Determination of cadmium by FAAS after on-line enrichment using a mini column packed with Amberlite XAD-2 loaded with TAM

被引:66
作者
Melo, MHA
Ferreira, SLC
Santelli, RE
机构
[1] Univ Fed Bahia, Inst Quim, BR-40170290 Salvador, BA, Brazil
[2] Univ Fed Fluminense, Dept Geoquim, BR-24020 Rio De Janeiro, Brazil
关键词
D O I
10.1016/S0026-265X(00)00029-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A procedure for determination of trace amounts of cadmium with on-line preconcentration has been proposed. It is based on chemical sorption of cadmium(II) ions onto a minicolumn packed with Amberlite XAD-2 resin loaded with TAM reagent. After being preconcentrated, cadmium was eluted by 0.50 mol l(-1) hydrochloric acid solution and straight determined on-line by flame atomic absorption spectrometry. A time-based technique was used. Chemical and flow variables as well as concomitant ions were studied in the developed procedure. Detection limit (3 sigma, n = 22) was 1.2 mu g l(-1) for the analytical curve within the range of 0-200.0 mu g l(-1). An enrichment factor of 108 and a retention efficiency of 75.6% were achieved for a sample volume of 5.2 mi (1 min preconcentration time). The precision of the continuous Preconcentration method, calculated as the relative standard deviation in sample solutions containing 10 mu g l(-1) of cadmium was 4 and 9%, regarding repeatability and reproducibility, respectively. The selectivity of the proposed procedure was also evaluated. Results showed that iron(III), zinc(II), nickel(II), manganese(II), aluminum(III), copper(II), lead(II), cobalt(II), chromium(VI) (200 mu g l(-1) concentration) did not interfere during cadmium determination (10.0 mu g l(-1) concentration). Achieved analytical results for rice flour (NIES 10c) and lobster hepatopancreas (TORT-1) standard reference materials were in good agreement with the certified values. A high sample throughput (40 samples h(-1)), preconcentration factor and simplicity are the main advantages in this analytical procedure. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:59 / 65
页数:7
相关论文
共 20 条
  • [1] Improved dithiocarbamate/oxine solvent extraction method for the preconcentration of trace metals from seawater using metal exchange back-extraction
    Batterham, GJ
    Parry, DL
    [J]. MARINE CHEMISTRY, 1996, 55 (3-4) : 381 - 388
  • [2] CRITICAL COMPARISON OF METHODS FOR MERCURY DETERMINATION IN FISH
    BORTOLI, A
    GEROTTO, M
    MARCHIORI, M
    MUNTAU, H
    REHNERT, A
    [J]. MIKROCHIMICA ACTA, 1995, 119 (3-4) : 305 - 310
  • [3] CHEN KL, 1992, HDB ORGANIC REAGENT
  • [4] PRECONCENTRATION OF TRACE-METALS IN RIVER WATERS BY THE APPLICATION OF CHELATE ADSORPTION ON AMBERLITE XAD-4
    ELCI, L
    SOYLAK, M
    DOGAN, M
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 342 (1-2): : 175 - 178
  • [5] Determination of cadmium in mussels by flame atomic absorption spectrometry with preconcentration on a chelating resin in a flow injection system
    Enríquez-Domínguez, MF
    Yebra-Biurrun, MC
    Bermejo-Barrera, MP
    [J]. ANALYST, 1998, 123 (01) : 105 - 108
  • [6] Fang Z., 1993, FLOW INJECTION SEPAR
  • [7] Separation and preconcentration of cobalt after sorption onto amberlite XAD-2 loaded with 2-(2-thiazolylazo)-p-cresol
    Ferreira, SLC
    de Brito, CF
    [J]. ANALYTICAL SCIENCES, 1999, 15 (02) : 189 - 191
  • [8] Nickel determination in saline matrices by ICP-AES after sorption on Amberlite XAD-2 loaded with PAN
    Ferreira, SLC
    de Brito, CF
    Dantas, AF
    de Araújo, NML
    Costa, ACS
    [J]. TALANTA, 1999, 48 (05) : 1173 - 1177
  • [9] An on-line continuous flow system for copper enrichment and determination by flame atomic absorption spectroscopy
    Ferreira, SLC
    Lemos, VA
    Moreira, BC
    Costa, ACS
    Santelli, RE
    [J]. ANALYTICA CHIMICA ACTA, 2000, 403 (1-2) : 259 - 264
  • [10] Collection of trace heavy metals on dithizone-impregnated admicelles for water analysis
    Hiraide, M
    Shibata, W
    [J]. ANALYTICAL SCIENCES, 1998, 14 (06) : 1085 - 1088