Insights about inductively coupled plasma optical emission spectroscopy interferences of major rare earth elements in complex e-waste feeds

被引:10
|
作者
Patil, Ajay B. [1 ,2 ]
Tarik, Mohamed [1 ]
Schuler, Albert J. [1 ]
Torrent, Laura [1 ]
Struis, Rudolf P. W. J. [1 ,2 ]
Ludwig, Christian [1 ,2 ]
机构
[1] Paul Scherrer Inst, Lab Bioenergy & Catalysis, Chem Proc & Mat Res Grp, Forschungsstr 111, CH-5232 Villagen, Switzerland
[2] Ecole Polytech Fed Lausanne EPFL, ENAC IIE GR LUD, CH-1015 Lausanne, Switzerland
关键词
Rare earths; E-waste; Inductively coupled plasma optical emission; spectroscopy; Spectral interferences; Method development; MUTUAL SPECTRAL INTERFERENCES; FLUORESCENT LAMP WASTE; MASS-SPECTROMETRY; STATISTICAL EVALUATION; NDFEB MAGNETS; RECOVERY; LIQUID; DISSOLUTION; EXTRACTION; SEPARATION;
D O I
10.1016/j.sab.2022.106399
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The advent of rare earth elements (REEs) with optoelectronic properties has shifted the technology paradigm from digital to a smart and hybrid world. Their substantial uses also resulted in a large piling up of e-waste. Therefore, e-waste is now a lucrative recycling target for the recovery of such critical raw materials. Their recycling from e-waste is often challenged by dilute metal concentration, complex composition, and difficult chemical characterisation. Generally, the characterisation of e-waste involves elemental determination techniques, such as inductively coupled plasma optical emission spectroscopy (ICP-OES) or inductively coupled plasma mass spectrometry (ICP-MS). ICP-OES is attractive for a recycling or research sector because it has a higher matrix tolerance and lower cost than ICP-MS. In this work, the intensity at 445 line positions measured by an ICP-OES instrument was compiled in a 2D diagram to map interferences by 27 prominent lines from 9 REEs. The second diagram shows the impact at 230 neighbouring line positions measured in each of, in total, 17 (i.e., 9 REEs and 8 non-REEs) single-standard solutions in terms of the concentration of the element type affected. The spectral interference correction algorithm proposed here had been developed by us for a recycling process to obtain pure Y, Eu, and Tb from fluorescent powder (FP) in spent lamps. The ICP-OES analysis and spectral interference correction approach presented here can be applied to any element and e-waste type. To underline this, the paper gives examples for elements in dissolved FP and surrogate NdFeB magnet samples.
引用
收藏
页数:11
相关论文
共 30 条
  • [1] Determination of Rare Earth Elements in Orchards Soil by Inductively Coupled Plasma-Atomic Emission Spectroscopy
    Xue, Jun
    Zhong, Lin-Sheng
    Fan, Yu-Lan
    Li, Xun
    ASIAN JOURNAL OF CHEMISTRY, 2013, 25 (02) : 752 - 754
  • [2] Determination of trace rare earth elements in fruits by microwave digestion coupled with inductively coupled plasma optical emission spectrometry
    Li, Fu-Kai
    Gong, Ai-Jun
    Qiu, Li-Na
    Zhang, Wei-Wei
    Li, Jing-Rui
    Liu, Yu
    Li, Jian-Di
    Gao, Ge
    Yuan, Xiao-Tao
    MICROCHEMICAL JOURNAL, 2019, 147 : 93 - 101
  • [3] Improved inductively coupled plasma optical emission spectroscopy analysis of trace elements in complex matrices by chemometric resolution
    Li, Pao
    Cai, Wensheng
    Shao, Xueguang
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (04) : 936 - 940
  • [4] Spectral interferences in the determination of traces of scandium, yttrium and rare earth elements in 'pure' rare earth matrices by inductively coupled plasma atomic emission spectrometry: Part VI - Ytterbium
    Aleksieva, L
    Daskalova, N
    Velichkov, S
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (08) : 1339 - 1350
  • [5] Spectral interferences in the determination of traces of scandium, yttrium and rare earth elements in ''pure'' rare earth matrices by inductively coupled plasma atomic emission spectrometry .3. Europium
    Daskalova, N
    Velichkov, S
    Slavova, P
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (07) : 733 - 768
  • [6] Spectral interferences in the determination of traces of scandium, yttrium and rare earth elements in 'pure' rare earth matrices by inductively coupled plasma atomic emission spectrometry. Part V - gadolinium and erbium
    Kostadinova, E
    Aleksieva, L
    Velichkov, S
    Daskalova, N
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2000, 55 (06) : 689 - 729
  • [7] Spectral interferences of rare earth elements observed with a high resolution inductively coupled plasma atomic emission spectrometer V. Spectral interferences of lutetium, thulium, yttrium and ytterbium as matrices
    Sun, ZH
    Gu, S
    Sun, DH
    Wang, XR
    Li, B
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2000, 20 (02) : 222 - 228
  • [8] Spectral interferences of rare earth elements observed with a high resolution inductively coupled plasma atomic emission spectrometer - VI. Spectral interferences of lanthanum, terbium, holmium and erbium as matrices
    Sun, ZH
    Li, B
    Sun, DH
    Zhang, ZG
    Ying, H
    Gu, S
    Wang, XR
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2001, 21 (01) : 57 - 61
  • [9] Cloud point extraction with/without chelating agent on-line coupled with inductively coupled plasma optical emission spectrometry for the determination of trace rare earth elements in biological samples
    Li, Yingjie
    Hu, Bin
    JOURNAL OF HAZARDOUS MATERIALS, 2010, 174 (1-3) : 534 - 540
  • [10] Spectral interferences in the determination of traces of scandium, yttrium and rare earth elements in "pure" rare earth matrices by inductively coupled plasma atomic emission spectrometry - Part VII - Terbium, dysprosium, holmium and thulium
    Kolibarska, I.
    Velichkov, S.
    Daskalova, N.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2008, 63 (05) : 603 - 606