Quantitative analytical methods for the determination of a new hypertension drug, CGS 25462, and its metabolites (CGS 25659 and CGS 24592) in human plasma by high-performance liquid chromatography

被引:5
作者
Maniara, WM [1 ]
Cipriano, A [1 ]
Powell, ML [1 ]
机构
[1] Novartis Corp, Div Pharmaceut, Res Dept, Summit, NJ 07901 USA
来源
JOURNAL OF CHROMATOGRAPHY B | 1998年 / 706卷 / 02期
关键词
CGS; 25462; 25659; 24592;
D O I
10.1016/S0378-4347(97)00566-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Two simple and sensitive reversed-phase high-performance liquid chromatography (HPLC) methods were developed and validated for the quantitative determination of a novel hypertension drug CGS 25462 and its major metabolites CGS 24592 and CGS 25659 in human plasma. CGS 25462 and CGS 25798 (internal standard) were purified by one-step liquid-liquid extraction with methylene chloride. The metabolites were analyzed on HPLC after plasma protein precipitation with 10% trichloroacetic acid (TCA). Separations were achieved on a Zorbax RX C-18 column. All compounds were detected by using a fluorescence detector. The excitation wavelength was 254 nm, and emission was monitored at 325+/-12.5 nm. Assessment of recovery and reproducibility indicated good accuracy and precision. Over the validation concentration range of 10 to 1000 ng/ml for CGS 25462 and 25 to 5000 ng/ml for both metabolites, overall mean relative recoveries were 96% for CGS 25462, 101% for CGS 25659 and 107% for CGS 24592, and the coefficients of variation were 4.6 to 13% for CGS 25462, 9.5 to 13% for CGS 25659 and 7.7 to 15% for CGS 24592. The limits of quantification (LOQs) were 10 ng/ml for CGS 25462 and 25 ng/ml for CGS 24592 and CGS 25659, which were of sufficient sensitivity to measure the concentrations of CGS 25462, CGS 25659 and CGS 24592 in plasma samples from normal volunteers following a single 800 mg oral dose. (C) 1998 Published by Elsevier Science B.V.
引用
收藏
页码:287 / 294
页数:8
相关论文
共 2 条