摘要
A linearly scanned three-dimensional (3-D) ultrasound imaging system is considered. The transducer array is initially oriented along the a: axis and aimed in the y direction. After being tilted by an angle theta about the x axis, and then swiveled by an angle phi about the y axis, it is translated in the z direction, in steps of size d, to acquire a series of parallel two-dimensional (2-D) images. From these, the 3-D image is reconstructed, using the nominal values of the parameters (phi, theta, d). Thus, any systematic or random errors in these, relative to their actual values (phi(0), theta(0), d(0)), will respectively cause distortions or variances in length, area, and volume in the reconstructed 3-D image, relative to the 3-D object. Here, we analyze these effects, Compact linear approximations are derived for the relative distortions as functions of the parameter errors, and hence, for the relative variances as functions of the parameter variances. Also, exact matrix formulas for the relative distortions are derived for arbitrary values of (phi, theta, d) and (phi(0), theta(0) d(0)). These were numerically compared to the linear approximations and to measurements from simulated 3-D images of a cubical object and real 3-D images of a wire phantom. In every case tested, the theory was confirmed within experimental error (0.5%).
引用
收藏
页码:632 / 651
页数:20
相关论文
共 14 条
[1]  
Baba K., 1997, 3 DIMENSIONAL ULTRAS, V1
[2]   3-DIMENSIONAL AND 4-DIMENSIONAL CARDIOVASCULAR ULTRASOUND IMAGING - A NEW ERA FOR ECHOCARDIOGRAPHY [J].
BELOHLAVEK, M ;
FOLEY, DA ;
GERBER, TC ;
KINTER, TM ;
GREENLEAF, JF ;
SEWARD, JB .
MAYO CLINIC PROCEEDINGS, 1993, 68 (03) :221-240
[3]  
DOWNEY DB, 1995, RADIOLOGY, V197, P539
[4]  
DUCK FA, 1998, ULTRASOUND MED MSS
[5]   3-D ultrasound imaging: A review [J].
Fenster, A ;
Downey, DB .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1996, 15 (06) :41-51
[6]  
FENSTER A, 1995, Patent No. 5454371
[7]   Three-dimensional ultrasound imaging [J].
Nelson, TR ;
Pretorius, DH .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (09) :1243-1270
[8]  
NELSON TR, 1999, 3 DIMENSIONAL ULTRAS
[9]   3-DIMENSIONAL AND 4-DIMENSIONAL ECHOCARDIOGRAPHY [J].
OFILI, EO ;
NANDA, NC .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (08) :669-675
[10]   3-DIMENSIONAL COLOR DOPPLER IMAGING [J].
PICOT, PA ;
RICKEY, DW ;
MITCHELL, R ;
RANKIN, RN ;
FENSTER, A .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (02) :95-104