Continuous Monitoring of Blood Pressure Using a Wrist-Worn Cuffless Device

被引:17
作者
Sayer, Gabriel [1 ]
Piper, Greta [2 ]
Vorovich, Esther [3 ]
Raikhelkar, Jayant [1 ]
Kim, Gene H. [4 ]
Rodgers, Daniel [4 ]
Shimbo, Daichi [1 ]
Uriel, Nir [1 ]
机构
[1] Columbia Univ, Irving Med Ctr, Dept Med, New York, NY 10027 USA
[2] NYU Langone Hlth, Dept Surg, New York, NY USA
[3] Northwestern Univ, Div Cardiol, Chicago, IL 60611 USA
[4] Univ Chicago, Med Ctr, Sect Cardiol, Dept Med, Chicago, IL 60637 USA
关键词
blood pressure; blood pressure measurement; monitoring; cardiovascular disease; hypertension; remote monitoring; TASK-FORCE;
D O I
10.1093/ajh/hpac020
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
BACKGROUND Measuring an accurate blood pressure (BP) in ambulatory patients is critical for therapeutic decisions. An accurate noninvasive device that measures BP continuously could provide a more comprehensive assessment of BP. The aim of this study was to determine whether the LiveOne device, a cuffless wristband that provides continuous BP measurements, is accurate in comparison to an invasive arterial line (A-line). METHODS We enrolled hospitalized patients and healthy volunteers in this multicenter study. All patients had an invasive A-line. Waveforms were simultaneously downloaded from the A-line and LiveOne device. The primary outcome was the correlation between the LiveOne device and the A-line for systolic BP (SBP) and diastolic BP (DBP). RESULTS Thirty-four patients with 233 measurement periods were included. Mean age was 60.7 +/- 15.2 years, and 15 (44%) were female. The correlation for SBP was 0.91 and for DBP was 0.85. The mean band error was 0.0 +/- 6.9 mm Hg for SBP and 1.2 +/- 5.7 mm Hg for DBP. The mean absolute error was 8.2 +/- 5.8 mm Hg for SBP and 6.4 +/- 3.9 mm Hg for DBP. For SBP, 98% of LiveOne measurements were within 15 mm Hg and for DBP, 92% of LiveOne measurements were within 10 mm Hg. CONCLUSIONS The LiveOne device provides continuous, noninvasive BP measurements that are accurate in comparison to A-line measurements. The portability and unobtrusive nature of this device and the ability to provide continuous BP measurements may offer advantages over currently available BP monitors.
引用
收藏
页码:407 / 413
页数:7
相关论文
共 15 条
[1]   Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial [J].
Abraham, William T. ;
Adamson, Philip B. ;
Bourge, Robert C. ;
Aaron, Mark F. ;
Costanzo, Maria Rosa ;
Stevenson, Lynne W. ;
Strickland, Warren ;
Neelagaru, Suresh ;
Raval, Nirav ;
Krueger, Steven ;
Weiner, Stanislav ;
Shavelle, David ;
Jeffries, Bradley ;
Yadav, Jay S. .
LANCET, 2011, 377 (9766) :658-666
[2]  
American National Standard, 8106022013 ANSIAAMII
[3]   The importance of using 24-hour and nighttime blood pressure for the identification of white coat hypertension: Data from the Jackson Heart Study [J].
Anstey, D. Edmund ;
Colantonio, Lisandro D. ;
Yano, Yuichiro ;
Booth, John N. ;
Muntner, Paul .
JOURNAL OF CLINICAL HYPERTENSION, 2018, 20 (08) :1176-1182
[4]   Comparison of 24-hour Holter Monitoring with 14-day Novel Adhesive Patch Electrocardiographic Monitoring [J].
Barrett, Paddy M. ;
Komatireddy, Ravi ;
Haaser, Sharon ;
Topol, Sarah ;
Sheard, Judith ;
Encinas, Jackie ;
Fought, Angela J. ;
Topol, Eric J. .
AMERICAN JOURNAL OF MEDICINE, 2014, 127 (01) :95.e11-95.e17
[5]   Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010 [J].
Bilo, Grzegorz ;
Zorzi, Cristina ;
Munera, Juan E. Ochoa ;
Torlasco, Camilla ;
Giuli, Valentina ;
Parati, Gianfranco .
BLOOD PRESSURE MONITORING, 2015, 20 (05) :291-294
[6]   A novel cuffless device for self-measurement of blood pressure: concept, performance and clinical validation [J].
Boubouchairopoulou, N. ;
Kollias, A. ;
Chiu, B. ;
Chen, B. ;
Lagou, S. ;
Anestis, P. ;
Stergiou, G. S. .
JOURNAL OF HUMAN HYPERTENSION, 2017, 31 (07) :479-482
[7]   Daytime and nighttime blood pressure as predictors of death and cause-specific cardiovascular events in hypertension [J].
Fagard, Robert H. ;
Celis, Hilde ;
Thijs, Lutgarde ;
Staessen, Jan A. ;
Clement, Denis L. ;
De Buyzere, Marc L. ;
De Bacquer, Dirk A. .
HYPERTENSION, 2008, 51 (01) :55-61
[8]  
GBD 2017 Risk Factor Collaborators, 2019, LANCET, V393, P132
[9]   Current and Future Use of Insertable Cardiac Monitors [J].
Giancaterino, Shaun ;
Lupercio, Florentino ;
Nishimura, Marin ;
Hsu, Jonathan C. .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (11) :1383-1396
[10]   Large-Scale Assessment of a Smartwatch to Identify Atrial Fibrillation [J].
Perez, Marco V. ;
Mahaffey, Kenneth W. ;
Hedlin, Haley ;
Rumsfeld, John S. ;
Garcia, Ariadna ;
Ferris, Todd ;
Balasubramanian, Vidhya ;
Russo, Andrea M. ;
Rajmane, Amol ;
Cheung, Lauren ;
Hung, Grace ;
Lee, Justin ;
Kowey, Peter ;
Talati, Nisha ;
Nag, Divya ;
Gummidipundi, Santosh E. ;
Beatty, Alexis ;
Hills, Mellanie True ;
Desai, Sumbul ;
Granger, Christopher B. ;
Desai, Manisha ;
Turakhia, Mintu P. .
NEW ENGLAND JOURNAL OF MEDICINE, 2019, 381 (20) :1909-1917