A CR-39 based thermal neutron dosimeter

被引:3
|
作者
Ahmad, N
Matiullah
Kenawy, MA
机构
[1] Ctr Nucl Studies, Islamabad, Pakistan
[2] Ain Shams Univ, Univ Coll Girls, Dept Phys, Cairo, Egypt
关键词
CR-39; LiF films; activation foils; personal neutron dosimetry; vacuum coating unit; beta-gamma coincidence;
D O I
10.1016/S1350-4487(97)00111-X
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Both chemically and electrochemically etched CR-39 detectors are used for the detection of fast neutrons. In practical situations, fast neutrons are always accompanied by thermal neutrons. Therefore, the response of the CR-39 based dosimeter has to be extended to thermal neutrons. To do so, a radiator/converter libe LiF is introduced in front of the CR-39 detector and an optimum thickness of the LiF film is determined so that a response similar to that of the fast neutrons is achieved. Thin films of natural LiF were prepared and the response of the CR-39 detector was studied as a function of the film thickness.