A Comparison of the Performance of High QE Photomultiplier Tubes to Conventional Photomultiplier Tubes

被引:1
作者
Liang, H. [1 ]
Oldendick, J. [1 ]
Li, Y. [1 ]
Ordonez, C. E. [1 ]
Chang, W. [1 ]
机构
[1] Rush Univ, Med Ctr, Dept Diagnost Radiol, Chicago, IL 60612 USA
来源
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5 | 2009年
关键词
D O I
10.1109/NSSMIC.2009.5401835
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have carried out experiments to assess the performance advantage of High Quantum Efficiency (HQE) Photomultiplier Tubes (PMTs) over conventional PMTs for possible use in a cardiac SPECT system. In particular, we have conducted tests of pulse height resolution and position sensitivity of HQE PMTs and compared their performance to conventional PMTs. Furthermore, we have conducted a cathode sensitivity scan and pulse height resolution scan of four HQE PMTs and four conventional PMTs. The HQE PMTs outperformed the conventional PMTs in both pulse height resolution and position sensitivity. Finally, we tested the four HQE PMTs on a 3 mm x 3 mm pixellated NaI(Tl) detector module and were able to resolve the pixels with 8.3% average energy resolution.
引用
收藏
页码:3615 / 3617
页数:3
相关论文
共 1 条
[1]   C-SPECT-A Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential [J].
Chang, Wei ;
Ordonez, Caesar E. ;
Liang, Haoning ;
Li, Yusheng ;
Liu, Jingai .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) :2659-2671