Kalman-Based Calibration Algorithm for AgaMatrix Continuous Glucose Monitoring System

被引:6
作者
Singh, Abhinoy Kumar [1 ,2 ]
Rebec, Mihailo V. [3 ]
Haidar, Ahmad [1 ]
机构
[1] McGill Univ, Dept Biomed Engn, Montreal, PQ H3A 2B4, Canada
[2] IIT Indore, Dept Elect Engn, Indore 453552, India
[3] WaveForm Technol Inc, Wilsonville, OR 97070 USA
关键词
Glucose; Calibration; Kalman filters; Blood; Noise measurement; Current measurement; Estimation; Continuous glucose monitoring; glucose sensor; Kalman filter; parameter estimation; INTEGRAL EQUATIONS; DECONVOLUTION; ACCURACY; SENSORS; STATE;
D O I
10.1109/TCST.2020.3003450
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A continuous glucose monitoring system is composed of a glucose sensor and an estimation (calibration) algorithm. The glucose sensor is composed of an electrode that is inserted under the skin and generates a noisy electrical current in response to interstitial glucose levels. The relationship between the electrical current and the interstitial glucose levels varies between individuals and within the same individual (with sensor wear time). The estimation algorithm infers the glucose levels in the blood from the noisy electrical current signal and uses intermittent capillary glucose measurements to account for intra-and inter-individual variability. In this article, we propose a novel real-time Kalman filter-based estimation algorithm that is composed of three steps: 1) noise filtering step; 2) compartment matching step using sequential Kalman filter; and 3) parameters estimation step using the conventional Kalman filter. The initial estimate of the parameters and covariance matrix are extracted offline using a nonlinear cubature Kalman filter. Our algorithm is compared with four alternative algorithms using 20 sensor data sets. Each data set was generated over a seven-day sensor wear period, during which patients were tested on three in-clinic days (days 1, 4, and 7). The comparison is based on the mean absolute relative difference (MARD) between frequent reference glucose measurements and the sensor glucose levels. MARD for days 1, 4, and 7 were 10.3%, 10.7%, and 11.9% for our algorithm and 12.46%, 15.75%, and 14.17% for the alternative algorithm, respectively. Our algorithm is currently being integrated into a commercial continuous glucose monitoring system (AgaMatrix, Inc., Salem, NH, USA).
引用
收藏
页码:1257 / 1267
页数:11
相关论文
共 50 条
  • [21] An Interstitial Fluid Transdermal Extraction System for Continuous Glucose Monitoring
    Yu, Haixia
    Li, Dachao
    Roberts, Robert C.
    Xu, Kexin
    Tien, Norman C.
    JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2012, 21 (04) : 917 - 925
  • [22] Non-invasive impedance based continuous glucose monitoring system
    Talary, M. S.
    Dewarrat, F.
    Huber, D.
    Falco-Jonasson, L.
    Caduff, A.
    13TH INTERNATIONAL CONFERENCE ON ELECTRICAL BIOIMPEDANCE AND THE 8TH CONFERENCE ON ELECTRICAL IMPEDANCE TOMOGRAPHY 2007, 2007, 17 : 636 - +
  • [23] Improving the Accuracy of Pdot-Based Continuous Glucose Monitoring by Using External Ratiometric Calibration
    Sun, Kai
    Liu, Siyang
    Liu, Jing
    Ding, Zhaoyang
    Jiang, Yifei
    Zhang, Jicheng
    Chen, Haobin
    Yu, Jiangbo
    Wu, Changfeng
    Chiu, Daniel T.
    ANALYTICAL CHEMISTRY, 2021, 93 (04) : 2359 - 2366
  • [24] Cell based metabolic barriers to glucose diffusion: Macrophages and continuous glucose monitoring
    Klueh, Ulrike
    Frailey, Jackman T.
    Qiao, Yi
    Antar, Omar
    Kreutzera, Donald L.
    BIOMATERIALS, 2014, 35 (10) : 3145 - 3153
  • [25] A method for continuous glucose monitoring in interstitial fluid based on microdialysis
    Li, D. (dchli@tju.edu.cn), 1600, Tianjin University (46): : 482 - 486
  • [26] Toward a Continuous Intravascular Glucose Monitoring System
    Beier, Brooke
    Musick, Katherine
    Matsumoto, Akira
    Panitch, Alyssa
    Nauman, Eric
    Irazoqui, Pedro
    SENSORS, 2011, 11 (01) : 409 - 424
  • [27] Measurement delay associated with the Guardian® RT continuous glucose monitoring system
    Wei, C.
    Lunn, D. J.
    Acerini, C. L.
    Allen, J. M.
    Larsen, A. M.
    Wilinska, M. E.
    Dunger, D. B.
    Hovorka, R.
    DIABETIC MEDICINE, 2010, 27 (01) : 117 - 122
  • [28] A novel glucose sensor system with Au nanoparticles based on microdialysis and coenzymes for continuous glucose monitoring
    Pan, M
    Guo, XS
    Cai, Q
    Li, G
    Chen, YQ
    SENSORS AND ACTUATORS A-PHYSICAL, 2003, 108 (1-3) : 258 - 262
  • [29] Criteria for Personalised Choice of a Continuous Glucose Monitoring System: An Expert Opinion
    Di Molfetta, Sergio
    Rossi, Antonio
    Boscari, Federico
    Irace, Concetta
    Laviola, Luigi
    Bruttomesso, Daniela
    DIABETES THERAPY, 2024, : 2263 - 2278
  • [30] Design of a Real-time Self-adjusting Calibration Algorithm to Improve the Accuracy of Continuous Blood Glucose Monitoring
    Ziru Jia
    Lijuan Huang
    Hongying Liu
    Yonghong Huang
    Wang Li
    Xitian Pi
    Xiaolin Zheng
    Applied Biochemistry and Biotechnology, 2020, 190 : 1163 - 1176