Comparison of the C-MAC® videolaryngoscope with the Macintosh, Glidescope®, and Airtraq® laryngoscopes in easy and difficult laryngoscopy scenarios in manikins

被引:88
作者
McElwain, J.
Malik, M. A.
Harte, B. H.
Flynn, N. M.
Laffey, J. G. [1 ]
机构
[1] Natl Univ Ireland, Dept Anaesthesia, Galway Univ Hosp, Galway, Ireland
关键词
CERVICAL-SPINE IMMOBILIZATION; CLOSED CLAIMS ANALYSIS; TRACHEAL INTUBATION; NOVICE PERSONNEL; TRUVIEW EVO2(R); CLINICAL-TRIAL; AIRWAY; MANAGEMENT; AIRWAYSCOPE(R); ANESTHETISTS;
D O I
10.1111/j.1365-2044.2010.06307.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
P>The C-MAC (R) comprises a Macintosh blade connected to a video unit. The familiarity of the Macintosh blade, and the ability to use the C-MAC as a direct or indirect laryngoscope, may be advantageous. We wished to compare the C-MAC with Macintosh, Glidescope (R) and Airtraq (R) laryngoscopes in easy and simulated difficult laryngoscopy. Thirty-one experienced anaesthetists performed tracheal intubation in an easy and difficult laryngoscopy scenario. The duration of intubation attempts, success rates, number of intubation attempts and of optimisation manoeuvres, the severity of dental compression, and difficulty of device use were recorded. In easy laryngoscopy, the duration of tracheal intubation attempts were similar with the C-MAC, Macintosh and Airtraq laryngoscopes; the Glidescope performed less well. The C-MAC and Airtraq provided the best glottic views, but the C-MAC was rated as the easiest device to use. In difficult laryngo-scopy the C-MAC demonstrated the shortest tracheal intubation times. The Airtraq provided the best glottic view, with the Macintosh providing the worst view. The C-MAC was the easiest device to use.
引用
收藏
页码:483 / 489
页数:7
相关论文
共 25 条
[1]   ADVERSE RESPIRATORY EVENTS IN ANESTHESIA - A CLOSED CLAIMS ANALYSIS [J].
CAPLAN, RA ;
POSNER, KL ;
WARD, RJ ;
CHENEY, FW .
ANESTHESIOLOGY, 1990, 72 (05) :828-833
[2]   The American Society of Anesthesiologists closed claims project: What have we learned, how has it affected practice, and how will it affect practice in the future? [J].
Cheney, FW .
ANESTHESIOLOGY, 1999, 91 (02) :552-556
[3]   RAPIDITY AND ACCURACY OF TRACHEAL INTUBATION IN A MANNEQUIN - COMPARISON OF THE FIBEROPTIC WITH THE BULLARD LARYNGOSCOPE [J].
DYSON, A ;
HARRIS, J ;
BHATIA, K .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (02) :268-270
[4]   Difficult Airway Society guidelines for management of the unanticipated difficult intubation [J].
Henderson, JJ ;
Popat, MT ;
Latto, IP ;
Pearce, AC .
ANAESTHESIA, 2004, 59 (07) :675-694
[5]   Patient Safety Watching over the medical device industry [J].
Lenzer, Jeanne .
BRITISH MEDICAL JOURNAL, 2009, 338 :b2321
[6]   Comparison of the GlideScope® with the Macintosh laryngoscope for tracheal intubation in patients with simulated difficult airway [J].
Lim, Y ;
Yeo, SW .
ANAESTHESIA AND INTENSIVE CARE, 2005, 33 (02) :243-247
[7]   Evaluation of the Airtraq® and Macintosh laryngoscopes in patients at increased risk for difficult tracheal intubation [J].
Maharaj, C. H. ;
Costello, J. F. ;
Harte, B. H. ;
Laffey, J. G. .
ANAESTHESIA, 2008, 63 (02) :182-188
[8]   A comparison of direct and indirect laryngoscopes and the ILMA in novice users: a manikin study [J].
Maharaj, C. H. ;
McDonnell, J. G. ;
Harte, B. H. ;
Laffey, J. G. .
ANAESTHESIA, 2007, 62 (11) :1161-1166
[9]   The Airtraq® as a rescue airway device following failed direct laryngoscopy:: a case series [J].
Maharaj, C. H. ;
Costello, J. F. ;
McDonnell, J. G. ;
Harte, B. H. ;
Laffey, J. G. .
ANAESTHESIA, 2007, 62 (06) :598-601
[10]   Retention of tracheal intubation skills by novice personnel:: a comparison of the Airtraq® and Macintosh laryngoscopes [J].
Maharaj, C. H. ;
Costello, J. ;
Higgins, B. D. ;
Harte, B. H. ;
Laffey, J. G. .
ANAESTHESIA, 2007, 62 (03) :272-278