Comparison of different biomaterials for glaucoma drainage devices - Part 2

被引:91
作者
Ayyala, RS
Michelini-Norris, B
Flores, A
Haller, E
Margo, CE
机构
[1] Tulane Univ, Sch Med, Dept Ophthalmol, Med Ctr, New Orleans, LA 70112 USA
[2] Med Ctr Louisiana, Dept Ophthalmol, New Orleans, LA USA
[3] Univ S Florida, Dept Ophthalmol, Tampa, FL USA
[4] Univ S Florida, Dept Pathol, Tampa, FL USA
[5] Watson Clin, Dept Ophthalmol, Lakeland, FL USA
关键词
D O I
10.1001/archopht.118.8.1081
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: Inflammation associated with biomaterials may contribute to the failure of glaucoma drainage devices. Objective: To compare the inflammatory reaction associated with the insertion of Krupin silicone, Molteno polypropylene, and Acrosof end plates in the subconjunctival space of rabbits. Methods: Similar-sized glaucoma end plates made of 3 different biomaterials were sutured to the sclera in the superotemporal quadrant of the rabbit eye. Thirty eyes of 15 albino New Zealand rabbits were randomly assigned to the 3 groups. Conjunctival vascular hyperemia was graded in a masked fashion among the 3 groups. At the end of 3 weeks, the enucleated eyes were examined histologically and by scanning electron microscopy. Results: Molteno polypropylene was associated with more inflammation both in clinical observations and based on histological grading. Silicone and Acrosof were associated with less intense inflammation. One polypropylene end plate was extruded on day 21. Conclusions: Polypropylene appears to be more inflammatory than silicone. Flexible biomaterials appear to be less inflammatory than rigid ones. Clinical Relevance: Bleb failure following glaucoma drainage device implantation could be related to the biomaterial-associated inflammation. Choosing a biomaterial with the least inflammatory potential might enhance the success rate of the glaucoma drainage device.
引用
收藏
页码:1081 / 1084
页数:4
相关论文
共 6 条
[1]   A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma [J].
Ayyala, RS ;
Zurakowski, D ;
Smith, JA ;
Monshizadeh, R ;
Netland, PA ;
Richards, DW ;
Layden, WE .
OPHTHALMOLOGY, 1998, 105 (10) :1968-1976
[2]  
Ayyala RS, 1999, ARCH OPHTHALMOL-CHIC, V117, P233
[3]   COMPARISON OF HEXAMETHYLDISILAZANE (HMDS), PELDRI-II, AND CRITICAL-POINT DRYING METHODS FOR SCANNING ELECTRON-MICROSCOPY OF BIOLOGICAL SPECIMENS [J].
BRAY, DF ;
BAGU, J ;
KOEGLER, P .
MICROSCOPY RESEARCH AND TECHNIQUE, 1993, 26 (06) :489-495
[4]   Biocompatibility of poly(methyl methacrylate), silicone, and AcrySof intraocular lenses: Randomized comparison of the cellular reaction on the anterior lens surface [J].
Hollick, EJ ;
Spalton, DJ ;
Ursell, PG ;
Pande, MV .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (03) :361-366
[5]  
Molteno AC., 1988, GLAUCOMA P 4 INT S N, P192
[6]   CLINICAL-EXPERIENCE WITH THE BAERVELDT GLAUCOMA DRAINAGE IMPLANT [J].
SIEGNER, SW ;
NETLAND, PA ;
URBAN, RC ;
WILLIAMS, AS ;
RICHARDS, DW ;
LATINA, MA ;
BRANDT, JD .
OPHTHALMOLOGY, 1995, 102 (09) :1298-1307