Effect of improved navigation performance on the accuracy of implant placement in total hip arthroplasty with a CT-based navigation system

被引:28
作者
Nakahara, Ichiro [1 ]
Kyo, Takayuki [2 ]
Kuroda, Yasuo [1 ]
Miki, Hidenobu [1 ]
机构
[1] Osaka Natl Hosp, Dept Orthopaed Surg, Chuo Ku, 2-1-14 Oenzaka, Osaka 5400006, Japan
[2] Bellland Gen Hosp, Dept Orthopaed Surg, Naka Ku, 500-3 Higashiyama, Sakai, Osaka, Japan
关键词
Total hip arthroplasty; CT-based navigation; Implant alignment; Implant position; Accuracy; TOMOGRAPHY-BASED NAVIGATION; ACETABULAR COMPONENT; POLYETHYLENE; ORIENTATION; STABILITY; POSITION; WEAR;
D O I
10.1007/s10047-018-1041-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A computed tomography (CT)-based navigation system is one of the support tools to place implant with appropriate alignment and position in total hip arthroplasty (THA). To determine whether the higher performance of the navigation would further improve the accuracy of implant placement in the clinical setting, we retrospectively compared the navigation accuracy of two different versions of a navigation system. The newer version of the navigation system had an upgraded optical sensor with superior positional accuracy. Navigation accuracy, defined as differences between postoperative measurements on CT images and intraoperative records on the navigation system, of 49 THAs performed with the newer version of the navigation system was compared with that of 49 THAs performed with the older version. With the newer version, the mean absolute accuracy (95% limits of agreement) of implant alignment was 1.2 degrees (+/- 3.3 degrees) for cup inclination, 1.0 degrees (+/- 2.4 degrees) for cup anteversion, 2.0 degrees (+/- 4.9 degrees) for stem anteversion, and 1.1 degrees (+/- 2.4 degrees) for stem valgus angle. The accuracy of the implant position was 1.5 mm (+/- 3.1 mm), 1.3 mm (+/- 3.0 mm), and 1.5 mm (+/- 3.1 mm) for cup x-, y-, and z-axes, respectively, 1.6 mm (+/- 3.2 mm), 1.4 mm (+/- 2.9 mm), and 1.5 mm (+/- 2.7 mm) for stem x-, y-, and z-axes, respectively, and 2.4 mm (+/- 4.5 mm) for leg length discrepancy. The values for the newer version were significantly more accurate with less variation compared to those of the older version. With upgraded navigation performance, more accurate implant placement was demonstrated in the clinical setting.
引用
收藏
页码:340 / 347
页数:8
相关论文
共 37 条
[1]   Stability and leg length equality in total hip arthroplasty [J].
Austin, MS ;
Hozack, WJ ;
Sharkey, PF ;
Rothman, RH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (03) :88-90
[2]   Differences between the wear couples metal-on-polyethylene and ceramic-on-ceramic in the stability against dislocation of total hip replacement [J].
Bader, R ;
Steinhauser, E ;
Zimmermann, S ;
Mittelmeier, W ;
Scholz, R ;
Busch, R .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (06) :711-718
[3]  
Barrack Robert L, 2003, J Am Acad Orthop Surg, V11, P89
[4]  
Chen PY, 2005, J FORMOS MED ASSOC, V104, P370
[5]   Accuracy of combined anteversion in image-free navigated total hip arthroplasty: stem-first or cup-first technique? [J].
Fukunishi, Shigeo ;
Nishio, Shoji ;
Fujihara, Yuki ;
Okahisa, Shohei ;
Takeda, Yu ;
Fukui, Tomokazu ;
Yoshiya, Shinichi .
INTERNATIONAL ORTHOPAEDICS, 2016, 40 (01) :9-13
[6]   Comparison of navigation accuracy in THA between the mini-anterior and -posterior approaches [J].
Hananouchi, Takehito ;
Takao, Masaki ;
Nishii, Takashi ;
Miki, Hidenobu ;
Iwana, Daiki ;
Yoshikawa, Hideki ;
Sugano, Nobuhiko .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (01) :20-25
[7]   Effect of CT slice thickness on accuracy of implant positioning in navigated total hip arthroplasty [J].
Hirasawa, N. ;
Matsubara, M. ;
Ishii, K. ;
Hagio, S. ;
Okuda, N. ;
Sekiya, I. ;
Muneta, T. .
COMPUTER AIDED SURGERY, 2010, 15 (4-6) :83-89
[8]   Voxel similarity measures for 3-D serial MR brain image registration [J].
Holden, M ;
Hill, DLG ;
Denton, ERE ;
Jarosz, JM ;
Cox, TCS ;
Rohlfing, T ;
Goodey, J ;
Hawkes, DJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (02) :94-102
[9]   Accuracy of angle and position of the cup using computed tomography-based navigation systems in total hip arthroplasty [J].
Iwana, Daiki ;
Nakamura, Nobuo ;
Miki, Hidenobu ;
Kitada, Makoto ;
Hananouchi, Takehito ;
Sugano, Nobuhiko .
COMPUTER AIDED SURGERY, 2013, 18 (5-6) :187-194
[10]   Navigated non-image-based positioning of the acetabulum during total hip replacement [J].
Jenny, Jean-Yves ;
Boeri, Cyril ;
Dosch, Jean-Claude ;
Uscatu, Marius ;
Ciobanu, Eugen .
INTERNATIONAL ORTHOPAEDICS, 2009, 33 (01) :83-87