Objective: This study aimed to evaluate a new ion-exchange HPLC system to measure hemoglobin A1c (HbA1c) in comparison to two other widely used HPLC systems. Methods: Analytical performance and hemoglobin variants detection was assessed on the new Arkray/ELITech ADAMS HA-8180V (R), in parallel to BioRad Variant (TM) II Turbo 2.0 (R) and Tosoh Bioscience HLC (R)-723G8. A method comparison and concordance for the classification of patients according to the American Diabetes Association (ADA) categories was performed using kappa test. Results: ADAMS HA-8180V (R) demonstrated excellent within-run imprecision (0%) and between-run imprecision: <= 1.21% using blood sample and <= 0.94% using quality controls. Method comparison of ADAMS HA-8180V (R) with the two other systems yielded very high coefficient correlation (r > 0.995). A very good agreement (kappa value >= 0.81) was observed between methods for the classification of patients according to the ADA categories. In addition, all systems were able to detect the presence of most classical variants such as HbC, HbS, HbD, and HbE. Conclusion: The Arkray/ELITech ADAMS HA-8180V (R) demonstrated high analytical performance similar to previous systems such as Biorad (TM) Variant II Turbo 2.0 (R) and Tosoh Bioscience HLC (R)-723G8. The three systems allow a high-quality HbA1c measurement and appeared interchangeable for diabetes diagnosis or for the therapeutic monitoring of patients without hemoglobin variants. (C) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:428 / 434
页数:7
相关论文
共 19 条
[1]  
[Anonymous], 2010, Diabetes Care, V33 Suppl 1, pS4, DOI 10.2337/dc10-S004
[2]  
[Anonymous], J INT FED CLIN CHEM
[3]   Reporting variant hemoglobins discovered during hemoglobin A1c analysis - Common practices in clinical laboratories [J].
Behan, Kristina Jackson ;
Storey, Natalie M. ;
Lee, Hong-Kee .
CLINICA CHIMICA ACTA, 2009, 406 (1-2) :124-128
[4]   Multicentre evaluation of the Tosoh HbA1c G8 analyser [J].
Chapelle, Jean-Paul ;
Teixeira, Jelda ;
Maisin, Diane ;
Assink, Hans ;
Barla, Gerhard ;
Stroobants, An K. ;
Delzenne, Barend ;
van den Eshof, Wouter .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (03) :365-371
[5]  
Dewitte K, 2002, CLIN CHEM, V48, P799
[6]   The meaning of kappa: Probabilistic concepts of reliability and validity revisited [J].
GuggenmoosHolzmann, I .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 1996, 49 (07) :775-782
[7]   IFCC reference system for measurement of hemoglobin Alc in human blood and the National Standardization Schemes in the United States, Japan, and Sweden:: A method-comparison study [J].
Hoelzel, W ;
Weykamp, C ;
Jeppsson, JO ;
Miedema, K ;
Barr, JR ;
Goodall, I ;
Hoshino, T ;
John, WG ;
Kobold, U ;
Little, R ;
Mosca, A ;
Mauri, P ;
Paroni, R ;
Susanto, F ;
Takei, I ;
Thienpont, L ;
Umemoto, M ;
Wiedmeyer, HM .
CLINICAL CHEMISTRY, 2004, 50 (01) :166-174
[8]   Approved IFCC reference method for the measurement of HbA1c in human blood [J].
Jeppsson, JO ;
Kobold, U ;
Barr, J ;
Finke, A ;
Hoelzel, W ;
Hoshino, T ;
Miedema, K ;
Mosca, A ;
Mauri, P ;
Paroni, R ;
Thienpont, L ;
Umemoto, M ;
Weykamp, C .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2002, 40 (01) :78-89
[9]   Six of Eight Hemoglobin A1c Point-of-Care Instruments Do Not Meet the General Accepted Analytical Performance Criteria [J].
Lenters-Westra, Erna ;
Slingerland, Robbert J. .
CLINICAL CHEMISTRY, 2010, 56 (01) :44-52
[10]   Effects of hemoglobin C, D, E, and S traits on measurements of HbA1c by six methods [J].
Lin, Chia-Ni ;
Emery, Todd J. ;
Little, Randie R. ;
Hanson, Steve E. ;
Rohlfing, Curt L. ;
Jaisson, Stephane ;
Gillery, Philippe ;
Roberts, William L. .
CLINICA CHIMICA ACTA, 2012, 413 (7-8) :819-821