Studies on corrosion resistance and bio-activity of plasma spray deposited hydroxylapatite (HA) based TiO2 and ZrO2 dispersed composite coatings on titanium alloy (Ti-6Al-4V) and the same after post spray heat treatment

被引:52
作者
Kumari, Renu [1 ,2 ]
Majumdar, Jyotsna Dutta [1 ]
机构
[1] IIT, Dept Met & Maters Engn, Kharagpur 721302, W Bengal, India
[2] NIFFT, Dept Met & Maters Engn, Ranchi, Jharkhand, India
关键词
Plasma spraying; Ti-6Al-4V; HA; TiO2; ZrO2; Corrosion; Bioactivity; YTTRIA-STABILIZED ZIRCONIA; IN-VITRO BIOACTIVITY; HYDROXYAPATITE COATINGS; MECHANICAL-PROPERTIES; BOND COAT; MICROSTRUCTURE; BEHAVIOR; BIOCOMPATIBILITY; LAYER;
D O I
10.1016/j.apsusc.2017.05.208
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
In the present study, the effect of plasma spray deposited hydroxylapatite (HA) based TiO2 dispersed (HA + 50 wt.% TiO2), coating and post spray heat treatment to be referred as HA-TiO2 (heat treated at 650 degrees C for 2 h) and ZrO2 dispersed (HA + 10 wt.% ZrO2), to be referred as HA-ZrO2 coating (heat treated at 750 degrees C for 2 h) on corrosion resistance and bioactivity of Ti-6Al-4V substrate has been undertaken. There is partial decomposition of HA to tri-calcium-phosphate (Ca-3(PO4)(2)) and formation of CaTiO3 phase in HA-TiO2 coating and CaZrO3 phase in the HA-ZrO2 coating. Corrosion study in Hank's solution shows that there is shifting of corrosion potential (E-corr) towards active potential (-1.1 V(SCE) for as-sprayed and post spray heat treated HA-TiO2 coating, -1.1 V(SCE) for as-sprayed HA-ZrO2 coating and -1 V(SCE) for HA-ZO(2) coating after post spray heat treatment), and deterioration in pitting corrosion (E-pit) resistance in as-sprayed coatings and the same after heat treatment (-0.7 V(SCE) for both HA-TiO2 and HA-ZrO2 coating as compared to as received substrate (-0.3 V(SCE)). The corrosion rate was increased for both the coatings with a maximum increase in HA-ZrO2 coating. Bioactivity test shows a higher degree of apatite deposition in as-sprayed coating and the same after heat treatment as compared to as received Ti-6Al-4V though the as-sprayed one showed a superior behavior. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:935 / 943
页数:9
相关论文
共 38 条
[1]  
Adamson W., 1990, PHYS CHEM SURFACE
[2]   TISSUE-REACTIONS AROUND A HYDROXYAPATITE-COATED HIP-PROSTHESIS - CASE-REPORT OF A RETRIEVED SPECIMEN [J].
BUMA, P ;
GARDENIERS, JWM .
JOURNAL OF ARTHROPLASTY, 1995, 10 (03) :389-395
[3]   Corrosion behaviour of plasma sprayed coatings [J].
Celik, E ;
Ozdemir, I ;
Avci, E ;
Tsunekawa, Y .
SURFACE & COATINGS TECHNOLOGY, 2005, 193 (1-3) :297-302
[4]   Plasma spraying of zirconia-reinforced hydroxyapatite composite coatings on titanium .1. Phase, microstructure and bonding strength [J].
Chang, E ;
Chang, WJ ;
Wang, BC ;
Yang, CY .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1997, 8 (04) :193-200
[5]   Plasma spraying of zirconia-reinforced hydroxyapatite composite coatings on titanium .2. Dissolution behaviour in simulated body fluid and bonding degradation [J].
Chang, E ;
Chang, WJ ;
Wang, BC ;
Yang, CY .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1997, 8 (04) :201-211
[6]   Effect of heat treatment on characteristics of plasma sprayed hydroxyapatite coatings [J].
Chen, Chun-Cheng ;
Ding, Shinn-Jyh .
MATERIALS TRANSACTIONS, 2006, 47 (03) :935-940
[7]   Plasma-sprayed zirconia bond coat as an intermediate layer for hydroxyapatite coating on titanium alloy substrate [J].
Chou, BY ;
Chang, E .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2002, 13 (06) :589-595
[8]   Effects of hydroxylapatite coating crystallinity on biosolubility, cell attachment efficiency and proliferation in vitro [J].
Chou, L ;
Marek, B ;
Wagner, WR .
BIOMATERIALS, 1999, 20 (10) :977-985
[9]   THE EFFECT OF CALCIUM-PHOSPHATE CERAMIC COMPOSITION AND STRUCTURE ON INVITRO BEHAVIOR .1. DISSOLUTION [J].
DUCHEYNE, P ;
RADIN, S ;
KING, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (01) :25-34
[10]  
Fu L, 2002, J AM CERAM SOC, V85, P800