An Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT Imaging

被引:197
|
作者
Lois, Cristina [1 ,2 ]
Jakoby, Bjoern W. [1 ,3 ,4 ]
Long, Misty J. [1 ]
Hubner, Karl F. [1 ]
Barker, David W. [1 ]
Casey, Michael E. [3 ]
Conti, Maurizio [3 ]
Panin, Vladimir Y. [3 ]
Kadrmas, Dan J. [5 ]
Townsend, David W. [1 ]
机构
[1] Univ Tennessee, Med Ctr, Knoxville, TN USA
[2] Univ Santiago de Compostela, Dept Fis Particulas, Santiago De Compostela, Spain
[3] Siemens Healthcare Mol Imaging, Knoxville, TN 37932 USA
[4] Univ Surrey, Dept Phys, Guildford GU2 5XH, Surrey, England
[5] Univ Utah, Dept Radiol, Salt Lake City, UT 84132 USA
关键词
TOF; PET; PET/CT; lesion detectability; oncology; POSITRON-EMISSION-TOMOGRAPHY; RECONSTRUCTION; PERFORMANCE; SCANNER;
D O I
10.2967/jnumed.109.068098
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The introduction of fast scintillators with good stopping power for 511-keV photons has renewed interest in time-of-flight (TOF) PET. The ability to measure the difference between the arrival times of a pair of photons originating from positron annihilation improves the image signal-to-noise ratio (SNR). The level of improvement depends upon the extent and distribution of the positron activity and the time resolution of the PET scanner. While specific estimates can be made for phantom imaging, the impact of TOF PET is more difficult to quantify in clinical situations. The results presented here quantify the benefit of TOF in a challenging phantom experiment and then assess both qualitatively and quantitatively the impact of incorporating TOF information into the reconstruction of clinical studies. A clear correlation between patient body mass index and gain in SNR was observed in this study involving 100 oncology patient studies, with a gain due to TOF ranging from 1.1 to 1.8, which is consistent with the 590-ps time resolution of the TOF PET scanner. The visual comparison of TOF and non-TOF images performed by two nuclear medicine physicians confirmed the advantages of incorporating TOF into the reconstruction, advantages that include better definition of small lesions and image details, improved uniformity, and noise reduction.
引用
收藏
页码:237 / 245
页数:9
相关论文
共 50 条
  • [1] Clinical impact of time-of-flight PET
    Soussan, M.
    Galas, J-L
    Ouvrier, M-J
    Pop, G.
    Neuman, A.
    Weinmann, P.
    MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2011, 35 (07): : 401 - 405
  • [2] Update on Time-of-Flight PET Imaging
    Surti, Suleman
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (01) : 98 - 105
  • [3] Impact of Time-of-Flight on PET Tumor Detection
    Kadrmas, Dan J.
    Casey, Michael E.
    Conti, Maurizio
    Jakoby, Bjoern W.
    Lois, Cristina
    Townsend, David W.
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (08) : 1315 - 1323
  • [4] Improving time resolution in time-of-flight PET
    Conti, Maurizio
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S194 - S198
  • [5] Recent developments in time-of-flight PET
    Vandenberghe, S.
    Mikhaylova, E.
    D'Hoe, E.
    Mollet, P.
    Karp, J. S.
    EJNMMI PHYSICS, 2016, 3 (01):
  • [6] Physical and clinical performance of the mCT time-of-flight PET/CT scanner
    Jakoby, B. W.
    Bercier, Y.
    Conti, M.
    Casey, M. E.
    Bendriem, B.
    Townsend, D. W.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08) : 2375 - 2389
  • [7] A Clinical and Experimental Comparison of Time of Flight PET/MRI and PET/CT Systems
    Oprea-Lager, Daniela E.
    Yaqub, Maqsood
    Pieters, Indra C.
    Reinhard, Rinze
    van Moorselaar, Reindert J. A.
    van den Eertwegh, Alfons J. M.
    Hoekstra, Otto S.
    Lammertsma, Adriaan A.
    Boellaard, Ronald
    MOLECULAR IMAGING AND BIOLOGY, 2015, 17 (05) : 714 - 725
  • [8] Time-of-Flight Information Improved the Detectability of Subcentimeter Spheres Using a Clinical PET/CT Scanner
    Hashimoto, Naoki
    Morita, Keishin
    Tsutsui, Yuji
    Himuro, Kazuhiko
    Baba, Shingo
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2018, 46 (03) : 268 - 273
  • [9] Focus on time-of-flight PET: the benefits of improved time resolution
    Conti, Maurizio
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (06) : 1147 - 1157
  • [10] The Effect of Defective PET Detectors in Clinical Simultaneous [18F]FDG Time-of-Flight PET/MR Imaging
    ter Voert, Edwin E. G. W.
    Delso, Gaspar
    Barbosa, Felipe de Galiza
    Huellner, Martin
    Veit-Haibach, Patrick
    MOLECULAR IMAGING AND BIOLOGY, 2017, 19 (04) : 626 - 635