Kinetic parameter estimation from SPECT cone-beam projection measurements

被引:48
作者
Huesman, RH [1 ]
Reutter, BW
Zeng, GL
Gullberg, GT
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Lab, Ctr Funct Imaging, Berkeley, CA 94720 USA
[2] Univ Utah, Dept Radiol, Salt Lake City, UT 84132 USA
关键词
D O I
10.1088/0031-9155/43/4/024
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Kinetic parameters are commonly estimated from dynamically acquired nuclear medicine data, by first reconstructing a dynamic sequence of images and subsequently fitting the parameters to time-activity curves generated from regions of interest overlaid upon the image sequence. Biased estimates can result from images reconstructed using inconsistent projections of a time-varying distribution of radiopharmaceutical acquired by a rotating SPECT system. If the SPECT data are acquired using cone-beam collimators wherein the gantry rotates so that the focal point of the collimators always remains in a plane, additional biases can arise from images reconstructed using insufficient, as well as truncated, projection samples. To overcome these problems we have investigated the estimation of kinetic parameters directly from SPECT cone-beam projection data by modelling the data acquisition process. To accomplish this it was necessary to parametrize the spatial and temporal distribution of the radiopharmaceutical within the SPECT field of view. In a simulated chest image volume, kinetic parameters were estimated for simple one-compartment models for four myocardial regions of interest. Myocardial uptake and washout parameters estimated by conventional analysis of noiseless simulated cone-beam data had biases ranging between 3-26% and 0-28%, respectively. Parameters estimated directly from the noiseless projection data were unbiased as expected, since the model used for fitting was faithful to the simulation. Statistical uncertainties of parameter estimates for 10 000 000 events ranged between 0.2-9% for the uptake parameters and between 0.3-6% for the washout parameters.
引用
收藏
页码:973 / 982
页数:10
相关论文
共 14 条
[1]  
[Anonymous], 1988, Nonlinear regression analysis and its applications
[3]   MODEL-BASED ESTIMATION FOR DYNAMIC CARDIAC STUDIES USING ECT [J].
CHIAO, PC ;
ROGERS, WL ;
CLINTHORNE, NH ;
FESSLER, JA ;
HERO, AO .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (02) :217-226
[4]   MODEL-BASED ESTIMATION WITH BOUNDARY SIDE INFORMATION OR BOUNDARY REGULARIZATION [J].
CHIAO, PC ;
ROGERS, WL ;
FESSLER, JA ;
CLINTHORNE, NH ;
HERO, AO .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (02) :227-234
[5]   IMPLEMENTATION OF 3-DIMENSIONAL IMAGE-RECONSTRUCTION FOR MULTIRING POSITRON TOMOGRAPHS [J].
DEFRISE, M ;
TOWNSEND, D ;
GEISSBUHLER, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1361-1372
[6]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[7]   LEAST-SQUARES ALGORITHM FOR REGION-OF-INTEREST EVALUATION IN EMISSION TOMOGRAPHY [J].
FORMICONI, AR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (01) :90-100
[8]   A NEW FAST ALGORITHM FOR THE EVALUATION OF REGIONS OF INTEREST AND STATISTICAL UNCERTAINTY IN COMPUTED-TOMOGRAPHY [J].
HUESMAN, RH .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (05) :543-552
[9]  
HUESMAN RH, 1997, J NUCL MED S5, V38, P222
[10]   DIRECT RECONSTRUCTION OF FUNCTIONAL PARAMETERS FOR DYNAMIC SPECT [J].
LIMBER, MA ;
LIMBER, MN ;
CELLER, A ;
BARNEY, JS ;
BORWEIN, JM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1249-1256