ;
Menciassi, Arianna ;
Schmitt, Lothar ;
Dario, Paolo .
SENSORS AND ACTUATORS A-PHYSICAL, 2007, 133 (01) :128-140
[9]  
Kerr David, 2008, J Diabetes Sci Technol, V2, P450
[10]  
Tanaka S., 2007, TRANSDUCERS '07 & Eurosensors XXI. 2007 14th International Conference on Solid-State Sensors, Actuators and Microsystems, P2183, DOI 10.1109/SENSOR.2007.4300600