Tracheal Intubation of a Difficult Airway Using Airway Scope, Airtraq, and Macintosh Laryngoscope: A Comparative Manikin Study of Inexperienced Personnel

被引:33
作者
Liu, Liangji [1 ]
Tanigawa, Koichi [1 ]
Kusunoki, Shinji [1 ]
Tamura, Tomoko [1 ]
Ota, Kohei [1 ]
Yamaga, Satoshi [1 ]
Kida, Yoshiko [1 ]
Otani, Tadatsugu [1 ]
Sadamori, Takuma [1 ]
Takeda, Taku [1 ]
Iwasaki, Yasumasa [1 ]
Hirohashi, Nobuyuki [1 ]
机构
[1] Hiroshima Univ, Dept Emergency & Crit Care Med, Hiroshima 7348551, Japan
关键词
VIDEO LARYNGOSCOPE; PENTAX-AWS; PENTAX-AWS(R); PERFORMANCE;
D O I
10.1213/ANE.0b013e3181d2aad7
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: The Airway Scope (AWS) (Pentax-AWS, Hoya Corp., Tokyo, Japan) and the Airtraq (ATQ) (Prodol, Vizcaya, Spain) have similarities in the novel structures of their blades. In this study, we evaluated the ease of use of the AWS and ATQ compared with the Macintosh laryngoscope (ML) by inexperienced personnel in a simulated manikin difficult airway. METHODS: Twenty-four fifth-year medical students with no previous experience in tracheal intubation participated in this study. We used an advanced patient simulator (SimMan (R), Laerdal Medical, Stavanger, Norway) to simulate difficult airway scenarios including cervical spine rigidity, limited mouth opening, and pharyngeal obstruction. The sequences in selecting devices and scenarios were randomized. Success rates for tracheal intubation, and the time required for visualization of the glottis, tracheal intubation, and inflation of the lungs, and the number of optimization maneuvers and dental click sounds were analyzed. The 3 different intubation devices were tested in 4 different scenarios by 24 students. RESULTS: Both the AWS and ATQ had very high success rates of tracheal intubation compared with the ML (AWS 100%*; ATQ 98%*; and ML 89%; *P < 0.05 AWS, ATQ versus ML). The time to intubation with the AWS was significantly shorter than with the ATQ and ML (AWS 11 6 seconds; ATQ 16 12 seconds; and ML 16 11 seconds; *P < 0.05 AWS versus ATQ, ML). The number of optimization maneuvers with the AWS was significantly lower than with the ATQ and ML. There were significantly more audible dental click sounds with the ML than with the AWS and ATQ. CONCLUSION: Both the AWS and ATQ may be suitable devices for difficult intubation by inexperienced personnel in this manikin simulated scenario. Further studies in a clinical setting are necessary to confirm these findings. (Anesth Ana Ig 2010;110:1049-55)
引用
收藏
页码:1049 / 1055
页数:7
相关论文
共 15 条
  • [1] Use of the Pentax-AWS® in 293 Patients with Difficult Airways
    Asai, Takashi
    Liu, Eugene H.
    Matsumoto, Sanae
    Hirabayashi, Yoshihiro
    Seo, Norimasa
    Suzuki, Akihiro
    Toi, Takashi
    Yasumoto, Kazumasa
    Okuda, Yasuhisa
    [J]. ANESTHESIOLOGY, 2009, 110 (04) : 898 - 904
  • [2] A comparison of two techniques for inserting the Airtraq™ laryngoscope in morbidly obese patients
    Dhonneur, G.
    Ndoko, S. K.
    Amathieu, R.
    Attias, A.
    Housseini, L. E. L.
    Polliand, C.
    Tual, L.
    [J]. ANAESTHESIA, 2007, 62 (08) : 774 - 777
  • [3] Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study
    Enomoto, Y.
    Asai, T.
    Arai, T.
    Kamishima, K.
    Okuda, Y.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) : 544 - 548
  • [4] Hapgood A, 2008, ANAESTHESIA, V63, P889, DOI 10.1111/j.1365-2044.2008.05620.x
  • [5] Airway Scope: early clinical experience in 405 patients
    Hirabayashi, Yoshihiro
    Seo, Norimasa
    [J]. JOURNAL OF ANESTHESIA, 2008, 22 (01) : 81 - 85
  • [6] Description and first clinical application of AirWay Scope for tracheal intubation
    Koyama, Jun-ichi
    Aoyama, Tatsuro
    Kusano, Yoshikazu
    Seguchi, Tatsuya
    Kawagishi, Kyutaro
    Iwashita, Tomomi
    Okamoto, Kazufumi
    Okudera, Hiroshi
    Takasuna, Hiroshi
    Hongo, Kazuhiro
    [J]. JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 2006, 18 (04) : 247 - 250
  • [7] Comparison of the Airway Scope®, gum elastic bougie and fibreoptic bronchoscope in simulated difficult tracheal intubation:: a manikin study
    Koyama, Y.
    Inagawa, G.
    Miyashita, T.
    Kikuchi, T.
    Miura, N.
    Miki, T.
    Kurihara, R.
    Kamiya, Y.
    Goto, T.
    [J]. ANAESTHESIA, 2007, 62 (09) : 936 - 939
  • [8] Learning and performance of tracheal intubation by novice personnel:: a comparison of the Airtraq® and Macintosh laryngoscope
    Maharaj, C. H.
    Costello, J. F.
    Higgins, B. D.
    Harte, B. H.
    Laffey, J. G.
    [J]. ANAESTHESIA, 2006, 61 (07) : 671 - 677
  • [9] Evaluation of intubation using the Airtraq® or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy -: a manikin study
    Maharaj, CH
    Higgins, BD
    Harte, BH
    Laffey, JG
    [J]. ANAESTHESIA, 2006, 61 (05) : 469 - 477
  • [10] Tracheal intubation by inexperienced medical residents using the Airtraq and Macintosh laryngoscopes - a manikin study
    Maharaj, Chrisen H.
    Chonghaile, Martina Ni
    Higgins, Brendan D.
    Harte, Brian H.
    Laffey, John G.
    [J]. AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2006, 24 (07) : 769 - 774