Proximal Fiber Tip Damage during Holmium: YAG and Thulium Fiber Laser Ablation of Kidney Stones

被引:2
作者
Wilson, Christopher R. [1 ]
Hardy, Luke A. [1 ]
Irby, Pierce B. [2 ]
Fried, Nathaniel M. [1 ,2 ]
机构
[1] Univ N Carolina, Dept Phys & Opt Sci, Charlotte, NC 28223 USA
[2] Carolinas Med Ctr, McKay Dept Urol, Charlotte, NC 28203 USA
来源
PHOTONIC THERAPEUTICS AND DIAGNOSTICS XII | 2016年 / 9689卷
关键词
ablation; damage; fiber; holmium; kidney stones; lithotripsy; proximal; thulium; OPTICAL-FIBERS; LITHOTRIPSY;
D O I
10.1117/12.2207895
中图分类号
TH742 [显微镜];
学科分类号
摘要
The Thulium fiber laser (TFL) is being studied as an alternative to Holmium: YAG laser for lithotripsy. TFL beam originates within an 18-mu m-core thulium doped silica fiber, and its near single mode, Gaussian beam profile enables transmission of higher laser power through smaller fibers than possible during Holmium laser lithotripsy. This study examines whether TFL beam profile also reduces proximal fiber tip damage compared to Holmium laser multimodal beam. TFL beam at wavelength of 1908 nm was coupled into 105-mu m-core silica fibers, with 35-mJ energy, 500-mu s pulse duration, and pulse rates of 50-500 Hz. For each pulse rate, 500,000 pulses were delivered. Magnified images of proximal fiber surfaces were taken before and after each trial. For comparison, 20 single-use, 270-mu m-core fibers were collected after clinical Holmium laser lithotripsy procedures using standard settings (600 mJ, 350 mu s, 6 Hz). Total laser energy, number of laser pulses, and laser irradiation time were recorded, and fibers were rated for damage. For TFL studies, output power was stable, and no proximal fiber damage was observed after delivery of 500,000 pulses at settings up to 35 mJ, 500 Hz, and 17.5 W average power. In contrast, confocal microscopy images of fiber tips after Holmium lithotripsy showed proximal fiber tip degradation in all 20 fibers. The proximal fiber tip of a 105-mu m-core fiber transmitted 17.5 W of TFL power without degradation, compared to degradation of 270-mu m-core fibers after transmission of 3.6 W of Holmium laser power. The smaller and more uniform TFL beam profile may improve fiber lifetime, and potentially reduce costs for the surgical disposables as well.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Fiber-optic manipulation of urinary stone phantoms using holmium:YAG and thulium fiber lasers
    Blackmon, Richard L.
    Case, Jason R.
    Trammell, Susan R.
    Irby, Pierce B.
    Fried, Nathaniel M.
    JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (02)
  • [32] A Tapered Distal Fiber Tip for Thulium Fiber Laser Lithotripsy
    Blackmon, Richard L.
    Irby, Pierce B.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS VI, 2010, 7548
  • [33] Navigating urolithiasis treatment: assessing the practicality and performance of thulium fiber laser, holmium YAG, and thulium YAG in real-world scenarios
    Catalina Solano
    Mariela Corrales
    Frederic Panthier
    Luigi Candela
    Steeve Doizi
    Olivier Traxer
    World Journal of Urology, 2023, 41 : 2627 - 2636
  • [34] Navigating urolithiasis treatment: assessing the practicality and performance of thulium fiber laser, holmium YAG, and thulium YAG in real-world scenarios
    Solano, Catalina
    Corrales, Mariela
    Panthier, Frederic
    Candela, Luigi
    Doizi, Steeve
    Traxer, Olivier
    WORLD JOURNAL OF UROLOGY, 2023, 41 (10) : 2627 - 2636
  • [35] Understanding the ablation rate of Holmium:YAG and thulium fiber lasers. Perspectives from an in vitro study
    Basulto-Martinez, Mario
    Proietti, Silvia
    Pavia, Maria Pia
    Yeow, Yuyi
    Eisner, Brian H. H.
    Giusti, Guido
    UROLITHIASIS, 2023, 51 (01)
  • [36] Collateral damage to the ureter and Nitinol stone baskets during thulium fiber laser lithotripsy
    Wilson, Christopher R.
    Hardy, Luke A.
    Irby, Pierce B.
    Fried, Nathaniel M.
    LASERS IN SURGERY AND MEDICINE, 2015, 47 (05) : 403 - 410
  • [37] Understanding the ablation rate of Holmium:YAG and thulium fiber lasers. Perspectives from an in vitro study
    Mario Basulto-Martínez
    Silvia Proietti
    Maria Pia Pavia
    Yuyi Yeow
    Brian H. Eisner
    Guido Giusti
    Urolithiasis, 51
  • [38] Assessment of Holmium:YAG, Pulsed-Thulium:YAG and Thulium Fiber Lasers for Urinary Stone Ablation. In Vitro Study
    Sierra, Alba
    Panthier, Frederic
    Castillo, Esther
    Mercade, Asier
    Peri, Lluis
    Alcaraz, Antonio
    Traxer, Olivier
    Lopez, Juan Manuel
    Luque, Maria Pilar
    JOURNAL OF ENDOUROLOGY, 2024, 38 (11) : 1209 - 1216
  • [39] Comparing dusting and fragmenting efficiency using the new SuperPulsed thulium fiber laser versus a 120 W Holmium:YAG laser
    Chew, Ben H.
    Koo, Kyo Chul
    Halawani, Abdulghafour
    Lundeen, Colin J.
    Knudsen, Bodo E.
    Molina, Wilson R.
    INVESTIGATIVE AND CLINICAL UROLOGY, 2023, 64 (03) : 265 - 271
  • [40] Thermal Injury and Laser Efficiency with Holmium YAG and Thulium Fiber Laser-An In Vitro Study
    Sierra, Alba
    Corrales, Mariela
    Kolvatzis, Merkourios
    Panthier, Frederic
    Pinero, Adria
    Traxer, Olivier
    JOURNAL OF ENDOUROLOGY, 2022, 36 (12) : 1599 - 1606