Portable ozone calibration source independent of changes in temperature, pressure and humidity for research and regulatory applications

被引:2
作者
Birks, John W. [1 ]
Williford, Craig J. [1 ]
Andersen, Peter C. [1 ]
Turnipseed, Andrew A. [1 ]
Strunk, Stanley [1 ]
Ennis, Christine A. [1 ]
机构
[1] 2B Technol Inc, 2100 Cent Ave, Boulder, CO 80301 USA
关键词
INDUCED FLUORESCENCE SPECTROSCOPY; LASER-INDUCED FLUORESCENCE; ABSORPTION CROSS-SECTIONS; TROPOSPHERIC OH RADICALS; POPCORN FIELD CAMPAIGN; WATER-VAPOR; UV-ABSORBENCY; INTERFERENCES; SESQUITERPENES; RESOLUTION;
D O I
10.5194/amt-11-4797-2018
中图分类号
P4 [大气科学(气象学)];
学科分类号
0706 ; 070601 ;
摘要
A highly portable ozone (O-3) calibration source that can serve as a U.S. EPA level 4 transfer standard for the calibration of ozone analyzers is described and evaluated with respect to analytical figures of merit and effects of ambient pressure and humidity. Reproducible mixing ratios of ozone are produced by the photolysis of oxygen in O-3-scrubbed ambient air by UV light at 184.9 nm light from a low-pressure mercury lamp. By maintaining a constant volumetric flow rate (thus constant residence time within the photolysis chamber), the mixing ratio produced is independent of both pressure and temperature and can be varied by varying the lamp intensity. Pulse width modulation of the lamp with feedback from a photodiode monitoring the 253.7 nm emission line is used to maintain target ozone mixing ratios in the range 30-1000 ppb. In order to provide a constant ratio of intensities at 253.7 and 184.9 nm, the photolysis chamber containing the lamp is regulated at a temperature of 40 degrees C. The resulting O-3 calibrator has a response time for step changes in output ozone mixing ratio of < 30 s and precision (sigma(p)) of 0.4 % of the output mixing ratio for 10 s measurements (e.g., sigma(p) = +/- 0.4 ppb for 100 ppb of O-3). Ambient humidity was found to affect the output mixing ratio of ozone primarily by dilution of the oxygen precursor. This potential humidity interference could be up to a few percent in extreme cases but is effectively removed by varying the lamp intensity to compensate for the reduced oxygen concentration based on feedback from a humidity sensor.
引用
收藏
页码:4797 / 4807
页数:11
相关论文
共 37 条
[1]   THE PHOTOCHEMISTRY OF ATMOSPHERIC WATER VAPOR [J].
BATES, DR ;
NICOLET, M .
JOURNAL OF GEOPHYSICAL RESEARCH, 1950, 55 (03) :301-327
[2]  
Beard M. E., 1977, EPA600477005 OFF RES
[3]  
BIRKS JW, 1998, TOP ENVIRONM CHEM, P233
[4]  
Buckley TJ, 2003, 6369 NISTIR
[5]  
Burkholder J.B., 2015, JPL PUBLICATION, V15-10, DOI DOI 10.13140/RG.2.1.2504.2806
[6]   Absorption cross sections for water vapor from 183 to 193 nm [J].
Cantrell, CA ;
Zimmer, A ;
Tyndall, GS .
GEOPHYSICAL RESEARCH LETTERS, 1997, 24 (17) :2195-2198
[7]   Absorption cross-section measurements of water vapour and oxygen at 185 nm.: Implications for the calibration of field instruments to measure OH, HO2 and RO2 radicals [J].
Creasey, DJ ;
Heard, DE ;
Lee, JD .
GEOPHYSICAL RESEARCH LETTERS, 2000, 27 (11) :1651-1654
[8]  
Finlayson-Pitts B.J., 2000, Chemistry of the Upper and Lower Atmosphere: Theory, Experiments, and Applications, Vfirst
[9]  
Gotz F.W. P., 1934, Proceedings of the Royal Society of London. Series A, V145, P416, DOI DOI 10.1098/RSPA.1934.0109
[10]  
GUINVARC'H PIERRE, 1959, ADVANCES IN CHEM SER, V21, P416