A Comparative Study of Single-pulse and Double-pulse Laser-Induced Breakdown Spectroscopy with Uranium-containing Samples

被引:32
|
作者
Skrodzki, Patrick J. [1 ]
Becker, Jason R. [1 ]
Diwakar, Prasoon K. [1 ]
Harilal, Sivanandan S. [2 ]
Hassanein, Ahmed [1 ]
机构
[1] Purdue Univ, Ctr Mat Extreme Environm CMUXE, Sch Nucl Engn, 400 Cent Dr, W Lafayette, IN 47907 USA
[2] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
Double-pulse laser-induced breakdown spectroscopy; DPLIBS; Nuclear forensics; Uranium; Signal enhancement; Plasma diagnostics; WAVELENGTH; PLASMA; ENHANCEMENT; ABLATION; THORIUM;
D O I
10.1177/0003702815626670
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Laser-induced breakdown spectroscopy (LIBS) holds potential advantages in special nuclear material (SNM) sensing and nuclear forensics, which require rapid analysis, minimal sample preparation, and stand-off distance capability. SNM, such as U, however, result in crowded emission spectra with LIBS, and characteristic emission lines are challenging to discern. It is well-known that double-pulse LIBS (DPLIBS) improves the signal intensity for analytes over conventional single-pulse LIBS (SPLIBS). This study investigates the U signal in a glass matrix using DPLIBS and compares it to signal obtained using SPLIBS. Double-pulse LIBS involves sequential firing of a 1.06 mu m Nd:YAG pre-pulse and 10.6 mu m TEA CO2 heating pulse in a near collinear geometry. Optimization of experimental parameters including inter-pulse delay and energy follows identification of characteristic lines for the bulk analyte Ca and the minor constituent analyte U for both DPLIBS and SPLIBS. Spatial and temporal coupling of the two pulses in the proposed DPLIBS technique yields improvements in analytical merits with a negligible increase in damage to the sample compared to SPLIBS. Subsequently, the study discusses optimum plasma emission conditions of U lines and relative figures of merit in both SPLIBS and DPLIBS. Investigation into plasma characteristics also addresses plausible mechanisms related to the observed U analyte signal variation between SPLIBS and DPLIBS.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 50 条
  • [21] Detection of Tantalum-Niobium Ores Using Double-Pulse Laser-Induced Breakdown Spectroscopy Based on a Single Laser
    Chen Shihang
    Liu Li
    Deng Zhiwei
    Hao Zhongqi
    Shi Jiulin
    He Xingdao
    LASER & OPTOELECTRONICS PROGRESS, 2023, 60 (13)
  • [22] DOUBLE-PULSE LASER-INDUCED BREAKDOWN SPECTROSCOPY WITH 1030 AND 257.5 nm WAVELENGTH FEMTOSECOND LASER PULSES
    Balachninaite, O.
    Baskevicius, A.
    Stankeviciute, K.
    Kurselis, K.
    Sirutkaitis, V.
    LITHUANIAN JOURNAL OF PHYSICS, 2010, 50 (01): : 105 - 110
  • [23] Single-Pulse Underwater Laser-Induced Breakdown Spectroscopy with Nongated Detection Scheme
    Matsumoto, Ayumu
    Tamura, Ayaka
    Fukami, Kazuhiro
    Ogata, Yukio H.
    Sakka, Tetsuo
    ANALYTICAL CHEMISTRY, 2013, 85 (08) : 3807 - 3811
  • [24] UV femtosecond single-pulse and collinear double-pulse laser-induced breakdown spectroscopy (LIBS) for depth-resolved characterization of nano-scaled films
    Giannakaris, N.
    Siozos, P.
    Pinon, V.
    Banerjee, S. P.
    Sentis, M.
    Pedarnig, J. D.
    Anglos, D.
    APPLIED SURFACE SCIENCE, 2023, 640
  • [25] Investigation of the osteitis deformans phases in snake vertebrae by double-pulse laser-induced breakdown spectroscopy
    Galiova, M.
    Kaiser, J.
    Novotny, K.
    Ivanov, M.
    Fisakova, M. Nyvltova
    Mancini, L.
    Tromba, G.
    Vaculovic, T.
    Liska, M.
    Kanicky, V.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 398 (02) : 1095 - 1107
  • [26] A comparative study of single and double pulse of laser induced breakdown spectroscopy of silver
    Rashid, Babar
    Ahmed, Rizwan
    Ali, Raheel
    Baig, M. A.
    PHYSICS OF PLASMAS, 2011, 18 (07)
  • [27] Signal enhancement in collinear double-pulse laser-induced breakdown spectroscopy applied to different soils
    Nicolodelli, Gustavo
    Senesi, Giorgio Saverio
    Romano, Renan Amon
    de Oliveira Perazzoli, Ivan Luiz
    Bastos Pereira Milori, Debora Marcondes
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2015, 111 : 23 - 29
  • [28] Influence of pulse-to-pulse delay for 532 nm double-pulse laser-induced breakdown spectroscopy of technical polymers
    Viskup, R.
    Praher, B.
    Linsmeyer, T.
    Scherndl, H.
    Pedarnig, J. D.
    Heitz, J.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2010, 65 (11) : 935 - 942
  • [29] Application of Stand-off Double-Pulse Laser-Induced Breakdown Spectroscopy in Elemental Analysis of Magnesium Alloy
    Qi Lifeng
    Sun Lanxiang
    Xin Yong
    Cong Zhibo
    Li Yang
    Yu Haibin
    PLASMA SCIENCE & TECHNOLOGY, 2015, 17 (08) : 676 - 681
  • [30] Enhancement and stabilization of plasma using collinear long-short double-pulse laser-induced breakdown spectroscopy
    Cui, Minchao
    Deguchi, Yoshihiro
    Wang, Zhenzhen
    Fujita, Yuki
    Liu, Renwei
    Shiou, Fang-Jung
    Zhao, Shengdun
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2018, 142 : 14 - 22