A single setup approach for the MRI-based measurement and validation of the transfer function of elongated medical implants

被引:3
作者
Stijnman, Peter R. S. [1 ,2 ]
Erturk, M. Arcan [3 ]
van den Berg, Cornelis A. T. [1 ]
Raaijmakers, Alexander J. E. [1 ,2 ]
机构
[1] Ctr Image Sci UMC Utrecht, Computat Imaging Grp MRI Diagnost & Therapy, NL-3584 CX Utrecht, Netherlands
[2] Eindhoven Univ Technol, Dept Biomed Engn, Eindhoven, Brabant, Netherlands
[3] Medtron PLC, Restorat Therapies Grp, Implantables R&D, Minneapolis, MN USA
关键词
measurement; simulation; transfer function; validation; DEEP BRAIN-STIMULATION; SIMULATION;
D O I
10.1002/mrm.28840
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To propose a single setup using the MRI to both measure and validate the transfer function (TF) of linear implants. Conventionally, the TF of an implant is measured in one bench setup and validated using another. Methods It has been shown that the TF can be measured using MRI. To validate this measurement, the implant is exposed to different incident electric fields, while the temperature increase at the tip is monitored. For a good validation, the incident electric fields that the implant is exposed to should be orthogonal. We perform a simulation study on six different methods that change the incident electric field. Afterward, a TF measurement and validation study using the best method from the simulations is performed. This is done with fiberoptic temperature probes at 1.5 T for four linear implant structures using the proposed single setup. Results The simulation study showed that positioning local transmit coils at different locations along the lead trajectory has a similar validation quality compared with changing the implant trajectory (ie, the conventional validation method). For the validation study that was performed, an R-2 >= 0.91 was found for the four investigated leads. Conclusion A single setup to both measure and validate the transfer function using local transmit coils has been shown to work. The benefits of using the proposed validation method are that there is only one setup required instead of two and the implant trajectory is not varied; therefore, the relative distance between the leap tip and the temperature probe is constant.
引用
收藏
页码:2751 / 2765
页数:15
相关论文
共 24 条
[1]   A 12-month prospective study on gasserian stimulation for neuropathic facial pain [J].
Andre, MG ;
Ogrin, M ;
Henderson, J ;
Rosenow, JM .
NEUROSURGERY, 2005, 57 (02) :413-413
[2]  
Arlinghaus S., 1994, Practical Handbook of Curve Fitting
[3]   Measurements and simulation of RF heating of implanted stereo-electroencephalography electrodes during MR scans [J].
Bhusal, Bhumi ;
Bhattacharyya, Pallab ;
Baig, Tanvir ;
Jones, Stephen ;
Martens, Michael .
MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (04) :1676-1685
[4]   Should patients with brain implants undergo MRI? [J].
Erhardt, Johannes B. ;
Fuhrer, Erwin ;
Gruschke, Oliver G. ;
Leupold, Jochen ;
Wapler, Matthias C. ;
Hennig, Juergen ;
Stieglitz, Thomas ;
Korvink, Jan G. .
JOURNAL OF NEURAL ENGINEERING, 2018, 15 (04)
[5]   Reduction of Implant RF Heating Through Modification of Transmit Coil Electric Field [J].
Eryaman, Yigitcan ;
Akin, Burak ;
Atalar, Ergin .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (05) :1305-1313
[6]   Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants [J].
Golestanirad, Laleh ;
Iacono, Maria Ida ;
Keil, Boris ;
Angelone, Leonardo M. ;
Bonmassar, Giorgio ;
Fox, Michael D. ;
Herrington, Todd ;
Adalsteinsson, Elfar ;
LaPierre, Cristen ;
Mareyam, Azma ;
Wald, Lawrence L. .
NEUROIMAGE, 2017, 147 :577-588
[7]  
ISO TS, 2018, 10974 ISO TS
[8]  
Kozlov M, 2019, IEEE ANTENNAS PROP, P765, DOI [10.1109/apusncursinrsm.2019.8888481, 10.1109/APUSNCURSINRSM.2019.8888481]
[9]   Modeling radiofrequency responses of realistic multi-electrode leads containing helical and straight wires [J].
Kozlov, Mikhail ;
Horner, Marc ;
Kainz, Wolfgang .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2020, 33 (03) :421-437
[10]   Investigation of Parallel Radiofrequency Transmission for the Reduction of Heating in Long Conductive Leads in 3 Tesla Magnetic Resonance Imaging [J].
McElcheran, Clare E. ;
Yang, Benson ;
Anderson, Kevan J. T. ;
Golenstani-Rad, Laleh ;
Graham, Simon J. .
PLOS ONE, 2015, 10 (08)