A novel poly (4-methyl-1-pentene)/polypropylene (PMP/PP) thin film composite (TFC) artificial lung membrane for enhanced gas transport and excellent hemo-compatibility

被引:47
作者
Wang, Yiwen [1 ,2 ,3 ]
Liu, Yang [1 ,2 ,3 ]
Han, Qiu [1 ,3 ]
Lin, Haibo [1 ,3 ]
Liu, Fu [1 ,2 ,3 ]
机构
[1] Chinese Acad Sci, Ningbo Inst Mat Technol & Engn, Key Lab Marine Mat & Related Technol, Zhejiang Key Lab Marine Mat & Protect Technol, 1219 Zhongguan West Rd, Ningbo 315201, Peoples R China
[2] Univ Chinese Acad Sci, 19 A Yuquan Rd, Beijing 100049, Peoples R China
[3] Key Lab Biobased Polymer Mat Technol & Applicat Z, 1219 Zhongguan West Rd, Ningbo 315201, Peoples R China
关键词
ECMO; Composite membrane; Anticoagulant surface; Gas permeation; Plasma leakage; PMP/PP; TFC; OXYGENATION; COMPLEMENT; THROMBOSIS; MECHANISM;
D O I
10.1016/j.memsci.2022.120359
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Extracorporeal membrane oxygenation is a technique that provides short-term supports to the heart and lungs. It removes CO2 from the blood and provides enough oxygen, which is a huge help in the fight against COVID-19. As the key component, the artificial lung membranes have evolved in three generations including silicon, polypropylene and poly (4-methyl-1-pentene). Herein, we for the first time design and fabricate a novel poly (4methyl-1-pentene)/polypropylene (PMP/PP) thin film composite (TFC) membrane with the anticoagulant coating composed of poly (sodium 4-styrenesulfonate) and cross-linked poly (vinyl alcohol). Poly (sodium 4-styrenesulfonate) provides sulfonic acid groups to inhibit the coagulant factors (FVIII and FXII), and cross-linked poly (vinyl alcohol) increase the stability of the anticoagulant coating and further improve the hydrophilicity via abundant hydroxyl groups to depress the protein adsorption. Long-term anticoagulant property was demonstrated by whole human blood for 28 days. Blood compatibility was evaluated by hemolysis rate, anticoagulation activity (APTT, TT and PT), complement activation, platelet activation and contact activation. Pure CO2, O2 and N2 permeation rates were determined to evaluate the mass transfer properties of PMP/PP TFC membranes. Gas permeation results revealed that gas permeation flux increased in the TFC membranes because of the decrease of crystallinity. Overall, the so prepared PMP/PP membrane shows good CO2/O2 selectivity and blood compatibility as novel TFC artificial lung membrane.
引用
收藏
页数:13
相关论文
共 44 条
[1]  
Cheung P.Y., MECH PLATELET DYSFUN
[2]   The mechanisms of platelet dysfunction during extracorporeal membrane oxygenation in critically ill neonates [J].
Cheung, PY ;
Sawicki, G ;
Salas, E ;
Etches, PC ;
Schulz, R ;
Radomski, MW .
CRITICAL CARE MEDICINE, 2000, 28 (07) :2584-2590
[3]   Membranes in Extracorporeal Blood Oxygenation Technology [J].
Evseev, A. K. ;
Zhuravel, S., V ;
Alentiev, A. Yu ;
Goroncharovskaya, I., V ;
Petrikov, S. S. .
MEMBRANES AND MEMBRANE TECHNOLOGIES, 2019, 1 (04) :201-211
[4]   Is COVID-19 associated thrombosis caused by overactivation of the complement cascade? A literature review [J].
Fletcher-Sandersjoo, Alexander ;
Bellander, Bo-Michael .
THROMBOSIS RESEARCH, 2020, 194 :36-41
[5]   Extraordinary Transport Behavior of Gases in Isothermally Annealed Poly(4-Methyl-1-Pentene) Membranes [J].
Fu, Ywu-Jang ;
Lai, Cheng-Lee ;
Hu, Chien-Chieh ;
Sun, Yi-Ming ;
Wu, Shuan-Ying ;
Chen, Jung-Tsai ;
Huang, Shu-Hsien ;
Hung, Wei-Song ;
Lee, Kueir-Rarn .
JOURNAL OF POLYMER SCIENCE PART B-POLYMER PHYSICS, 2016, 54 (22) :2368-2376
[6]   Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes [J].
Gorbet, MB ;
Sefton, MV .
BIOMATERIALS, 2004, 25 (26) :5681-5703
[7]  
Hadaya Joseph, 2020, JAMA, V323, P2536, DOI 10.1001/jama.2020.9148
[8]  
Henchir K.A., 2004, LUNG SURFACTANTS
[9]   Surface monofunctionalized polymethyl pentene hollow fiber membranes by plasma treatment and hemocompatibility modification for membrane oxygenators [J].
Huang, Xin ;
Wang, Weiping ;
Zheng, Zhi ;
Fan, Wenling ;
Mao, Chun ;
Shi, Jialiang ;
Li, Lei .
APPLIED SURFACE SCIENCE, 2016, 362 :355-363
[10]   Safety and Efficacy of a Novel Pneumatically Driven Extracorporeal Membrane Oxygenation Device [J].
Karagiannidis, Christian ;
Joost, Thilo ;
Strassmann, Stephan ;
Weber-Carstens, Steffen ;
Combes, Alain ;
Windisch, Wolfram ;
Brodie, Daniel .
ANNALS OF THORACIC SURGERY, 2020, 109 (06) :1684-1691