A Comparative Study of Single-pulse and Double-pulse Laser-Induced Breakdown Spectroscopy with Uranium-containing Samples

被引:32
作者
Skrodzki, Patrick J. [1 ]
Becker, Jason R. [1 ]
Diwakar, Prasoon K. [1 ]
Harilal, Sivanandan S. [2 ]
Hassanein, Ahmed [1 ]
机构
[1] Purdue Univ, Ctr Mat Extreme Environm CMUXE, Sch Nucl Engn, 400 Cent Dr, W Lafayette, IN 47907 USA
[2] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
Double-pulse laser-induced breakdown spectroscopy; DPLIBS; Nuclear forensics; Uranium; Signal enhancement; Plasma diagnostics; WAVELENGTH; PLASMA; ENHANCEMENT; ABLATION; THORIUM;
D O I
10.1177/0003702815626670
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Laser-induced breakdown spectroscopy (LIBS) holds potential advantages in special nuclear material (SNM) sensing and nuclear forensics, which require rapid analysis, minimal sample preparation, and stand-off distance capability. SNM, such as U, however, result in crowded emission spectra with LIBS, and characteristic emission lines are challenging to discern. It is well-known that double-pulse LIBS (DPLIBS) improves the signal intensity for analytes over conventional single-pulse LIBS (SPLIBS). This study investigates the U signal in a glass matrix using DPLIBS and compares it to signal obtained using SPLIBS. Double-pulse LIBS involves sequential firing of a 1.06 mu m Nd:YAG pre-pulse and 10.6 mu m TEA CO2 heating pulse in a near collinear geometry. Optimization of experimental parameters including inter-pulse delay and energy follows identification of characteristic lines for the bulk analyte Ca and the minor constituent analyte U for both DPLIBS and SPLIBS. Spatial and temporal coupling of the two pulses in the proposed DPLIBS technique yields improvements in analytical merits with a negligible increase in damage to the sample compared to SPLIBS. Subsequently, the study discusses optimum plasma emission conditions of U lines and relative figures of merit in both SPLIBS and DPLIBS. Investigation into plasma characteristics also addresses plausible mechanisms related to the observed U analyte signal variation between SPLIBS and DPLIBS.
引用
收藏
页码:467 / 473
页数:7
相关论文
共 25 条
[1]   Double pulse laser ablation and plasma: Laser induced breakdown spectroscopy signal enhancement [J].
Babushok, V. I. ;
DeLucia, F. C., Jr. ;
Gottfried, J. L. ;
Munson, C. A. ;
Miziolek, A. W. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (09) :999-1014
[2]   Analysis and Spectral Assignments of Mixed Actinide Oxide Samples Using Laser-Induced Breakdown Spectroscopy (LIBS) [J].
Barefield, James E., II ;
Judge, Elizabeth J. ;
Berg, John M. ;
Willson, Stephen P. ;
Le, Loan A. ;
Lopez, Leon N. .
APPLIED SPECTROSCOPY, 2013, 67 (04) :433-440
[3]   Inductively coupled plasma mass spectrometry (ICP-MS) and laser ablation ICP-MS for isotope analysis of long-lived radionuclides [J].
Becker, JS .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2005, 242 (2-3) :183-195
[4]   Detection of Uranium Using Laser-Induced Breakdown Spectroscopy [J].
Chinni, Rosemarie C. ;
Cremers, David A. ;
Radziemski, Leon J. ;
Bostian, Melissa ;
Navarro-Northrup, Claudia .
APPLIED SPECTROSCOPY, 2009, 63 (11) :1238-1250
[5]   Line Selection and Parameter Optimization for Trace Analysis of Uranium in Glass Matrices by Laser-Induced Breakdown Spectroscopy (LIBS) [J].
Choi, Inhee ;
Chan, George C. -Y. ;
Mao, Xianglei ;
Perry, Dale L. ;
Russo, Richard E. .
APPLIED SPECTROSCOPY, 2013, 67 (11) :1275-1284
[6]   A comparison of single and double pulse laser-induced breakdown spectroscopy of aluminum samples [J].
Colao, F ;
Lazic, V ;
Fantoni, R ;
Pershin, S .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (07) :1167-1179
[7]   The importance of longer wavelength reheating in dual-pulse laser-induced breakdown spectroscopy [J].
Coons, R. W. ;
Harilal, S. S. ;
Hassan, S. M. ;
Hassanein, A. .
APPLIED PHYSICS B-LASERS AND OPTICS, 2012, 107 (03) :873-880
[8]   Three-dimensional analysis of laser induced plasmas in single and double pulse configuration [J].
Corsi, M ;
Cristoforetti, G ;
Giuffrida, M ;
Hidalgo, M ;
Legnaioli, S ;
Palleschi, V ;
Salvetti, A ;
Tognoni, E ;
Vallebona, C .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2004, 59 (05) :723-735
[9]   Reference lines in the optogalvanic spectra of uranium and thorium over the wavelength range 694-755 nm [J].
DeGraffenreid, W ;
Sansonetti, CJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 2002, 19 (07) :1711-1715
[10]   Role of laser pre-pulse wavelength and inter-pulse delay on signal enhancement in collinear double-pulse laser-induced breakdown spectroscopy [J].
Diwakar, P. K. ;
Harilal, S. S. ;
Freeman, J. R. ;
Hassanein, A. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2013, 87 :65-73