A Comparative Study of Single-pulse and Double-pulse Laser-Induced Breakdown Spectroscopy with Uranium-containing Samples

被引:32
|
作者
Skrodzki, Patrick J. [1 ]
Becker, Jason R. [1 ]
Diwakar, Prasoon K. [1 ]
Harilal, Sivanandan S. [2 ]
Hassanein, Ahmed [1 ]
机构
[1] Purdue Univ, Ctr Mat Extreme Environm CMUXE, Sch Nucl Engn, 400 Cent Dr, W Lafayette, IN 47907 USA
[2] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
Double-pulse laser-induced breakdown spectroscopy; DPLIBS; Nuclear forensics; Uranium; Signal enhancement; Plasma diagnostics; WAVELENGTH; PLASMA; ENHANCEMENT; ABLATION; THORIUM;
D O I
10.1177/0003702815626670
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Laser-induced breakdown spectroscopy (LIBS) holds potential advantages in special nuclear material (SNM) sensing and nuclear forensics, which require rapid analysis, minimal sample preparation, and stand-off distance capability. SNM, such as U, however, result in crowded emission spectra with LIBS, and characteristic emission lines are challenging to discern. It is well-known that double-pulse LIBS (DPLIBS) improves the signal intensity for analytes over conventional single-pulse LIBS (SPLIBS). This study investigates the U signal in a glass matrix using DPLIBS and compares it to signal obtained using SPLIBS. Double-pulse LIBS involves sequential firing of a 1.06 mu m Nd:YAG pre-pulse and 10.6 mu m TEA CO2 heating pulse in a near collinear geometry. Optimization of experimental parameters including inter-pulse delay and energy follows identification of characteristic lines for the bulk analyte Ca and the minor constituent analyte U for both DPLIBS and SPLIBS. Spatial and temporal coupling of the two pulses in the proposed DPLIBS technique yields improvements in analytical merits with a negligible increase in damage to the sample compared to SPLIBS. Subsequently, the study discusses optimum plasma emission conditions of U lines and relative figures of merit in both SPLIBS and DPLIBS. Investigation into plasma characteristics also addresses plausible mechanisms related to the observed U analyte signal variation between SPLIBS and DPLIBS.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 50 条
  • [31] Collinear double-pulse laser-induced breakdown spectroscopy based Cd profiling in the soil
    Ren, Jie
    Yang, Zihan
    Zhao, Yanru
    Yu, Keqiang
    OPTICS EXPRESS, 2022, 30 (21) : 37711 - 37726
  • [32] Investigation of double-pulse femtosecond laser induced breakdown spectroscopy of polymethyl methacrylate (PMMA)
    Wang, Mengmeng
    Wang, Sumei
    Cao, Zhitao
    Wang, Peng
    Wang, Cong
    LASER-BASED MICRO- AND NANOPROCESSING IX, 2015, 9351
  • [33] Effect of laser pulse energy on orthogonal double femtosecond pulse laser-induced breakdown spectroscopy
    Liu, Xiaoliang
    Sun, Shaohua
    Wang, Xiaoshan
    Liu, Zuoye
    Liu, Qingcao
    Ding, Pengji
    Guo, Zeqin
    Hu, Bitao
    OPTICS EXPRESS, 2013, 21 (13): : A704 - A713
  • [34] Hydrogen isotope detection in metal matrix using double-pulse laser-induced breakdown-spectroscopy
    Fantoni, Roberta
    Almaviva, Salvatore
    Caneve, Luisa
    Colao, Francesco
    Maddaluno, Giorgio
    Gasior, Pawel
    Kubkowska, Monika
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2017, 129 : 8 - 13
  • [35] Quantitative analysis of Pb in soil by femtosecond-nanosecond double-pulse laser-induced breakdown spectroscopy
    Zhao, Shangyong
    Song, Chao
    Gao, Xun
    Lin, Jingquan
    RESULTS IN PHYSICS, 2019, 15
  • [36] Effect of sample position on collinear femtosecond double-pulse laser-induced breakdown spectroscopy of silicon in air
    Qi, Hongxia
    Li, Suyu
    Qi, Ying
    Chen, Anmin
    Hu, Zhan
    Huang, Xuri
    Jin, Mingxing
    Ding, Dajun
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (06) : 1105 - 1111
  • [37] Multiple-element analysis of coal using collinear double-pulse laser-induced breakdown spectroscopy
    Ni, Qi
    He, Yong
    Weng, Wubin
    Zhu, Yanqun
    Wang, Zhihua
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (08) : 1652 - 1660
  • [38] Collinear double-pulse laser-induced breakdown spectroscopy as an in-situ diagnostic tool for wall composition in fusion devices
    Hai, Ran
    Liu, Ping
    Wu, Ding
    Ding, Hongbin
    Wu, Jing
    Luo, Guang-Nan
    FUSION ENGINEERING AND DESIGN, 2014, 89 (9-10) : 2435 - 2439
  • [39] Tomography of double-pulse laser-induced plasmas in the orthogonal geometry
    Kepes, Erik
    Gornushkin, Igor
    Porizka, Pavel
    Kaiser, Jozef
    ANALYTICA CHIMICA ACTA, 2020, 1135 : 1 - 11
  • [40] Effect of parameters on Si plasma emission in collinear double-pulse laser-induced breakdown spectroscopy
    Chen, Fang-Fang
    Su, Xue-Jiao
    Zhou, Wei-Dong
    FRONTIERS OF PHYSICS, 2015, 10 (05)