A New Automated Sample Transfer System for Instrumental Neutron Activation Analysis

被引:14
作者
Ismail, S. S. [1 ]
机构
[1] Vienna Univ Technol, Inst Atom, A-1020 Vienna, Austria
来源
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY | 2010年
关键词
D O I
10.1155/2010/389374
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fully automated and fast pneumatic transport system for short-time activation analysis was recently developed. It is suitable for small nuclear research reactors or laboratories that are using neutron generators and other neutron sources. It is equipped with a programmable logic controller, software package, and 12 devices to facilitate optimal analytical procedures. 550 ms were only necessary to transfer the irradiated capsule (diameter: 15 mm, length: 50 mm, weight: 4 gram) to the counting chamber at a distance of 20 meters using pressurized air (4 bars) as a transport gas.
引用
收藏
页数:8
相关论文
共 6 条
[1]   New pneumatic transfer systems for neutron activation analysis at the HANARO research reactor [J].
Chung, Yong-Sam ;
Kim, Sun-Ha ;
Moon, Jong-Hwa ;
Baek, Sung-Yeol ;
Kim, Young-Jin ;
Kim, Hark-Rho ;
Park, Il-Jin ;
Min, Kyung-Sik .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2008, 278 (03) :707-712
[2]   High counting efficiency Compton suppression system for automatic-INAA [J].
Ismail, S. S. .
INSTRUMENTATION SCIENCE & TECHNOLOGY, 2007, 35 (03) :307-322
[3]   Dynamic irradiation counting systems for fast, short-time activation analysis using a fully automatic moveable counting chamber and a digital spectrometer [J].
Ismail, SS ;
Brezovits, K ;
Klikovich, W .
INSTRUMENTATION SCIENCE & TECHNOLOGY, 2001, 29 (04) :255-266
[4]   Quality of high count rate high resolution short-time activation analysis [J].
Ismail, SS .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-LETTERS, 1996, 213 (04) :291-310
[5]  
Ismail SS, 2008, EGYPT J CHEM, V51, P29
[6]  
Parry S., 1991, Activation spectrometry in chemical analysis